An investigation of using log-file analysis for automated patient-specific quality assurance in MRgRT

被引:9
作者
Lim, Seng Boh [1 ]
Godoy Scripes, Paola [1 ]
Napolitano, Mary [2 ]
Subashi, Ergys [1 ]
Tyagi, Neelam [1 ]
Cervino Arriba, Laura [1 ]
Lovelock, Dale Michael [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
[2] Standard Imaging Inc, Middleton, WI USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2021年 / 22卷 / 09期
关键词
elekta unity; linacview; log-file analysis; MRI guided adaptive radiotherapy; MRgRT; patient specific quality assurance; MLC PERFORMANCE; IMRT; QA;
D O I
10.1002/acm2.13361
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA. Method All treatment plans were delivered in 1.5T Unity MR-Linac (Elekta). A Unity compatible version of LinacView (Standard Imaging) was commissioned to automatically monitor and analyze the log-files. A total of 220 fields were delivered and measured by ArcCheck-MR (Sun Nuclear) and LinacView. Thirty incorrectly matched fields were also delivered to check for error detection sensitivity. The gamma analysis, gamma, with 3%, 3 mm criteria was used in both ArcCheck-MR and LinacView. Additionally, the gantry angle, jaws, and multileaf collimators (MLC) positions reported in the log-file were compared with plan positions using TG-142 criteria. Result The gamma (3%, 3 mm) for the 190 plans were found to be between the range of 72.5%-100.0% and 95.4%-100.0% for ArcCheck-MR and LinacVeiw, respectively. All the delivered gantry angle and jaws were found to be within 0.2 degrees and 2 mm. MLCs that were outside the guard leaves or under the diaphragms were found to have more than 1.0 mm discrepancy. This was attributed to the linac internal override for these MLCs and had no dosimetric impact. Excluding these discrepancies, all MLC positions were found to be within 1.0 mm. The gamma (3%, 3 mm) for the 30 incorrectly matched fields were found to be 3.9%-84.8% and 0.1%-64.4% for ArcCheck-MR and LinacVeiw, respectively. Conclusion Significant ranked correlation demonstrates the automated log-file analysis can be used for PSQA and expedite the ART workflow. Ongoing PSQA will be compared with log-file analysis to investigate the longer term reproducibility and correlation.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 50 条
  • [21] Simplification of head and neck volumetric modulated arc therapy patient-specific quality assurance, using a Delta4 PT
    Sasaki, Motoharu
    Sugimoto, Wataru
    Ikushima, Hitoshi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (05) : 793 - 800
  • [22] A tool for patient-specific prediction of delivery discrepancies in machine parameters using trajectory log files
    Chuang, Kai-Cheng
    Giles, William
    Adamson, Justus
    MEDICAL PHYSICS, 2021, 48 (03) : 978 - 990
  • [23] Patient-Specific Quality Assurance Using Monte Carlo Dose Calculation and Elekta Log Files for Prostate Volumetric-Modulated Arc Therapy
    Katsuta, Yoshiyuki
    Kadoya, Noriyuki
    Fujita, Yukio
    Shimizu, Eiji
    Matsunaga, Kenichi
    Sawada, Kinya
    Matsushita, Haruo
    Majima, Kazuhiro
    Jingu, Keiichi
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2017, 16 (06) : 1220 - 1225
  • [24] Feasibility study of structural similarity index for patient-specific quality assurance
    Lee, Jae Choon
    Park, Hyeong Wook
    Kang, Young Nam
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (03):
  • [25] Quality assurance-based optimization (QAO): Towards improving patient-specific quality assurance in volumetric modulated arc therapy plans using machine learning
    Wall, Phillip D. H.
    Fontenot, Jonas D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 87 : 136 - 143
  • [26] Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT
    Agnew, Christina E.
    Irvine, Denise M.
    McGarry, Conor K.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (06): : 204 - 216
  • [27] Patient-specific quality assurance for intracranial cases in robotic radiosurgery system
    Koksal, Canan
    Akbas, Ugur
    Kesen, Nazmiye Donmez
    Okutan, Murat
    Bilge, Hatice
    Kemikler, Gonul
    JOURNAL OF BUON, 2018, 23 (01): : 179 - 184
  • [28] 2D dose reconstruction of IMRT patient-specific QA based on log file
    Mubarok, Sayid
    Wahyu, Wibowo Edy
    Supriyanto, Pawiro Ardjo
    RADIATION PHYSICS AND CHEMISTRY, 2020, 166
  • [29] An automated workflow for patient-specific quality control of contour propagation
    Beasley, William J.
    McWilliam, Alan
    Slevin, Nicholas J.
    Mackay, Ranald I.
    van Herk, Marcel
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (24) : 8577 - 8586
  • [30] ROC analysis in patient specific quality assurance
    Carlone, Marco
    Cruje, Charmainne
    Rangel, Alejandra
    McCabe, Ryan
    Nielsen, Michelle
    MacPherson, Miller
    MEDICAL PHYSICS, 2013, 40 (04)