Contrast-detail evaluation of a full-field digital mammography system

被引:0
作者
Darambara, DG [1 ]
Taibi, A [1 ]
Speller, RD [1 ]
Gambaccini, M [1 ]
机构
[1] UCL, Dept Med Phys & Bioengn, London WC1E 6JA, England
来源
1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3 | 1999年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have been investigating the potential of large-area flat-panel a-Si:H imaging devices for full-field digital mammography. The performance of a clinical mammographic system specially developed to provide geometric direct magnification equipped with an a-Si:H digital detector in conjunction with a series of Gd2O2S:Tb phosphor screens was evaluated in terms of image quality and entrance surface dose by using the contrast-detail CDMAM phantom over a range of exposure levels. We also compared the contrast-detail detectability properties of the a-Si:H-based digital imaging system with those of a conventional screen-film combination. The results from this study reveal the superiority of this prototype a-Si:H-based digital imaging system both in terms of contrast-detail resolution and significant entrance surface dose reduction in comparison with the conventional screen-film combination. These advantages are of particular significance in the breast imaging. Therefore, the combination of a-Si:H digital array with the direct magnification mammography system has the potential to fulfil the image quality requirements for full-field digital mammogarphy and to offer significant improvements in the efficient early detection of breast cancer.
引用
收藏
页码:217 / 221
页数:3
相关论文
共 6 条
[1]   Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation [J].
Antonuk, LE ;
ElMohri, Y ;
Siewerdsen, JH ;
Yorkston, J ;
Huang, W ;
Scarpine, VE ;
Street, RA .
MEDICAL PHYSICS, 1997, 24 (01) :51-70
[2]  
Bijkerk K., 1995, MANUAL CDMAM PHANTOM
[3]   X-RAY-IMAGING WITH AMORPHOUS SELENIUM - DETECTIVE QUANTUM EFFICIENCY OF PHOTOCONDUCTIVE RECEPTORS FOR DIGITAL MAMMOGRAPHY [J].
FAHRIG, R ;
ROWLANDS, JA ;
YAFFE, MJ .
MEDICAL PHYSICS, 1995, 22 (02) :153-160
[4]  
STEKLENSKI D, COMMUNICATION
[5]  
Stines J, 1998, J RADIOL, V79, P331
[6]  
WOLF CJM, 1996, EUROPEAN GUIDELINES