Design aspects and prototype test of a very precise TDC system implemented for the Multigap RPC of the ALICE-TOF

被引:44
作者
Akindinov, AV
Alici, A
Anselmo, F
Antonioli, P
Basile, M
Romeo, GC
Cifarelli, L
Cindolo, F
Cosenza, F
D'Antone, I
De Caro, A
De Pasquale, S
Di Bartolomeo, A
Girard, MF
Golovine, V
Guerzoni, M
Guida, M
Hatzifotiadou, D
Kaidalov, AB
Kim, DH
Kim, DW
Kisselev, SM
Laurenti, G
Lioublev, E
Lee, K
Lee, SC
Luvisetto, ML
Margotti, A
Martemiyanov, AN
Massera, F
Meneghini, S
Michinelli, R
Nania, R
Otiougova, P
Pancaldi, G
Pesci, A
Pilastrini, R
Pinazza, O
Polozov, PA
Rizzi, M
Scapparone, E
Scioli, G
Sellitto, S
Semeria, F
Smirnitski, AV
Tchoumakov, MM
Ugolini, E
Usenko, E
Valenti, G
Voloshin, KG
机构
[1] Univ Bologna, I-40126 Bologna, Italy
[2] Inst Theoret & Expt Phys, Moscow 117259, Russia
[3] Ist Nazl Fis Nucl, I-40126 Bologna, Italy
[4] Univ Salerno, Dipartimento Fis, I-84100 Salerno, Italy
[5] Ist Nazl Fis Nucl, Salerno, Italy
[6] World Lab, Lausanne, Switzerland
[7] Kangnung Natl Univ, Dept Phys, Kangnung, South Korea
[8] Inst High Energy Phys, Protvino, Russia
关键词
resistive plate chambers; ALICE; time of flight; time resolution; Time Digital Converter;
D O I
10.1016/j.nima.2004.07.023
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The ALICE Time-of-Flight system will be a large area (150 m(2)) detector made by Multigap RPC (MRPC). T he time digitisation is based on the High Performance TDC chip (HPTDC). Tests carried out on board prototypes are discussed, emphasising the optimisation of the effective time resolution of the chip when working in its Very High Resolution Mode. Lab bench tests and test beam results show that a 20 ps resolution has been achieved. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:178 / 182
页数:5
相关论文
共 3 条
  • [1] *ALICE COLL, 200216 ALICE COLL CE
  • [2] HATZIFOTIADOU D, 2004, NUCL INSTR METH A
  • [3] MOTA M, 2000, P 2000 NUCL SCI S ME