System for monitoring UV radiation level in phototherapy cabins in Poland

被引:3
作者
Narbutt, Joanna [1 ]
Pawlaczyk, Mariola [2 ]
Sysa-Jedrzejowska, Anna [3 ]
Krzyscin, Janusz W. [4 ]
Sobolewski, Piotr [4 ]
Rajewska-Wiech, Bonawentura [4 ]
Lesiak, Aleksandra [1 ]
机构
[1] Med Univ Lodz, Dept Dermatol, Lodz, Poland
[2] Poznan Univ Med Sci, Dept Biol & Environm Study, Poznan, Poland
[3] Univ Social Sci, Lodz, Poland
[4] Polish Acad Sci, Inst Geophys, PL-01452 Warsaw, Poland
关键词
phototherapy; UV radiation; medical cabinet; dose calibration; ULTRAVIOLET-B PHOTOTHERAPY; CALIBRATION; GUIDELINES; DOSIMETRY;
D O I
10.5114/aoms.2014.47834
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Ultraviolet phototherapy (UVP) is widely used in dermatological practice for the treatment of various skin diseases. Numerous studies support its beneficial curing effectiveness; however, overexposure to ultraviolet radiation can cause adverse health effects, such as sunburn reaction, erythema response, cataract, skin aging, etc. For these reasons, it is of special importance to monitor performance of UVP cabins using a calibration system to evaluate the UV doses incident upon the patient. Material and methods: A mechanized cabin control system (CCS) is proposed. It consists of radiometers with a wide and narrow field of view to estimate the body irradiation and to identify malfunctioning cabin tubes. Quality control and quality assurance procedures are developed to keep high accuracy of the calibration procedure. The CCS has been used in the examination of two different types of UVP cabins routinely working in Poland. Results: It allows precise calculation of UV doses and spatial variability of UV radiance inside the cabin, thus providing uncertainties of the doses assigned by medical staff. The CCS could potentially serve as a primary standard for monitoring various UVP cabins working in Poland. Conclusions: The methodology developed to quantify UV doses in UVP cabins may be easily extended to any UV radiation source.
引用
收藏
页码:1244 / 1254
页数:11
相关论文
共 14 条
[1]   An automated dosimetry system for testing whole-body ultraviolet phototherapy cabinets [J].
Currie, GD ;
Evans, AL ;
Smith, D ;
Martin, CJ ;
McCalman, S ;
Bilsland, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02) :333-346
[2]  
Diffey BL, 1997, 76 I PHYS ENG MED
[3]   ULTRAVIOLET-RADIATION IN THE TREATMENT OF SKIN-DISEASE [J].
GREEN, C ;
DIFFEY, BL ;
HAWK, JLM .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) :1-20
[4]  
HENSELER T, 1981, LANCET, V1, P853
[5]   An update and guidance on narrowband ultraviolet B phototherapy: a British Photodermatology Group workshop report [J].
Ibbotson, SH ;
Bilsland, D ;
Cox, NH ;
Dawe, RS ;
Diffey, B ;
Edwards, C ;
Farr, PM ;
Ferguson, J ;
Hart, G ;
Hawk, J ;
Lloyd, J ;
Martin, C ;
Moseley, H ;
McKenna, K ;
Rhodes, LE ;
Taylor, DK .
BRITISH JOURNAL OF DERMATOLOGY, 2004, 151 (02) :283-297
[6]   An insight into the contributions of self-shielding and lamp reflectors to patient exposure in phototherapy units [J].
Langmack, KA .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (01) :207-214
[7]   Vitamin D serum level changes in psoriatic patients treated with narrowband ultraviolet B phototherapy are related to the season of the irradiation [J].
Lesiak, Aleksandra ;
Narbutt, Joanna ;
Pawlaczyk, Mariola ;
Sysa-Jedrzejowska, Anna ;
Krzyscin, Janusz .
PHOTODERMATOLOGY PHOTOIMMUNOLOGY & PHOTOMEDICINE, 2011, 27 (06) :304-310
[8]   Variation in calibration of hand-held ultraviolet (UV) meters for psoralen plus UVA and narrow-band UVB phototherapy [J].
Lloyd, JJ .
BRITISH JOURNAL OF DERMATOLOGY, 2004, 150 (06) :1162-1166
[9]   The importance of radiometer angular response for ultraviolet phototherapy dosimetry [J].
Martin, CJ ;
Currie, GD ;
Pye, SD .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (04) :843-855
[10]   A study of the correction factor for ultraviolet phototherapy dose measurements made by the indirect method [J].
Martin, CJ ;
Clouting, H ;
Aitken, A .
BRITISH JOURNAL OF DERMATOLOGY, 2003, 149 (06) :1227-1231