Dose profile evaluation in digital breast tomosynthesis exposition using radiochromic film

被引:3
作者
Flores, Mabel B. [1 ]
Mourao, Arnaldo P. [1 ,2 ]
Oliveira, Fernando A. [2 ]
del Rio, Margarita C. [3 ]
机构
[1] Univ Fed Minas Gerais, Dept Engn Nucl, Belo Horizonte, MG, Brazil
[2] Ctr Fed Educ Tecnol Minas Gerais, Ctr Engn Biomed, Belo Horizonte, MG, Brazil
[3] Univ Complutense Madrid, Dept Radiol Rehabi & Fisioterapia, Madrid, Spain
关键词
Dosimetry; Mammography; DBT; Radiochromic film; MAMMOGRAPHY; CANCER; PHANTOMS; WOMEN;
D O I
10.1016/j.apradiso.2019.07.002
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Digital Breast Tomosynthesis (DBT) is an efficient method to diagnose changes in the breast tissues. However, it may promote some future detriment to the patient exposed in this test. The future effects can be evaluating using absorbed dose values. A mammographic phantom should be chosen to simulate approximately the properties of a compressed breast in terms of the primary and dispersed radiation spectrum that exits its lower surface. The use of PMMA has important advantages. It is cheaper than substitutes for specially manufactured fabrics and is more available. The objective of this work will be to estimate the dose profile that a real breast would receive during a DTB examination. For this purpose a breast phantom with acrylic materials (PMMA) and dimensions similar to those of a real breast will be constructed, and the dose will be measured on the surface (ESAK profile) and in the middle cut (air kerma profile) of the breast phantom using the radiochromic film.
引用
收藏
页码:140 / 144
页数:5
相关论文
共 26 条
  • [1] Phantoms for quality control procedures in digital breast tomosynthesis: dose assessment
    Bouwman, R. W.
    Diaz, O.
    van Engen, R. E.
    Young, K. C.
    den Heeten, G. J.
    Broeders, M. J. M.
    Veldkamp, W. J. H.
    Dance, D. R.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (13) : 4423 - 4438
  • [2] Brasil Instituto Nacional de Cancer Jose Alencar Gomes da Silva, 2019, EST 2018 INC CANC BR
  • [3] Assessing Improvement in Detection of Breast Cancer with Three-dimensional Automated Breast US in Women with Dense Breast Tissue: The Somoinsight Study
    Brem, Rachel F.
    Tabar, Laszlo
    Duffy, Stephen W.
    Inciardi, Marc F.
    Guingrich, Jessica A.
    Hashimoto, Beverly E.
    Lander, Marla R.
    Lapidus, Robert L.
    Peterson, Mary Kay
    Rapelyea, Jocelyn A.
    Roux, Susan
    Schilling, Kathy J.
    Shah, Biren A.
    Torrente, Jessica
    Wynn, Ralph T.
    Miller, Dave P.
    [J]. RADIOLOGY, 2015, 274 (03) : 663 - 673
  • [4] Development of a physical 3D anthropomorphic breast phantom
    Carton, Ann-Katherine
    Bakic, Predrag
    Ullberg, Christer
    Derand, Helen
    Maidment, Andrew D. A.
    [J]. MEDICAL PHYSICS, 2011, 38 (02) : 891 - 896
  • [5] Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol
    Dance, DR
    Skinner, CL
    Young, KC
    Beckett, JR
    Kotre, CJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) : 3225 - 3240
  • [6] Energy dependent calibration of XR-QA2 radiochromic film with monochromatic and polychromatic x-ray beams
    Di Lillo, F.
    Mettivier, G.
    Sarno, A.
    Tromba, G.
    Tomic, N.
    Devic, S.
    Russo, P.
    [J]. MEDICAL PHYSICS, 2016, 43 (01) : 583 - 588
  • [7] Internal breast dosimetry in mammography: Monte Carlo validation in homogeneous and anthropomorphic breast phantoms with a clinical mammography system
    Fedon, Christian
    Caballo, Marco
    Sechopoulos, Ioannis
    [J]. MEDICAL PHYSICS, 2018, 45 (08) : 3950 - 3961
  • [8] Ferlay J., 2021, Global Cancer Observatory: Cancer Today
  • [9] Cancer incidence and mortality worldwide: Sources, methods and major patterns in GLOBOCAN 2012
    Ferlay, Jacques
    Soerjomataram, Isabelle
    Dikshit, Rajesh
    Eser, Sultan
    Mathers, Colin
    Rebelo, Marise
    Parkin, Donald Maxwell
    Forman, David
    Bray, Freddie
    [J]. INTERNATIONAL JOURNAL OF CANCER, 2015, 136 (05) : E359 - E386
  • [10] Response of CR Detector Plates for Use in Mammography Equipment
    Flores, Mabel Bustos
    Mourao, Arnaldo Prata
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 1, 2019, 68 (01): : 117 - 122