Accuracy of Corneal Thickness by Swept-Source Optical Coherence Tomography and Scheimpflug Camera in Virgin and Treated Fuchs Endothelial Dystrophy

被引:11
作者
Arnalich-Montiel, Francisco [1 ]
Ortiz-Toquero, Sara [2 ]
Auladell, Clara [1 ]
Couceiro, Ana [1 ]
机构
[1] Ramon y Cajal Univ Hosp, Dept Ophthalmol, Carretera Colmenar Viejo Km 9,100, Madrid 28034, Spain
[2] Univ Valladolid, Sch Optometry, IOBA Eye Inst, Optometry Res Grp, Valladolid, Spain
关键词
Fuchs endothelial dystrophy; DSAEK; DMEK; optical coherence tomography; Scheimpflug tomography; corneal thickness; REPEATABILITY; PACHYMETRY; ANTERIOR; PENTACAM; KERATOCONUS;
D O I
10.1097/ICO.0000000000001530
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess intraobserver repeatability, intersession reproducibility, and agreement of swept-source Fourier-domain optical coherence tomography (SS-OCT) and the Scheimpflug camera in measuring corneal thickness in virgin and grafted eyes with Fuchs endothelial corneal dystrophy (FECD). Methods: Thirty-six control eyes, 35 FECD eyes, 30 FECD with corneal edema eyes, 25 Descemet stripping automated endothelial keratoplasty (DSAEK) eyes, and 29 Descemet membrane endothelial keratoplasty (DMEK) eyes were included. The apical center, pupillary center, and thinnest corneal thickness were determined in 3 consecutive images and repeated 2 weeks later. Repeatability and reproducibility coefficients, intraclass correlation coefficients, and 95% limits of agreement (LOA) between measurements were calculated. Agreement between devices was assessed using Bland- Altman analysis. Results: Corneal thickness measurements were highly reproducible and repeatable with both systems. SS-OCT showed better repeatability in all corneal locations in the normal, FECD, FECD with edema, DSAEK, and DMEK groups (coefficient of variation <= 0.60%, <= 0.36%, <= 0.43%, <= 1.09%, and <= 0.48%, respectively) than the Scheimpflug (coefficient of variation <= 1.15%, <= 0.92%, <= 1.10%, <= 1.25%, and <= 1.14%, respectively). Between-session 95% LOA for SS-OCT was less than 3% for all groups except for the FECD with edema group, being almost double using the Scheimpflug camera. Differences between instruments were statistically significant in all groups and locations (P < 0.01) except in the DSAEK group (P <= 0.51); however, SS-OCT underestimated all measurements. Conclusions: SS-OCT provides better reproducible and repeatable measurements of corneal thickness than those obtained with the Scheimpflug camera in patients with FECD or an endothelial transplant. Variations between examinations higher than the 95% LOA observed in our study should raise awareness of changes in the endothelial function.
引用
收藏
页码:727 / 733
页数:7
相关论文
共 50 条
  • [31] Reliability and repeatability of swept-source Fourier-domain optical coherence tomography and Scheimpflug imaging in keratoconus
    Szalai, Eszter
    Berta, Andras
    Hassan, Ziad
    Modis, Laszlo, Jr.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (03) : 485 - 494
  • [32] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    EYE AND VISION, 2022, 9 (01)
  • [33] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Robert Herber
    Janine Lenk
    Lutz E. Pillunat
    Frederik Raiskup
    Eye and Vision, 9
  • [34] Agreement and repeatability of corneal tomography in healthy eyes using a new swept-source OCT, a rotating Scheimpflug camera, and a dual Scheimpflug-Placido system
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2022, 48 (02) : 190 - 198
  • [35] Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer
    Li, Xin
    Lei, Chak Seng
    Ning, Rui
    Liu, Luze
    Chen, Aodong
    Yang, Xinning
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Huang, Jinhai
    EYE AND VISION, 2025, 12 (01)
  • [36] Topographic Variations of Choroidal Thickness in Healthy Eyes on Swept-Source Optical Coherence Tomography
    Touhami, Sara
    Philippakis, Elise
    Mrejen, Sarah
    Couturier, Aude
    Casteran, Celine
    Levent, Priscille
    Tadayoni, Ramin
    Gaudric, Alain
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (03)
  • [37] Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes
    Jhanji, Vishal
    Yang, Bingzhi
    Yu, Marco
    Ye, Cong
    Leung, Christopher K. S.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 41 (08) : 735 - 745
  • [38] Comparison of corneal measurements in keratoconus eyes with two swept-source-optical coherence tomography devices and a Scheimpflug device
    Alexandra Bograd
    Isabelle Himmel
    Isabel B. Pfister
    Theo G. Seiler
    Beatrice E. Frueh
    Christoph Tappeiner
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2024, 262 : 891 - 901
  • [39] Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit systems
    Lee, Yong Woo
    Choi, Chul Young
    Yoon, Geun Young
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (05) : 1018 - 1029
  • [40] Swept-source optical coherence tomography imaging of the retinochoroid and beyond
    Vira, Jayesh
    Marchese, Alessandro
    Singh, Rohan Bir
    Agarwal, Aniruddha
    EXPERT REVIEW OF MEDICAL DEVICES, 2020, 17 (05) : 413 - 426