A two-stage rotary blood pump design with potentially lower blood trauma: a computational study

被引:28
|
作者
Thamsen, Bente [1 ]
Mevert, Ricardo [1 ]
Lommel, Michael [1 ]
Preikschat, Philip [1 ]
Gaebler, Julia [1 ]
Krabatsch, Thomas [2 ]
Kertzscher, Ulrich [1 ]
Hennig, Ewald [2 ]
Affeld, Klaus [1 ]
机构
[1] Charite, Biofluid Mech Lab, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] Deutsch Herzzentrum Berlin, Dept Cardiothorac & Vasc Surg, Berlin, Germany
来源
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS | 2016年 / 39卷 / 04期
关键词
Blood trauma; Computational fluid dynamics; Rotary blood pump; Two-stage pump; VENTRICULAR ASSIST DEVICES; SHEAR-STRESS; HEARTMATE II; DAMAGE; THROMBOSIS;
D O I
10.5301/ijao.5000482
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aim: In current rotary blood pumps, complications related to blood trauma due to shear stresses are still frequently observed clinically. Reducing the rotor tip speed might decrease blood trauma. Therefore, the aim of this project was to design a two-stage rotary blood pump leading to lower shear stresses. Methods: Using the principles of centrifugal pumps, two diagonal rotor stages were designed with an outer diameter of 22 mm. The first stage begins with a flow straightener and terminates with a diffusor, while a volute casing behind the second stage is utilized to guide fluid to the outlet. Both stages are combined into one rotating part which is pivoted by cup-socket ruby bearings. Details of the flow field were analyzed employing computational fluid dynamics (CFD). A functional model of the pump was fabricated and the pressure-flow dependency was experimentally assessed. Results: Measured pressure-flow performance of the developed pump indicated its ability to generate adequate pressure heads and flows with characteristic curves similar to centrifugal pumps. According to the CFD results, a pressure of 70 mmHg was produced at a flow rate of 5 L/min and a rotational speed of 3200 rpm. Circumferential velocities could be reduced to 3.7 m/s as compared to 6.2 m/s in a clinically used axial rotary blood pump. Flow fields were smooth with well-distributed pressure fields and comparatively few recirculation or vortices. Substantially smaller volumes were exposed to high shear stresses > 150 Pa. Conclusions: Hence, blood trauma might be reduced with this design. Based on these encouraging results, future in vitro investigations to investigate actual blood damage are intended.
引用
收藏
页码:178 / 183
页数:6
相关论文
共 44 条
  • [41] Development of design methods of a centrifugal blood pump with in vitro tests, flow visualization, and computational fluid dynamics: Results in hemolysis tests
    Takiura, K
    Masuzawa, T
    Endo, S
    Wakisaka, Y
    Tatsumi, E
    Taenaka, Y
    Takano, H
    Yamane, T
    Nishida, M
    Asztalos, B
    Konishi, Y
    Miyazoe, Y
    Ito, K
    ARTIFICIAL ORGANS, 1998, 22 (05) : 393 - 398
  • [42] Study on the hemodynamic effects of different pulsatile working modes of a rotary blood pump using a microfluidic platform that realizes in vitro cell culture effectively
    Liang, Lixue
    Wang, Xueying
    Chen, Dong
    Sethu, Palaniappan
    Giridharan, Guruprasad A.
    Wang, Yanxia
    Wang, Yu
    Qin, Kai-Rong
    LAB ON A CHIP, 2024, 24 (09) : 2428 - 2439
  • [43] Computational fluid dynamics and digital particle image velocimetry study of the flow through an optimized micro-axial blood pump
    Triep, M
    Brücker, C
    Schröder, W
    Siess, T
    ARTIFICIAL ORGANS, 2006, 30 (05) : 384 - 391
  • [44] Computational fluid dynamic study of hemodynamic effects on aortic root blood flow of systematically varied left ventricular assist device graft anastomosis design
    Callington, Andrew
    Long, Quan
    Mohite, Prashant
    Simon, Andre
    Mittal, Tarun Kumar
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2015, 150 (03): : 696 - 704