Keratometry versus total corneal refractive power: Analysis of measurement repeatability with 5 different devices in normal eyes with low astigmatism

被引:29
作者
Fityo, Sylvia [1 ]
Buehren, Jens [1 ]
Shajari, Mehdi [1 ]
Kohnen, Thomas [1 ]
机构
[1] Goethe Univ Frankfurt, Dept Ophthalmol, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
关键词
INTRAOCULAR-LENS POWER; ROTATING SCHEIMPFLUG CAMERA; CATARACT-SURGERY; TOPOGRAPHY;
D O I
10.1016/j.jcrs.2015.11.046
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To analyze the repeatability of corneal astigmatism measurements with 2 autokeratometers (IOLMaster 500 and Lenstar LS 900), 2 Placido topographers (Keratron Scout and Atlas 9000), and a Scheimpflug tomographer (Pentacam HR). SETTING: Department of Ophthalmology, Goethe University, Frankfurt, Germany. DESIGN: Prospective case series. METHODS: The inclusion criteria were a minimum age of 18 years, corneal astigmatism of 3.00 diopters or less, and patient's consent to participate in the study. The main exclusion criteria were corneal pathologies, trauma, previous surgery, hard contact lens wear in the previous 3 months, and soft contact lens wear in the previous 6 weeks. In addition to manifest refraction and visual acuity measurements, all subjects had 2 full measurements of 1 eye using each device. Anterior corneal astigmatism and total corneal refractive power via calculations derived from the Scheimpflug tomographer were evaluated. Statistical analysis was performed using the Bland-wAltman method and multiple linear regression analysis. RESULTS: Forty-five eyes of 45 subjects (mean age 53.11 years +/- 10.43 [SD]; 23 left eyes; 33 women) were included. The mean intermeasurement differences were approximately zero for all devices. The relative coefficient of repeatability of anterior corneal astigmatism, steep axis, and the vector analysis parameter J180 were lowest with the Scheimpflug tomographer. Total corneal refractive power measured with the Scheimpflug tomographer showed the highest repeatability in the larger diameter zones. CONCLUSION: The Scheimpflug tomographer yielded higher repeatability of corneal astigmatism measurements, with total corneal refractive power measurements being as repeatable as established metrics. (C) 2016 ASCRS and ESCRS
引用
收藏
页码:569 / 576
页数:8
相关论文
共 22 条
  • [1] A NEW METHOD OF ANALYZING VECTORS FOR CHANGES IN ASTIGMATISM
    ALPINS, NA
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1993, 19 (04) : 524 - 533
  • [2] Scheimpflug and topography systems in ophthalmologic diagnostics
    Auffarth G.U.
    Borkensein A.F.M.
    Ehmer A.
    Mannsfeld A.
    Rabsilber T.M.
    Holzer M.P.
    [J]. Der Ophthalmologe, 2008, 105 (9): : 810 - 817
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] Intra- and Interindividual Comparison of Corneal Refraction Measured by IOL-Master vs. Corneal Topography
    Cucera, A.
    Lang, G. K.
    Buchwald, H. -J.
    [J]. KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2008, 225 (11) : 957 - 962
  • [5] Prevalence of corneal astigmatism before cataract surgery
    Ferrer-Blasco, Teresa
    Montes-Mico, Robert
    Peixoto-de-Matos, Sofia C.
    Gonzalez-Meijome, Jose M.
    Cervino, Alejandro
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) : 70 - 75
  • [6] Origins of the Keratometer and its Evolving Role in Ophthalmology
    Gutmark, Ron
    Guyton, David L.
    [J]. SURVEY OF OPHTHALMOLOGY, 2010, 55 (05) : 481 - 497
  • [7] The Prevalence of Astigmatism and its Determinants in a Rural Population of Iran: The "Nooravaran Salamat" Mobile Eye Clinic Experience
    Hashemi, Hassan
    Rezvan, Farhad
    Yekta, Abbas Ali
    Hashemi, Maryam
    Norouzirad, Reza
    Khabazkhoob, Mehdi
    [J]. MIDDLE EAST AFRICAN JOURNAL OF OPHTHALMOLOGY, 2014, 21 (02) : 175 - 181
  • [8] Herrmann C, 2008, OPHTHALMOLOGE, V105, P193, DOI 10.1007/s00347-008-1696-0
  • [9] Precision of a new device for biometric measurements in pseudophakic eyes
    Hildebrandt, A. L.
    Auffarth, G. U.
    Holzer, M. P.
    [J]. OPHTHALMOLOGE, 2011, 108 (08): : 739 - 744
  • [10] Validity of the keratometric index: Evaluation by the Pentacam rotating Scheimpflug camera
    Ho, Jau-Der
    Tsai, Ching-Yao
    Tsai, Ray Jui-Fang
    Kuo, Li-Lin
    Tsai, I-Lun
    Liou, Shiow-Wen
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (01) : 137 - 145