Evaluation of a dual chamber implantable cardioverter defibrillator for the treatment of atrial and ventricular arrhythmias

被引:12
|
作者
Boriani, G
Wollmann, C
Biffi, M
Kühl, M
Schuchert, A
Sperzel, J
Stiller, S
Gasparini, G
Böcker, D
机构
[1] Univ Bologna, Inst Cardiol, Policlin S Orsola, I-40138 Bologna, Italy
[2] Univ Munster, Dept Cardiol, D-4400 Munster, Germany
[3] Guidant Inc, St Paul, MN USA
[4] Univ Hamburg, Hamburg, Germany
[5] Univ Ulm, D-89069 Ulm, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2003年 / 26卷 / 01期
关键词
atrial tachyarrhythmia; atrialftbrillation; implantable cardioverter defibrillator; detection time; cardioversion; defibrillation;
D O I
10.1046/j.1460-9592.2003.00072.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Eighty-nine patients with a documented history of atrial tachyarrhythmias or fibrillation (AF) received a cardioverter defibrillator designed to selectively differentiate and treat atrial and ventricular arrhythmias. Twenty-two patients received a coronary sinus lead and, therefore, could use a separate shock vector for selective treatment of atrial tachyarrhythmias/AE The device is designed to treat tachyarrhythmias with antitachycardia pacing (ATP) and/or shock therapy using an atrial and/or a ventricular shock vector. Patients underwent induction and shock termination of atrial or dual tachyarrhythmias (AF/VF) to verify proper device function and to measure the arrhythmia detection time with enhancements and preventive algorithms programmed On and Off, respectively. Detection time for 329 VF inductions was 2.41 +/- 0.64 seconds with enhancements On and 2.29 +/- 0.47 with enhancements Off (NS). At implant or predischarge, 283 AF and/or AF/VF (121 atrial and 162 atrial/ventricular fibrillation) were induced. Shock conversion efficacy was 89.8% with AF conversion energies ranging from 0.9 to 27 J. Thirteen of the 23 patients had atrial shock conversions using the separate shock vector with an average conversion energy of 1.9 +/- 1.41. (range 0.5-5 J) A During follow-up the efficacy of ATP on atrial tachyarrhythmias was 59% and the efficacy of delivered shocks on AF was 85%. This new dual chamber cardioverter defibrillator appropriately detected and classified atrial arrhythmias, and shock therapy for AF was highly effective. The detection algorithm differentiated atrial tachyarrhythmia/AF and did not delay VF detection. The separate shock vector converted induced AF with energies ranging from 0.6 to 5 J.
引用
收藏
页码:461 / 465
页数:5
相关论文
共 50 条
  • [1] Diagnosis and therapy of atrial tachyarrhythmias in the dual chamber implantable cardioverter defibrillator
    Dijkman, B
    Wellens, HJJ
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (11) : 1196 - 1205
  • [2] Use of atrial and ventricular electrograms from a dual chamber implantable cardioverter defibrillator to elucidate a complex dysrhythmia
    Greenberg, RM
    Degeratu, FT
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10): : 2002 - 2004
  • [3] Implantable Dual-Chamber Defibrillator for the Selective Treatment of Spontaneous Atrial and Ventricular Arrhythmias: Arrhythmia Incidence and Device Performance
    Andreas Schuchert
    Giuseppe Boriani
    Christian Wollmann
    Mauro Biffi
    Martin Kühl
    Johannes Sperzel
    Sascha Stiller
    Gianni Gasparini
    Dirk Böcker
    Journal of Interventional Cardiac Electrophysiology, 2005, 12 : 149 - 156
  • [4] Implantable dual-chamber defibrillator for the selective treatment of spontaneous atrial and ventricular arrhythmias:: Arrhythmia incidence and device performance
    Schuchert, A
    Boriani, G
    Wollmann, C
    Biffi, M
    Kühl, M
    Sperzel, J
    Stiller, S
    Gasparini, G
    Böcker, D
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2005, 12 (02) : 149 - 156
  • [5] Clinical experience with a dual-chamber implantable cardioverter defibrillator to treat atrial tachyarrhythmias
    Gold, MR
    Sulke, N
    Schwartzman, DS
    Mehra, R
    Euler, DE
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2001, 12 (11) : 1247 - 1253
  • [6] Intradevice interaction in a dual chamber implantable cardioverter defibrillator preventing ventricular tachyarrhythmia detection
    Shivkumar, K
    Feliciano, Z
    Boyle, NG
    Wiener, I
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (11) : 1285 - 1288
  • [7] Implantable dual-chamber cardioverter-defibrillator-pacemaker
    Pfeiffer D.
    Mende M.
    Hagendorff A.
    Current Cardiology Reports, 2000, 2 (4) : 335 - 340
  • [8] Atrial oversensing in a dual chamber implantable cardioverter-defibrillator due to electromechanical lead interaction
    Cohen, TJ
    Ibrahim, B
    Quan, WL
    HEARTWEB, 1997, 3 (01): : U6 - U10
  • [9] The Effect of Air Pollution on Ventricular Arrhythmias in Patients with Implantable Cardioverter Defibrillator
    Shahabi, Javad
    Azhari, Amirhossein
    Safari, Mohsen
    Rahimi-Pordanjani, Amin
    RESEARCH IN CARDIOVASCULAR MEDICINE, 2023, 12 (04) : 108 - 111
  • [10] Anxiety and Risk of Ventricular Arrhythmias or Mortality in Patients With an Implantable Cardioverter Defibrillator
    Habibovic, Mirela
    Pedersen, Susanne S.
    van den Broek, Krista C.
    Theuns, Dominic A. M. J.
    Jordaens, Luc
    van der Voort, Pepijn H.
    Alings, Marco
    Denollet, Johan
    PSYCHOSOMATIC MEDICINE, 2013, 75 (01): : 36 - 41