Comparison of single- and dual-coil active pectoral defibrillation lead systems

被引:65
|
作者
Gold, MR [1 ]
Olsovsky, MR [1 ]
Pelini, MA [1 ]
Peters, RW [1 ]
Shorofsky, SR [1 ]
机构
[1] Univ Maryland, Med Syst, Sch Med, Dept Med,Div Cardiol, Baltimore, MD 21201 USA
关键词
D O I
10.1016/S0735-1097(98)00103-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. The purpose of this study was to compare defibrillation thresholds with lead systems consisting of an active left pectoral electrode and either single or dual transvenous coils. Background. Lead systems that include an active pectoral pulse generator reduce defibrillation thresholds and permit transvenous defibrillation in nearly all patients. A further improvement in defibrillation efficacy is desirable to allow for smaller pulse generators with a reduced maximal output. Methods. This prospective study was performed in 50 consecutive patients. Each patient was evaluated with two lead configurations with the order of testing randomized. Shocks were delivered between the right ventricular coil and either an active can alone (single coil) or an active can with the proximal atrial coil (dual coil). The right ventricular coil was the cathode for the first phase of the biphasic defibrillation waveform. Results. Delivered energy at the defibrillation threshold was 10.1 +/- 5.0 J for the single-coil configuration and 8.7 +/- 4.0 J for the dual-coil configuration (p < 0.02). Moreover, 98% of patients had low (less than or equal to 15 J) thresholds with the dual-coil lead system, compared with 88% of patients with the single-coil configuration (p = 0.05). Leading edge voltage (p < 0.001) and shock impedance (p < 0.001) were also decreased with the dual coil configuration, although peak current was increased (p < 0.001). Conclusions. A dual-coil, active pectoral lead system reduces defibrillation energy requirements compared with a single-coil, unipolar configuration. (C) 1998 by the American College of Cardiology.
引用
收藏
页码:1391 / 1394
页数:4
相关论文
共 50 条
  • [1] Optimization of atrial defibrillation with a dual-coil, active pectoral lead system
    Rashba, EJ
    Shorofsky, SR
    Peters, RW
    Gold, MR
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (07) : 790 - 794
  • [2] Clinical predictors of atrial defibrillation thresholds with a dual-coil, active pectoral lead system
    Rashba, EJ
    Shorofsky, SR
    Brown, T
    Peters, RW
    Gold, MR
    HEART RHYTHM, 2005, 2 (01) : 49 - 54
  • [3] Dual-coil vs single-coil active pectoral implantable defibrillator lead systems: defibrillation energy requirements and probability of defibrillation success at multiples of the defibrillation energy requirements
    Schulte, B
    Sperzel, J
    Carlsson, J
    Schwarz, T
    Ehrlich, W
    Pitschner, HF
    Neuzner, J
    EUROPACE, 2001, 3 (03): : 177 - 180
  • [4] REDUCTION OF DEFIBRILLATION THRESHOLDS WITH AN ACTIVE PECTORAL PULSE-GENERATOR AND A DUAL COIL LEAD
    GOLD, MR
    FOSTER, AH
    SHOROFSKY, SR
    CIRCULATION, 1995, 92 (08) : 1611 - 1611
  • [5] Effect of an active abdominal pulse generator on defibrillation thresholds with a dual-coil, transvenous ICD lead system
    Rashba, Eric J.
    Farasat, Morteza
    Kirk, Malcolm M.
    Shorofsky, Stephen R.
    Peters, Robert W.
    Gold, Michael R.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (06) : 617 - 620
  • [6] Clinical outcomes of patient with single-coil vs. dual-coil implantable cardioverter defibrillation lead for secondary prevention
    Korai, K.
    Ishibashi, K.
    Kamakura, T.
    Wada, M.
    Yamagata, K.
    Inoue, Y.
    Miyamoto, K.
    Nagase, S.
    Noda, T.
    Aiba, T.
    Noguchi, T.
    Yasuda, S.
    Kusano, K.
    EUROPEAN HEART JOURNAL, 2018, 39 : 1226 - 1226
  • [7] Failure Rates of Single- Versus Dual-Coil Nonrecalled Sprint Quattro Defibrillator Leads
    Shariff, Nasir
    Alluri, Krishna
    Saba, Samir
    AMERICAN JOURNAL OF CARDIOLOGY, 2015, 115 (02): : 202 - 205
  • [8] Comparison of single-coil lead versus dual-coil lead of implantable cardioverter defibrillator on lead-related venous complications in a canine model
    Yu, Ziqing
    Wu, Yuan
    Qin, Shengmei
    Wang, Jingfeng
    Chen, Xueying
    Chen, Ruizhen
    Su, Yangang
    Ge, Junbo
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2018, 52 (02) : 195 - 201
  • [9] Comparison of single-coil lead versus dual-coil lead of implantable cardioverter defibrillator on lead-related venous complications in a canine model
    Ziqing Yu
    Yuan Wu
    Shengmei Qin
    Jingfeng Wang
    Xueying Chen
    Ruizhen Chen
    Yangang Su
    Junbo Ge
    Journal of Interventional Cardiac Electrophysiology, 2018, 52 : 195 - 201
  • [10] Predictors of High Defibrillation Threshold in Patients With Implantable Cardioverter-Defibillator Using a Transvenous Dual-Coil Lead
    Mizukami, Kazuya
    Yokoshiki, Hisashi
    Mitsuyama, Hirofumi
    Watanabe, Masaya
    Tenma, Taro
    Matsui, Yoshiro
    Tsutsui, Hiroyuki
    CIRCULATION JOURNAL, 2015, 79 (01) : 77 - 84