Clinical application of a model-based cardiac stroke volume estimation method

被引:1
|
作者
Smith, Rachel [1 ]
Balmer, Joel [1 ]
Pretty, Christopher G. [1 ]
Shaw, Geoffrey M. [2 ]
Chase, J. Geoffrey [1 ]
机构
[1] Univ Canterbury, Dept Mech Engn, Canterbury, New Zealand
[2] Christchurch Hosp Intens Care Unit, Christchurch, New Zealand
来源
IFAC PAPERSONLINE | 2020年 / 53卷 / 02期
基金
欧盟地平线“2020”;
关键词
Pulse contour analysis; Pressure contour analysis; Windkessel model; Stroke volume; Cardiac output; Hemodynamic monitoring; Intensive care; PRESSURE; OUTPUT; PULSE; ANESTHESIA; ACCURACY; SHOCK;
D O I
10.1016/j.ifacol.2020.12.435
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A system is needed for monitoring stroke volume (SV) and cardiac output (CO) in unstable patients which is non-additionally invasive, reproducible and reliable in a variety of physiological states. This study evaluates SV estimation accuracy of a non-additionally invasive pulse contour analysis method implemented using a 3-element Windkessel model. The model lumps the properties of the arterial system into 3 parameters: characteristic impedance of the proximal aorta (Z), and resistance (R) and compliance (C) of the systemic arteries. Parameter products ZC and RC are dynamically identified from measured femoral arterial pressure waveforms, and Z is a static parameter obtained by calibration. The accuracy of the model is evaluated for a cohort of 9 liver transplant patients, using thermodilution as a reference method. Data were obtained from Vital Data Bank (VitalDB). The study thus provides independent assessment of a pulse contour analysis, proven in animal studies, in an uncontrolled clinical environment. The model tracked trends in SV well over the course of the surgery. However, the 95% range for percentage error was -88% to +53%, outside acceptable limits of +/- 45%. Main areas contributing to error for the model include the changing extent of reflected waves in the arterial system, dynamic response characteristics of fluid-filled pressure catheters, and the assumption of fixed Z parameter. Further investigation is needed to consider the contribution of these factors to SV estimation error by the model. Copyright (C) 2020 The Authors.
引用
收藏
页码:16137 / 16142
页数:6
相关论文
共 50 条
  • [1] Incorporating pulse wave velocity into model-based pulse contour analysis method for estimation of cardiac stroke volume
    Smith, Rachel
    Balmer, Joel
    Pretty, Christopher G.
    Mehta-Wilson, Tashana
    Desaive, Thomas
    Shaw, Geoffrey M.
    Chase, J. Geoffrey
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 195
  • [2] Clinically applicable model-based method, for physiologically accurate flow waveform and stroke volume estimation
    Balmer, Joel
    Pretty, Christopher G.
    Davidson, Shaun
    Mehta-Wilson, Tashana
    Desaive, Thomas
    Smith, Rachel
    Shaw, Geoffrey M.
    Chase, J. Geoffrey
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 185
  • [3] Tube-load model: A clinically applicable pulse contour analysis method for estimation of cardiac stroke volume
    Smith, Rachel
    Murphy, Liam
    Pretty, Christopher G.
    Desaive, Thomas
    Shaw, Geoffrey M.
    Chase, J. Geoffrey
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2021, 204
  • [4] Model-Based Estimation of the Cardiac Output in Exercise Scenario
    Dan, Ana Maria
    2016 12TH IEEE INTERNATIONAL SYMPOSIUM ON ELECTRONICS AND TELECOMMUNICATIONS (ISETC'16), 2016, : 285 - 288
  • [5] Linear model-based estimation of blood pressure and cardiac output for Normal and Paranoid cases
    Mohamed Abdelkader Aboamer
    Ahmad Taher Azar
    Khaled Wahba
    Abdallah S. A. Mohamed
    Neural Computing and Applications, 2014, 25 : 1223 - 1240
  • [6] Linear model-based estimation of blood pressure and cardiac output for Normal and Paranoid cases
    Aboamer, Mohamed Abdelkader
    Azar, Ahmad Taher
    Wahba, Khaled
    Mohamed, Abdallah S. A.
    NEURAL COMPUTING & APPLICATIONS, 2014, 25 (06) : 1223 - 1240
  • [7] Deep Learning-Based Stroke Volume Estimation Outperforms Conventional Arterial Contour Method in Patients with Hemodynamic Instability
    Moon, Young-Jin
    Moon, Hyun S.
    Kim, Dong-Sub
    Kim, Jae-Man
    Lee, Joon-Kyu
    Shim, Woo-Hyun
    Kim, Sung-Hoon
    Hwang, Gyu-Sam
    Choi, Jae-Soon
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (09)
  • [8] SIMULATION METHOD FOR CARDIAC STROKE VOLUME ESTIMATION BY INTRACARDIAC ELECTRICAL-IMPEDANCE MEASUREMENT
    BARAK, C
    LEVIATAN, Y
    INBAR, GF
    HOEKSTEIN, KN
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (05) : 473 - 480
  • [9] Model-based Measurement of the Stroke Volume of a VAD in Close-to-Real Conditions
    Sulej, Wojciech
    Murawski, Krzysztof
    Pustelny, Tadeusz
    13TH CONFERENCE ON INTEGRATED OPTICS: SENSORS, SENSING STRUCTURES, AND METHODS, 2018, 10830
  • [10] Application of the flow convergence region method to the determination of stroke volume and cardiac output in patients with mitral stenosis
    Horibe, N
    Matsumoto, M
    Deng, YB
    CANADIAN JOURNAL OF CARDIOLOGY, 1996, 12 (04) : 363 - 369