The Effects of Plasma Prothrombine Time and Activated Partial Thromboplastin Time Based on Different Instruments and Methods

被引:0
作者
Hernaningsih, Yetti [1 ]
Butarbutar, Trieva Verawaty [2 ]
机构
[1] Univ Airlangga, Fac Med, Dept Clin Pathol, Surabaya, East Java, Indonesia
[2] Dr Doris Sylvanus Gen Hosp, Lab Clin Pathol, Clin Pathol, Palangka Raya, Central Kaliman, Indonesia
关键词
CoaDATA; 501; Sysmex CS2100i; STA compact; Optical method; Electromechanical method; MECHANICAL CLOT DETECTION; COAGULATION; VALIDATION; ANALYZER;
D O I
10.7860/JCDR/2019/37999.13172
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Many laboratories have more than one coagulometer with different types and methods that are used simultaneously. In patients requiring Plasma Prothrombine Time (PPT) and Activated Partial Thromboplastin Time (APTT), follow-up treatments may be performed on different instruments. Aim: To analyse whether there are differences in the value of PPT and APTT between 3 instruments with different methods. Materials and Methods: This study took 64 samples of patients with End Stage Renal Disease (ESRD) who underwent haemodialysis in Haemodialysis Centre of Dr. Soetomo General Hospital Surabaya between June 2017 and October 2017. All samples were carried out by examination of PPT and APTT on three instruments: CoaDATA 501 with one wavelength optical detection, Sysmex CS2100i with three wavelengths optical detection, and STA Compact using electromechanical methods. Results: Significantly different results were obtained in the PPT comparison between the group of CoaData with Sysmex CS2100i (p <= 0.001) and CoaData with STA Compact (p <= 0.001). While in the intergroup APTT, all gave significantly different test results (p<0.001). Conclusion: Because of difference value of PPT and APTT in difference instrument and method, therefore serial tests should be performed in the same instrument to obtain valid and reliable results.
引用
收藏
页码:EC21 / EC25
页数:5
相关论文
共 29 条
[1]  
Adcock DM, 2005, APPROVED GUIDELINE, V25, P23
[2]   Discrepancy in Optical & Mechanical Method in Coagulation Tests in a Turbid Sample [J].
Aggarwal, Swati ;
Nayak, Deepak M. ;
Manohar, Chethan .
INDIAN JOURNAL OF HEMATOLOGY AND BLOOD TRANSFUSION, 2014, 30 :S402-S404
[3]  
[Anonymous], COADATA 501 MANUAL O, P1
[4]  
[Anonymous], CLSIH57A
[5]   Comparison of optical and mechanical clot detection for routine coagulation testing in a large volume clinical laboratory [J].
Bai, Bing ;
Christie, Douglas J. ;
Gorman, Robert T. ;
Wu, Jogin R. .
BLOOD COAGULATION & FIBRINOLYSIS, 2008, 19 (06) :569-576
[6]   Biological Variation in Tests of Hemostasis (vol 34, pg 635, 2008) [J].
Banfi, Giuseppe ;
Del Fabbro, Massimo .
SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2009, 35 (01) :119-126
[7]  
Clinical and Laboratory Standards Institute (CLSI), 2003, H03A5 CLSI
[8]  
Crighton Gemma, 2013, Methods Mol Biol, V992, P73, DOI 10.1007/978-1-62703-339-8_5
[9]  
Diagnostica Stago, 2013, MAT SAF DAT SHEET ST
[10]  
Diagnostica Stago, 2013, MAT SAF DAT SHEET CA