A retrospective study of the hearing results obtained after stapedotomy by the implantation of two Teflon pistons with a different diameter

被引:27
作者
Grolman, W
Tange, RA
de Bruijn, AJG
Hart, AAM
Schouwenburg, PF
机构
[1] Univ Amsterdam, Acad Med Ctr, Dept Otorhinolaryngol, NL-1105 AZ Amsterdam, Netherlands
[2] Univ Amsterdam, Acad Med Ctr, Dept Med Stat, NL-1105 AZ Amsterdam, Netherlands
关键词
hearing loss; otosclerosis; stapedectomy; teflon ossicular implant; audiometry;
D O I
10.1007/BF02439972
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
This study reports the evaluation of hearing results after implantation of a Teflon piston of a different diameter in cases of otosclerosis requiring stapedotomy. By random selection, a Teflon piston with a shaft diameter of 0.3 mm was inserted in 34 cases and a piston with a shaft diameter of 0.3 mm in 26 cases. A retrospective analysis of the pre-and postsurgery audiological results of these two patient groups was carried out by microcomputer. A repeated statistical measures analysis of variance was used to test and estimate the air-conduction frequency-specific differences between the two prostheses with respect to changes in pre-and postoperative healing. The results of this comparative study of the two Teflon pistons with different diameters indicate statistically a greater hearing gain for the 0.4-mm prosthesis, especially in the lower frequencies.
引用
收藏
页码:422 / 424
页数:3
相关论文
共 13 条
[1]  
Beales PH, 1987, S BROWNS OTOLARYNGOL, V3, P319
[2]  
Dubreuil C, 1994, Ann Otolaryngol Chir Cervicofac, V111, P249
[3]  
FISCH U, 1982, AM J OTOL, V4, P112
[4]  
FISH U, 1994, TYMPANOPLASTY MASTOI, P214
[5]  
HERZOG JA, 1991, AM J OTOL, V12, P16
[6]   CONSIDERATIONS ABOUT SURGICAL APPROACH IN STAPEDECTOMY [J].
MARQUET, J ;
VANCAMP, KJ ;
CRETEN, WL .
ACTA OTO-LARYNGOLOGICA, 1972, 74 (06) :406-&
[7]  
MARQUET J, 1983, J LARYNGOL OTOL, P78
[8]  
MARQUET JFE, 1990, OTOSCLEROSIS, P83
[9]  
PORTMANN M, 1975, TRAITE TECHNIQUE CHI, V1, P108
[10]  
SHEA J J Jr, 1958, Ann Otol Rhinol Laryngol, V67, P932