Stapes prosthesis attachment: The effect of crimping on sound transfer in otosclerosis surgery

被引:69
作者
Huber, AM
Ma, FR
Felix, H
Linder, T
机构
[1] Univ Zurich Hosp, Dept Otorhinolaryngol Head & Neck Surg, CH-8091 Zurich, Switzerland
[2] Beijing Med Univ, Third Teaching Hosp & Clin Med Coll, Dept Otolaryngol, Beijing 100083, Peoples R China
[3] Kantonsspit Luzern, Dept Otorhinolaryngol Head & Neck Surg, Luzern, Switzerland
关键词
stapes surgery; stapedotomy; otosclerosis; prosthesis; laser Doppler;
D O I
10.1097/00005537-200305000-00015
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objectives/Hypothesis: Although in stapes surgery successful hearing improvement may be achieved in the majority of patients, unsatisfactory closure of the air-bone gap can be recorded. One of many reasons for unexpected failures of stapes surgery may be the insufficient crimping of a stapes prosthesis onto the incus. The objectives of the study were to assess the amount of sound transmission loss in response to the quality of prosthesis crimping and to identify a required loop attachment pattern to obtain good sound transmission results. Study Design: Experimental. Methods A temporal bone model was developed to measure the sound transmission properties between incus and prosthesis on 17 fresh human temporal bones. The attachment of a titanium stapes piston was assessed without crimping, followed by loose crimping and tight fixation to the incus, using scanning laser Doppler interferometry, endoscopic photography, micro grinding technique, and scanning electron microscopy. An algorithm had to be developed to simulate acoustical stimulation using electromechanical stimulation. Results: Optimal tight crimping of the stapes piston revealed consistent good sound transfer function ranging from 0 to 7 dB loss, and loss remained, on average, at 2 dB. The mean transmission losses for conditions of loose crimping and no crimping were surprisingly small (within 10 dB). However, these unusual crimping conditions allowed a wide range of losses up to 28 dB. A close coupling at least at two opposite points was obligatory to obtain consistently good results. Conclusions: Perfect hearing reconstruction necessitates ideal crimping of a prosthesis to obtain consistently good results. However, the final functional gain depends on many different intraoperative and postoperative factors.
引用
收藏
页码:853 / 858
页数:6
相关论文
共 16 条
  • [1] Asai M, 1999, ACTA OTO-LARYNGOL, V119, P356
  • [2] BIRCH L, 1986, J LARYNGOL OTOL, V100, P1
  • [3] Malleostapedotomy in revision surgery for otosclerosis
    Fisch, U
    Acar, GÖ
    Huber, AM
    [J]. OTOLOGY & NEUROTOLOGY, 2001, 22 (06) : 776 - 785
  • [4] Fisch U., 1994, Tympanoplasty, mastoidectomy and stapes surgery, V1
  • [5] MEASUREMENT OF THE OSSICULAR VIBRATION RATIO IN HUMAN TEMPORAL BONES BY USE OF A VIDEO MEASURING SYSTEM
    GYO, K
    ARITOMO, H
    GOODE, RL
    [J]. ACTA OTO-LARYNGOLOGICA, 1987, 103 (1-2) : 87 - 95
  • [6] HOUSE JW, 1993, OTOLARYNG CLIN N AM, V26, P389
  • [7] Intraoperative assessment of stapes movement
    Huber, A
    Linder, T
    Ferrazzini, M
    Schmid, S
    Dillier, N
    Stoeckli, S
    Fisch, U
    [J]. ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2001, 110 (01) : 31 - 35
  • [8] Evaluation of eardrum laser Doppler interferometry as a diagnostic tool
    Huber, AM
    Schwab, C
    Linder, T
    Stoeckli, SJ
    Ferrazzini, M
    Dillier, N
    Fisch, U
    [J]. LARYNGOSCOPE, 2001, 111 (03) : 501 - 507
  • [9] SOUND PRESSURE GAIN PRODUCED BY THE HUMAN MIDDLE-EAR
    KUROKAWA, H
    GOODE, RL
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 1995, 113 (04) : 349 - 355
  • [10] Stapes surgery: How precisely do different prostheses attach to the long process of the incus with different instruments and different surgeons?
    Kwok, P
    Fisch, U
    Strutz, J
    May, J
    [J]. OTOLOGY & NEUROTOLOGY, 2002, 23 (03) : 289 - 295