Confirmatory analysis of acetylgestagens in plasma using liquid chromatography-tandem mass spectrometry

被引:30
作者
Mortensen, Sarah Kelly [1 ]
Pedersen, Mikael [1 ]
机构
[1] Danish Inst Food & Vet Res, DK-2860 Soborg, Denmark
关键词
gestagens; liquid chromatography-tandem mass spectrometry (LC-MS/MS); monitoring; plasma; residues; steroids;
D O I
10.1016/j.aca.2006.10.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A confirmatory method has been developed and validated for the determination of chlormadinone acetate (CMA), megestrol acetate (NIGA), melengestrol acetate (MLA) and medroxyprogesterone acetate (MPA) in bovine and porcine plasma. Analytes are extracted from plasma samples using matrix-assisted liquid-liquid extraction (LLE) on Extrelut NT columns followed by C18 solid-phase extraction (SPE). Analytes were analysed using liquid chromatography-tandem mass spectrometry (I-C-MS/MS), and quantification was performed using matrix-matched calibration standards in combination with deuterated internal standards. In accordance with Commission Decision 2002/657/EC, two ion transitions were monitored for each analyte. Decision limits (CC alpha) were estimated by analysing 20 blank plasma samples and ranged from 0.1 to 0.2 ng mL(-1). Detection capabilities (CC beta) were estimated using 20 plasma samples fortified at 0.5 ng mL(-1) and were < 0.5 ng mL(-1). In the range 0.5-2 ng mL(-1), the mean intra-laboratory reproducibility of the analytes ranged from 6 to 18% (%R.S.D.). Analytes were shown to be stable in fortified plasma samples for > 8 months when stored at -20 degrees C. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:217 / 222
页数:6
相关论文
共 17 条
[1]   Exposure to exogenous estrogens in food: possible impact on human development and health [J].
Andersson, AM ;
Skakkebaek, NE .
EUROPEAN JOURNAL OF ENDOCRINOLOGY, 1999, 140 (06) :477-485
[2]  
[Anonymous], 2002, Off. J. Eur. Comm, VL221, P8
[3]  
[Anonymous], 1996, Official Journal of the European Communities, VL125, P3
[4]   Detection of melengestrol acetate residues in plasma and edible tissues of heifers [J].
Daxenberger, A ;
Meyer, K ;
Hageleit, M ;
Meyer, HHD .
VETERINARY QUARTERLY, 1999, 21 (04) :154-158
[5]   High-performance liquid chromatography-tandem mass spectrometry validation of medroxyprogesterone acetate in products of pork origin and serum [J].
Giannetti, L ;
Barchi, D ;
Fiorucci, F ;
Gennuso, E ;
Sanna, P ;
Pallagrosi, M ;
Neri, B .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2005, 43 (07) :333-336
[6]  
Hageleit M, 2001, FOOD ADDIT CONTAM, V18, P285, DOI 10.1080/02652030010024311
[7]   DETERMINATION OF ANABOLIC ESTERS IN OILY FORMULATIONS AND PLASMA IN HUSBANDRY USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND GAS-CHROMATOGRAPHY - MASS-SELECTIVE DETECTION [J].
HOOIJERINK, D ;
SCHILT, R ;
VANBENNEKOM, E ;
BROUWER, B .
ANALYST, 1994, 119 (12) :2617-2622
[8]  
HOOIJERINK H, 2003, ANAL CHIM ACTA, V483, P269
[9]   Liquid chromatography-tandem mass spectrometry of some anabolic steroids [J].
Joos, PE ;
Van Ryckeghem, M .
ANALYTICAL CHEMISTRY, 1999, 71 (20) :4701-4710
[10]  
Kim SM, 2001, RAPID COMMUN MASS SP, V15, P2041