An efficient procedure for tomotherapy treatment plan verification using the on-board detector

被引:9
作者
Pisaturo, O. [1 ]
Mieville, F. [1 ]
Tercier, P-A [1 ]
Allal, A. S. [1 ]
机构
[1] HFR Hop Fribourgeois, Dept Radiat Oncol, Fribourg, Switzerland
关键词
tomotherapy; treatment plan verification; dose calculation model; DELIVERY VERIFICATION; HELICAL TOMOTHERAPY; DOSE VERIFICATION; IMRT; FILM;
D O I
10.1088/0031-9155/60/4/1625
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this work, a fast and simple procedure for tomotherapy treatment plan verification using the on-board detector (OBD) has been developed. This procedure allows verification of plans with static and dynamic jaws (TomoEDGE). A convolution-based calculation model has been derived in order to link the leaf control sinogram from the treatment planning system to the data acquired by the OBD during a static couch procedure. The convolution kernel has been optimized using simple plans calculated in the Tomotherapy Cheese phantom. The optimal kernel has been found to be a lorentzian function, whose parameter Gamma is 0.186 for the 1 cm jaw opening, 0.232 for the 2.5 cm jaw opening and 0.373 for the 5 cm jaw opening. The evaluation has been performed with gamma-index analysis. The dose criterion was 3% of the 95th percentile of the dose distribution and the distance-to-agreement criterion is 2 mm. In order to validate the procedure, it has been applied to around 50 clinical treatment plans, which had already been validated by the Delta4 phantom (Scandidos, Sweden). 96% of the tested plans have passed the criteria. Concerning the other 4%, significant discrepancies between the leaf pattern in the leaf control sinogram and the OBD data have been shown, which might be due to differences in the leaf open time. This corresponds also to a higher sensitivity of this method over the Delta4, adding the possibility of better monitoring the treatment delivery.
引用
收藏
页码:1625 / 1639
页数:15
相关论文
共 14 条
[1]   Single-Arc IMRT? [J].
Bortfeld, Thomas ;
Webb, Steve .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (01) :N9-N20
[2]  
Chen Q, 2011, MED PHYS, V39, P143
[3]  
ESTRO, 2008, GUID VER IMRT
[4]  
Handsfield L L, 2014, MED PHYS, V41
[5]   A feasible method for clinical delivery verification and dose reconstruction in tomotherapy [J].
Kapatoes, JM ;
Olivera, GH ;
Ruchala, KJ ;
Smilowitz, JB ;
Reckwerdt, PJ ;
Mackie, TR .
MEDICAL PHYSICS, 2001, 28 (04) :528-542
[6]   Delivery verification in sequential and helical tomotherapy [J].
Kapatoes, JM ;
Olivera, GH ;
Reckwerdt, PJ ;
Fitchard, EE ;
Schloesser, EA ;
Mackie, TR .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) :1815-1841
[7]  
Kong C W, 2012, BIOMED IMAGING INTER, V8, pe14
[8]  
Low D A, 1998, MED PHYS, V25, P656
[9]   TOMOTHERAPY - A NEW CONCEPT FOR THE DELIVERY OF DYNAMIC CONFORMAL RADIOTHERAPY [J].
MACKIE, TR ;
HOLMES, T ;
SWERDLOFF, S ;
RECKWERDT, P ;
DEASY, JO ;
YANG, J ;
PALIWAL, B ;
KINSELLA, T .
MEDICAL PHYSICS, 1993, 20 (06) :1709-1719
[10]  
Mohammadi M, 2012, J MED PHYS, V37, P200