Comparison Between Locking and Non-Locking Plates for Fixation of Metacarpal Fractures in an Animal Model

被引:30
作者
Ochman, Sabine
Doht, Stephanie
Paletta, Juergen
Langer, Martin
Raschke, Michael J.
Meffert, Rainer H.
机构
[1] Univ Hosp, Dept Trauma Hand & Reconstruct Surg, Munster, Germany
[2] Univ Wurzburg, Dept Trauma Hand Plast & Reconstruct Surg, Wurzburg, Germany
[3] Dept Orthoped Surg, Marburg, Germany
来源
JOURNAL OF HAND SURGERY-AMERICAN VOLUME | 2010年 / 35A卷 / 04期
关键词
Biomechanics; fracture; internal fixation; locking plate; metacarpal; DISTAL RADIUS FRACTURES; INTERNAL-FIXATION; PHALANGEAL FRACTURES; HAND FRACTURES; PROXIMAL HUMERUS; SCREW FIXATION; COMPLICATIONS; STIFFNESS; CONSTRUCT; FAILURE;
D O I
10.1016/j.jhsa.2010.01.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose The use of locking plates increases the primary load to failure, thereby reducing the rate of implant-related failure. The good clinical and biomechanical results of locking plates in long bones might be applicable to treatment of metacarpal fractures. The purpose of this study was to determine strength and stiffness of locking plates in a metacarpal fracture model with mono- and bicortical screw fixation in comparison to non-locking plate mono- and bicortical screw fixation, with both types of plates placed at the dorsal side of the bone. Methods Fresh second metacarpals from domestic pigs (n=40) were randomized in 4 equal groups. Short, oblique, mid-shaft fractures were generated, using a standardized 3-point bending method. Fractures were plated with non-locking, titanium, 1-mm-thick monocortical (group 1, n=10) or bicortical (group 2, n = 10) plates (Leibinger-Stryker; Stryker Corp, Freiburg, Germany). Newly designed locking titanium plates with the same width and thickness (Leibinger-Stryker) were used in the same manner for groups 3 (monocortical) and 4 (bicortical). The metacarpals were then tested to load to failure in a cantilever bending mode. Results Bicortical, non-locking fixation (group 2, 359 +/- 90 N) had a higher load to failure than monocortical non-locking fixation (group 1, 250 +/- 56 N) in testing the maximum load to failure (p < .01). There was no significant difference in stiffness between group 1 (46 +/- 12 N/mm) and group 2 (56 +/- 21 N/mm). The difference in maximum load to failure between monocortical (group 3, 440 +/- 85N) and bicortical (group 4, 378 +/- 116 N) locking plate stabilization was not significant. Also, there was no significant difference in stiffness between monocortical (group 3, 83 +/- 35 N/mm) and bicortical locking plates (group 4, 70 +/- 31 N/mm). Comparing non-locking (group 1) and locking plates in a monocortical fixation technique (group 3) demonstrated significant differences in maximum load to failure (group 1, 250 +/- 56 N; group 3, 440 +/- 85 N) and stiffness (group 1, 46 +/- 12 N/mm; group 3, 83 +/- 35 N/mm). The stability of monocortical locking plates was stronger, although not statistically significant, than the non-locking bicortical plates (load to failure, 440 +/- 85 N vs 359 +/- 90 N; stiffness, 83 +/- 35 N/mm vs 56 +/- 21 N/mm). Conclusions The new generation of locking plates can be used to achieve a higher stability for fixation of metacarpal fractures. Monocortical, stable fixation can minimize flexor tendon interference and probably reduce bone and soft tissue trauma. (J Hand Surg 2010;35A:597-603. Copyright (C) 2010 by the American Society for Surgery of the Hand. All rights reserved.)
引用
收藏
页码:597 / 603
页数:7
相关论文
共 50 条
[41]   Biomechanical Performance of Lateral Versus Dual Locking Plates for Calcaneal Fractures [J].
Maxwell, Abby B. ;
Owen, John R. ;
Gilbert, Todd M. ;
Romash, Michael M. ;
Wayne, Jennifer S. ;
Adelaar, Robert S. .
JOURNAL OF FOOT & ANKLE SURGERY, 2015, 54 (05) :830-835
[42]   Volar locking plate (VLP) versus non-locking plate (NLP) in the treatment of die-punch fractures of the distal radius, an observational study [J].
Zhang, Xiong ;
Hu, Chunhe ;
Yu, Kunlun ;
Bai, Jiangbo ;
Tian, Dehu ;
Xu, Yi ;
Zhang, Bing .
INTERNATIONAL JOURNAL OF SURGERY, 2016, 34 :142-147
[43]   Comparison of different locking plate fixation methods in lateral malleolus fractures [J].
Bilgetekin, Yenel Gurkan ;
Catma, Mehmet Faruk ;
Ozturk, Alper ;
Unlu, Serhan ;
Ersan, Onder .
FOOT AND ANKLE SURGERY, 2019, 25 (03) :366-370
[44]   Biomechanical Comparison of Polyaxial-Type Locking Plates and a Fixed-Angle Locking Plate for Internal Fixation of Distal Femur Fractures [J].
Otto, Randall J. ;
Moed, Berton R. ;
Bledsoe, J. Gary .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2009, 23 (09) :645-652
[45]   The outcome of surgical management of proximal humeral fractures using locking plates: comparison between Locking plates with different geometry [J].
Chen, Jiun-Liang ;
Chang, Fu-Chun ;
Lin, Shih-Jie ;
Chuang, Po-Yao ;
Peng, Kuo-Ti ;
Huang, Kuo-Chin ;
Huang, Tsan-Wen .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2018, 27 (12) :2159-2166
[46]   Fixation of Distal Femur Fractures With the Use of Periarticular Tibial Locking Plates [J].
Friedman, Lisa G. M. ;
Maniar, Hemil ;
Horwitz, Daniel S. .
TECHNIQUES IN ORTHOPAEDICS, 2024, 39 (02) :55-58
[47]   The effect of the combination of locking screws and non-locking screws on the torsional properties of a locking-plate construct [J].
Gordon, S. ;
Moens, N. M. M. ;
Runciman, J. ;
Monteith, G. .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2010, 23 (01) :7-13
[48]   Outcomes of locking plate fixation with fibular allograft augmentation for proximal humeral fractures in osteoporotic patients COMPARISON WITH LOCKING PLATE FIXATION ALONE [J].
Lee, S. H. ;
Han, S. S. ;
Yoo, B. M. ;
Kim, J. W. .
BONE & JOINT JOURNAL, 2019, 101B (03) :260-265
[49]   Subtrochanteric Fractures: Comparison of Proximal Femur Locking Plate and Intramedullary Locking Nail Fixation Outcome [J].
Mirbolook, Ahmadreza ;
Siavashi, Babak ;
Jafarinezhad, Adel Ebrahimpour ;
Jahromi, Sina Khajeh ;
Farahmand, Maral ;
Rad, Melina Roohi ;
Kahe, Mohammad Amin .
INDIAN JOURNAL OF SURGERY, 2015, 77 :S795-S798
[50]   Locking Plate Fixation for Canine Acetabular Fractures [J].
Piana, Francesco ;
Solano, Miguel ;
Kalff, Stephen ;
Yeadon, Russell .
VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2020, 33 (04) :294-300