Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part II: in vivo investigation

被引:88
作者
Khan, M
Smith, N
Jones, E
Finch, DS
Cameron, RE
机构
[1] Univ Cambridge, Dept Mat Sci & Met, Cambridge CB2 3QZ, England
[2] Stryker Howmedica Osteon, Staines, England
[3] Stryker Europe, CASA Grp, Limerick, Ireland
[4] Univ Colorado, Dept Mech Engn, Boulder, CO 80309 USA
基金
英国工程与自然科学研究理事会;
关键词
hip replacement prostheses; polyurethane; elstomer; compliance; biocompatibility; animal model;
D O I
10.1016/j.biomaterials.2004.02.064
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The polyurethane (PU) elastomer Corethane 80A (Corvita) is being considered as the acetabular bearing material in a novel total replacement hip joint. Its biostability was investigated in vitro (Analysis and evaluation of a biomedical polycarbonate urethane tested in an in vitro study and an ovine arthroplasty model. Part 1: material selection and evaluation, Biomaterials, in press) together with three other commercially available biomedical PUs: Pellethane 2363-80A (DOW Chemical), a polyhexamethylene oxide based PU, PHMO-PU (CSIRO, not supplied as a commercial product) and ChronoFlex AL-80A (CardioTech). From the in vitro studies, Corethane 80A displayed the best overall resistance to hydrolysis, ESC, MIO and calcification, followed by ChronoFlex 80A and PHMO-PU, with Pellethane 80A being the least stable. Building on the in vitro investigation, the follow-up ill vivo study (reported here) assessed Corethane 80A as the bearing layer in a prototype compliant layer acetabular cup, in a fully functioning ovine total hip arthoplasty (THA) model. PU degradation in the retrieved cups was analysed using a range of analytical and physical-testing methods including mechanical testing, differential scanning calorimetry, Fourier transform infrared spectroscopy and environmental scanning electron microscopy. The Corethane 80A functioned well in the THA model, with the bearing surfaces of the retrieved hip cups showing no significant evidence of biodegradation or wear damage after 3 years in vivo. The findings in this study provide compelling evidence for the biostability and effectiveness of acetabular cups incorporating a Corethane 80A compliant bearing layer. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:633 / 643
页数:11
相关论文
共 55 条
  • [1] ALEXANDER JW, 1999, T 45 ANN M ORTH RES, P906
  • [2] Hip joint forces in sheep
    Bergmann, G
    Graichen, F
    Rohlmann, A
    [J]. JOURNAL OF BIOMECHANICS, 1999, 32 (08) : 769 - 777
  • [3] A COMPARISON OF HIP-JOINT FORCES IN SHEEP, DOG AND MAN
    BERGMANN, G
    SIRAKY, J
    ROHLMANN, A
    KOELBEL, R
    [J]. JOURNAL OF BIOMECHANICS, 1984, 17 (12) : 907 - &
  • [4] Brumby SA, 1998, CLIN ORTHOP RELAT R, P229
  • [5] Bruns DP, 1996, VET COMP ORTHOPAED, V9, P158
  • [6] START UP AND STEADY-STATE FRICTION OF THIN POLYURETHANE LAYERS
    CARAVIA, L
    DOWSON, D
    FISHER, J
    [J]. WEAR, 1993, 160 (01) : 191 - 197
  • [7] CARBONE A, VET J BIOMATER DRAFT
  • [8] Cooper R A, 1992, J Arthroplasty, V7, P285, DOI 10.1016/0883-5403(92)90050-Z
  • [9] Dillon J.G., 1989, INFRARED SPECTROSCOP
  • [10] DOWSON D, 1986, FLUID FILM LUBRICATI