A two-stage rotary blood pump design with potentially lower blood trauma: a computational study

被引:28
|
作者
Thamsen, Bente [1 ]
Mevert, Ricardo [1 ]
Lommel, Michael [1 ]
Preikschat, Philip [1 ]
Gaebler, Julia [1 ]
Krabatsch, Thomas [2 ]
Kertzscher, Ulrich [1 ]
Hennig, Ewald [2 ]
Affeld, Klaus [1 ]
机构
[1] Charite, Biofluid Mech Lab, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] Deutsch Herzzentrum Berlin, Dept Cardiothorac & Vasc Surg, Berlin, Germany
来源
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS | 2016年 / 39卷 / 04期
关键词
Blood trauma; Computational fluid dynamics; Rotary blood pump; Two-stage pump; VENTRICULAR ASSIST DEVICES; SHEAR-STRESS; HEARTMATE II; DAMAGE; THROMBOSIS;
D O I
10.5301/ijao.5000482
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aim: In current rotary blood pumps, complications related to blood trauma due to shear stresses are still frequently observed clinically. Reducing the rotor tip speed might decrease blood trauma. Therefore, the aim of this project was to design a two-stage rotary blood pump leading to lower shear stresses. Methods: Using the principles of centrifugal pumps, two diagonal rotor stages were designed with an outer diameter of 22 mm. The first stage begins with a flow straightener and terminates with a diffusor, while a volute casing behind the second stage is utilized to guide fluid to the outlet. Both stages are combined into one rotating part which is pivoted by cup-socket ruby bearings. Details of the flow field were analyzed employing computational fluid dynamics (CFD). A functional model of the pump was fabricated and the pressure-flow dependency was experimentally assessed. Results: Measured pressure-flow performance of the developed pump indicated its ability to generate adequate pressure heads and flows with characteristic curves similar to centrifugal pumps. According to the CFD results, a pressure of 70 mmHg was produced at a flow rate of 5 L/min and a rotational speed of 3200 rpm. Circumferential velocities could be reduced to 3.7 m/s as compared to 6.2 m/s in a clinically used axial rotary blood pump. Flow fields were smooth with well-distributed pressure fields and comparatively few recirculation or vortices. Substantially smaller volumes were exposed to high shear stresses > 150 Pa. Conclusions: Hence, blood trauma might be reduced with this design. Based on these encouraging results, future in vitro investigations to investigate actual blood damage are intended.
引用
收藏
页码:178 / 183
页数:6
相关论文
共 44 条
  • [1] Design and Computational Evaluation of a Pediatric MagLev Rotary Blood Pump
    Tompkins, Landon H.
    Gellman, Barry N.
    Morello, Gino F.
    Prina, Steven R.
    Roussel, Thomas J.
    Kopechek, Jonathan A.
    Petit, Priscilla C.
    Slaughter, Mark S.
    Koenig, Steven C.
    Dasse, Kurt A.
    ASAIO JOURNAL, 2021, 67 (09) : 1026 - 1035
  • [2] COMPUTATIONAL FLOW OPTIMIZATION OF ROTARY BLOOD PUMP COMPONENTS
    ANTAKI, JF
    GHATTAS, O
    BURGREEN, GW
    HE, BC
    ARTIFICIAL ORGANS, 1995, 19 (07) : 608 - 615
  • [3] Numerical analysis of two-stage axial blood pump based on blood damage
    Zhou B.
    Jing T.
    Wang F.
    He Z.
    Shengwu Yixue Gongchengxue Zazhi/Journal of Biomedical Engineering, 2016, 33 (04): : 686 - 690and697
  • [4] Design and development strategy for the rotary blood pump
    Nose, Y
    ARTIFICIAL ORGANS, 1998, 22 (06) : 438 - 446
  • [5] Computational Parametric Study of the Axial and Radial Clearances in a Centrifugal Rotary Blood Pump
    Rezaienia, Mohammad Amin
    Paul, Gordon
    Avital, Eldad
    Rothman, Martin
    Korakianitis, Theodosios
    ASAIO JOURNAL, 2018, 64 (05) : 643 - 650
  • [6] Design and evaluation of a pulsatile flow rotary vane blood pump
    Li, Mengtang
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART C-JOURNAL OF MECHANICAL ENGINEERING SCIENCE, 2023, 237 (11) : 2561 - 2574
  • [7] A FLUID DYNAMIC ANALYSIS OF A ROTARY BLOOD PUMP FOR DESIGN IMPROVEMENT
    TREICHLER, J
    ROSENOW, SE
    DAMM, G
    NAITO, K
    OHARA, Y
    MIZUGUCHI, K
    MAKINOUCHI, K
    TAKATANI, S
    NOSE, Y
    ARTIFICIAL ORGANS, 1993, 17 (09) : 797 - 808
  • [8] A validated computational fluid dynamics model to estimate hemolysis in a rotary blood pump
    Arvand, A
    Hormes, M
    Reul, H
    ARTIFICIAL ORGANS, 2005, 29 (07) : 531 - 540
  • [9] Effect of the Center Post Establishment and Its Design Variations on the Performance of a Centrifugal Rotary Blood Pump
    Fang, Peng
    Du, Jianjun
    Yu, Shunzhou
    CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2020, 11 (04) : 337 - 349
  • [10] Computational Fluid Dynamics Simulation of the Sputnik Pediatric Rotary Blood Pump
    Telyshev, Dmitry
    Denisov, Maxim
    Satyukova, Anna
    Le, Tatyana
    PROCEEDINGS OF 2018 IEEE EAST-WEST DESIGN & TEST SYMPOSIUM (EWDTS 2018), 2018,