A proficiency testing of weight calibration laboratories in the milligram range

被引:0
作者
Ueki, Masaaki [1 ]
Ueda, Kazunaga [1 ]
机构
[1] NMIJ, AIST, Tsukuba, Ibaraki, Japan
来源
2006 SICE-ICASE INTERNATIONAL JOINT CONFERENCE, VOLS 1-13 | 2006年
关键词
proficiency testing; JCSS; mass calibration; accredited laboratory; E-n number;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A proficiency testing on the mass calibration of standard weights was carried out among 16 laboratories under the JCSS accreditation scheme run by the NITE. The NMIJ/AIST which act as the reference laboratory, has prepared 3 sets of stainless steel weights with nominal values of 200 mg, 20 mg and 2 mg. To evaluate the mass stability of the traveling standard weights, the calibrations of the 18 weights have been made 13 times in the course of the testing. The performance of the participating laboratories is evaluated using E. numbers. It has been confirmed that 15 laboratories have sufficient measurement capability in the milligram range. Only one outlier was a calibration result of a 200 mg weight reported by a participating laboratory.
引用
收藏
页码:160 / +
页数:2
相关论文
共 50 条
  • [21] Proficiency testing to evaluate the effectiveness of laboratories analyzing active ingredient in formulation samples
    Santilio, Angela
    Pompili, Chiara
    Cammarata, Roberto
    ACCREDITATION AND QUALITY ASSURANCE, 2021, 26 (4-5) : 217 - 227
  • [22] Proficiency testing to evaluate the effectiveness of laboratories analyzing active ingredient in formulation samples
    Angela Santilio
    Chiara Pompili
    Roberto Cammarata
    Accreditation and Quality Assurance, 2021, 26 : 217 - 227
  • [23] Performance of Canadian clinical laboratories processing throat culture proficiency testing surveys
    Restelli, Veronica
    Bhuvanendran, Suhanya
    Lee, Caleb
    Kwok, Esther
    Noble, Michael
    ACCREDITATION AND QUALITY ASSURANCE, 2014, 19 (06) : 445 - 450
  • [24] Proficiency testing programs for DNA typing laboratories offered by the College of American Pathologists
    Allen, RW
    Roby, RK
    Harrison, C
    Eisenberg, AJ
    Polesky, HF
    PROGRESS IN FORENSIC GENETICS 10, 2004, 1261 : 121 - 123
  • [25] Interlaboratory proficiency testing as a tool for improving performance in laboratories diagnosing bovine mastitis
    Pitkälä, A
    Gindonis, V
    Wallin, H
    Honkanen-Buzalski, T
    JOURNAL OF DAIRY SCIENCE, 2005, 88 (02) : 553 - 559
  • [26] Proficiency testing of laboratories for paralytic shellfish poisoning toxins in shellfish by QUASIMEME: A review
    Burrell, Stephen
    Crum, Steven
    Foley, Barry
    Turner, Andrew D.
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2016, 75 : 10 - 23
  • [27] Cause analysis for unsatisfactory results in proficiency testing activities: A case study of Brazilian calibration laboratories accredited under ISO/IEC 17025:2005 â
    Silva, M.A.F.
    Moura, M.H.
    Nogueira, R.
    Costa, S.R.R.
    International Journal of Metrology and Quality Engineering, 2013, 4 (02) : 87 - 95
  • [28] Results of a proficiency testing on the calibration of clinical liquid-in-glass thermometer
    Trisna, Beni Adi
    Larassati, Dwi
    Zaid, Ghufron
    ACCREDITATION AND QUALITY ASSURANCE, 2020, 25 (01) : 1 - 5
  • [29] The International Harmonized Protocol for the proficiency testing of analytical chemistry laboratories (IUPAC technical report)
    Thompson, M
    Ellison, SLR
    Wood, R
    PURE AND APPLIED CHEMISTRY, 2006, 78 (01) : 145 - 196
  • [30] PROFICIENCY TESTING FOR AIR POLLUTION TEST LABORATORIES BY INTER-LABORATORY COMPARISON SCHEMES
    Debiasi, R.
    Nicolau, M.
    Mitran, E.
    Nicolescu, I.
    Danciulescu, V.
    Stanescu, G. B.
    JOURNAL OF ENVIRONMENTAL PROTECTION AND ECOLOGY, 2006, 7 (01): : 204 - 211