共 4 条
In vivo alteration of Strata valve setting by vagus nerve stimulator-activating magnet
被引:6
作者:
Guilfoyle, Mathew R.
[1
]
Fernandes, H.
[1
]
Price, S.
[1
]
机构:
[1] Addenbrookes Hosp, Dept Neurosurg, Cambridge CB2 2QQ, England
关键词:
programmable valve;
Strata;
vagus nerve stimulator;
ventriculoperitoneal shunt;
D O I:
10.1080/02688690701243217
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
We report the unexpected adjustment of a Strata shunt valve by a vagus nerve stimulator-activating magnet, resulting in symptomatic under-drainage of CSF. Implantable devices using magnets for calibration and activation are becoming more prevalent within neurosurgery and the present case highlights the potential for adverse interactions.
引用
收藏
页码:41 / 42
页数:2
相关论文