Electrical Analysis of a Planar-type Digital X-ray Detector with Inert Gas

被引:5
作者
Yun, Min-seok [1 ]
Nam, Sang-hee [1 ]
Park, Sung-Kwang [2 ]
Jung, Suk-hee [3 ]
机构
[1] Inje Univ, Dept Biomed Engn, Gimhae 621749, South Korea
[2] Busan Paik Hosp, Pusan 614735, South Korea
[3] Infinite Inc, Seoul 110300, South Korea
关键词
Digital radiography; Gas detector; X-ray; Radiation detector; IMAGING PERFORMANCE; MERCURIC IODIDE; FILMS; SENSITIVITY; RADIOGRAPHY;
D O I
10.3938/jkps.56.1083
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Large-area digital X-ray image detectors are currently being investigated for medical diagnostic and non-destructive tests. In this paper, a new plat in detector for diagnostic X-ray (using an inert gas) imaging is proposed, and its electrical characteristics are presented. Almost gas detectors are designed in the shape of a chamber, so we made a planar-type panel sample based on a plasma display panel (PDP) technique. The transparent electrodes and dielectric layer were formed on the front glass. Additionally, the bottom electrodes and dielectric layer were formed on the rear glass. Then, we injected an inert gas (xenon and helium) with various component; ratios into the sample using by gas-filling equipment. We measured the dark current density and the X-ray sensitivity as functions of the electric field to investigate the electrical properties. From the results, a stabilized dark current density and a significant X-ray sensitivity were obtained. Also, the signal linearity was calculated, and a good linearity as a function of exposure dose could be realized over a wide diagnostic energy range. All linearity coefficient gamma values were higher than 0.9. These results mean that the passive matrix-addressed flat panel gas detector can be used as a digital x-ray detector.
引用
收藏
页码:1083 / 1087
页数:5
相关论文
共 24 条
[1]   Low-voltage characteristics of MgO-CaO films as a protective layer for AC plasma display panels by e-beam evaporation [J].
Cho, JH ;
Kim, R ;
Lee, KW ;
Son, CY ;
Yeom, GY ;
Kim, HJ ;
Kim, JY ;
Park, JW .
JOURNAL OF MATERIALS SCIENCE, 1999, 34 (20) :5055-5059
[2]   Effect of annealing on the X-ray detection properties of nano-sized polycrystalline lead oxide films [J].
Cho, Sungho ;
Kang, Sansik ;
Kwon, Chul ;
Yun, Min-Seok ;
Park, Jikoon ;
Park, Sungkwang ;
Choi, Jang-Yong ;
Mun, Chiwoong ;
Lee, Hyungwon ;
Nam, Sanghee .
JAPANESE JOURNAL OF APPLIED PHYSICS, 2008, 47 (09) :7393-7397
[3]   Defects in CdTe polycrystalline films grown by physical vapour deposition [J].
Corregidor, V ;
Saucedo, E ;
Fornaro, L ;
Castaño, JL ;
Dieguez, E .
MATERIALS SCIENCE AND ENGINEERING B-SOLID STATE MATERIALS FOR ADVANCED TECHNOLOGY, 2002, 91 :525-528
[4]   Development of gas microstrip detectors for digital X-ray imaging and radiation dosimetry [J].
Dixit, MS ;
Armitage, JC ;
Dubeau, J ;
Gobbi, DG ;
Johns, PC ;
Karlen, D ;
Oakham, FG ;
Waker, AJ .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1998, 47 (03) :809-813
[5]   Low dark current (0 0 l) mercuric iodide thick films for X-ray direct and digital imagers [J].
Fornaro, L ;
Cuña, A ;
Noguera, A ;
Aguiar, I ;
Pérez, M ;
Mussio, L ;
Gancharov, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (06) :3107-3110
[6]   Comparison between sublimation and evaporation as process for growing lead iodide polycrystalline films [J].
Fornaro, L ;
Saucedo, E ;
Mussio, L ;
Gancharov, A ;
Guimaraes, F ;
Hernandes, AC .
HARD X-RAY AND GAMMA-RAY DETECTOR PHYSICS III, 2001, 4507 :99-107
[7]   Toward epitaxial lead-iodide films for X-ray digital imaging [J].
Fornaro, L ;
Saucedo, E ;
Mussio, L ;
Gancharov, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) :2274-2278
[8]   System considerations in CCD-based x-ray imaging for digital chest radiography and digital mammography [J].
Hejazi, S ;
Trauernicht, DP .
MEDICAL PHYSICS, 1997, 24 (02) :287-297
[9]   Intrinsic spatial resolution of semiconductor X-ray detectors: a simulation study [J].
Hoheisel, M ;
Giersch, J ;
Bernhardt, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 531 (1-2) :75-81
[10]   Imaging performance of amorphous selenium flat-panel detector for digital fluoroscopy [J].
Hunt, DC ;
Tousignanth, O ;
Demers, Y ;
Laperrière, L ;
Rowlands, JA .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :226-234