Intravascular ultrasound assessment of neointima distribution and the length of stent that was free of intravascular ultrasound-detectable intimal hyperplasia in paclitaxel-eluting stents

被引:16
作者
Mintz, GS
Hong, MK
Raizner, AE
Lee, CW
Kim, JJ
Escolar, E
Fearnot, NE
Park, SW
Park, SJ
Weissman, NJ
机构
[1] Washington Hosp Ctr, Washington, DC 20010 USA
[2] Univ Ulsan, Coll Med, Dept Med, Seoul, South Korea
[3] Cardiovasc Res Fdn, New York, NY USA
[4] MED Inst Inc, W Lafayette, IN USA
[5] Methodist DeBakey Heart Ctr, Cardiovasc Angiog Anal Lab, Houston, TX USA
关键词
D O I
10.1016/j.amjcard.2004.08.072
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Using data from the ASian Paclitaxel-Eluting Stent Clinical Trial, a 3-center, randomized, placebo-controlled trial of nonpolymeric paclitaxel-coated stents with a single center, 81-patient intravascular ultrasound (IVUS) substudy, the length of a stent that was free of IVUS-detectable intimal hyperplasia measured 3.2 +/- 4.8 mm in placebo stents, 6.1 +/- 5.6 mm in low-dose stents, and 8.7 +/- 6.1 mm in high-dose stents (p = 0.0029). IVUS percent neointima volume obstruction con-elated with the length of this IVUS neointima-free segment (r = 0.785, p < 0.0001); angiographic late lumen loss and follow-up diameter stenosis also correlated with the IVUS neointima-free length of the stents (r = 0.670, p < 0.0001 and r = 0.679, p < 0.0001, respectively. (C) 2005 by Excerpta Medica Inc.
引用
收藏
页码:107 / 109
页数:3
相关论文
共 11 条
[1]   Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions [J].
Colombo, A ;
Drzewiecki, J ;
Banning, A ;
Grube, E ;
Hauptmann, K ;
Silber, S ;
Dudek, D ;
Fort, S ;
Schiele, F ;
Zmudka, K ;
Guagliumi, G ;
Russell, ME .
CIRCULATION, 2003, 108 (07) :788-794
[2]   Long-term follow-up of incomplete stent apposition in patients who received sirolimus-eluting stent for de novo coronary lesions - An intravascular ultrasound analysis [J].
Degertekin, M ;
Serruys, PW ;
Tanabe, K ;
Lee, CH ;
Sousa, JE ;
Colombo, A ;
Morice, MC ;
Ligthart, JMR ;
de Feyter, PJ .
CIRCULATION, 2003, 108 (22) :2747-2750
[3]   Patterns and mechanisms of in-stent restenosis - A serial intravascular ultrasound study [J].
Hoffmann, R ;
Mintz, GS ;
Dussaillant, GR ;
Popma, JJ ;
Pichard, AD ;
Satler, LF ;
Kent, KM ;
Griffin, J ;
Leon, MB .
CIRCULATION, 1996, 94 (06) :1247-1254
[4]   Paclitaxel coating reduces in-stent intimal hyperplasia in human coronary arteries - A serial volumetric intravascular ultrasound analysis from the ASian paclitaxel-eluting stent clinical trial (ASPECT) [J].
Hong, MK ;
Mintz, GS ;
Lee, CW ;
Song, JM ;
Han, KH ;
Kang, DH ;
Song, JK ;
Kim, JJ ;
Weissman, NJ ;
Fearnot, NE ;
Park, SW ;
Park, SJ .
CIRCULATION, 2003, 107 (04) :517-520
[5]   American College of Cardiology Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies (IVUS) - A report of the American College of Cardiology Task Force on Clinical Expert Consensus - Documents developed in collaboration with the European Society of Cardiology endorsed by the Society of Cardiac Angiography and Interventions [J].
Mintz, GS ;
Nissen, SE ;
Anderson, WD ;
Bailey, SR ;
Erbel, R ;
Fitzgerald, PJ ;
Pinto, FJ ;
Rosenfield, K ;
Siegel, RJ ;
Tuzcu, EM ;
Yock, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (05) :1478-1492
[6]   A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization. [J].
Morice, M ;
Serruys, PW ;
Sousa, JE ;
Fajadet, J ;
Hayashi, EB ;
Perin, M ;
Colombo, A ;
Schuler, G ;
Barragan, P ;
Guagliumi, G ;
Molnar, F ;
Falotico, R .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (23) :1773-1780
[7]   Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery [J].
Moses, JW ;
Leon, MB ;
Popma, JJ ;
Fitzgerald, PJ ;
Holmes, DR ;
O'Shaughnessy, C ;
Caputo, RP ;
Kereiakes, DJ ;
Williams, DO ;
Teirstein, PS ;
Jaeger, JL ;
Kuntz, RE .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 349 (14) :1315-1323
[8]   A paclitaxel-eluting stent for the prevention of coronary restenosis [J].
Park, SJ ;
Shim, WH ;
Ho, DS ;
Raizner, AE ;
Park, SW ;
Hong, MK ;
Lee, CW ;
Choi, DH ;
Jang, YS ;
Lam, R ;
Weissman, NJ ;
Mintz, GS .
NEW ENGLAND JOURNAL OF MEDICINE, 2003, 348 (16) :1537-1545
[9]   Vascular responses at proximal and distal edges of paclitaxel-eluting stents - Serial intravascular ultrasound analysis from the TAXUS II trial [J].
Serruys, PW ;
Degertekin, M ;
Tanabe, K ;
Russell, ME ;
Guagliumi, G ;
Webb, J ;
Hamburger, J ;
Rutsch, W ;
Kaiser, C ;
Whitbourn, R ;
Camenzind, E ;
Meredith, I ;
Reeves, F ;
Nienaber, C ;
Benit, E ;
Disco, C ;
Koglin, J ;
Colombo, A .
CIRCULATION, 2004, 109 (05) :627-633
[10]   Intravascular ultrasound findings in the multicenter, randomized, double-blind RAVEL (RAndomized study with the sirolimus-eluting VElocity balloon-expandable stent in the treatment of patients with de novo native coronary artery Lesions) Trial [J].
Serruys, PW ;
Degertekin, M ;
Tanabe, K ;
Abizaid, A ;
Sousa, JE ;
Colombo, A ;
Guagliumi, G ;
Wijns, W ;
Lindeboom, WK ;
Ligthart, J ;
de Feyter, PJ ;
Morice, MC .
CIRCULATION, 2002, 106 (07) :798-803