Phase-contrast MRI volume flow - a comparison of breath held and navigator based acquisitions

被引:22
作者
Andersson, Charlotta [1 ,3 ]
Kihlberg, Johan [1 ,4 ]
Ebbers, Tino [1 ]
Lindstrom, Lena [3 ]
Carlhall, Carl-Johan [1 ,2 ,5 ]
Engvall, Jan E. [1 ,2 ,5 ]
机构
[1] Linkoping Univ, Ctr Med Image Sci & Visualizat, SE-58183 Linkoping, Sweden
[2] Linkoping Univ, Dept Med & Hlth Sci, SE-58183 Linkoping, Sweden
[3] Linkoping Univ, Dept Clin Physiol, SE-60379 Norrkoping, Sweden
[4] Linkoping Univ, Dept Diagnost Radiol, SE-58185 Linkoping, Sweden
[5] Linkoping Univ, Dept Clin Physiol, SE-58185 Linkoping, Sweden
基金
英国医学研究理事会;
关键词
Phase-contrast flow; Magnetic resonance imaging; Segmentation; 2D Doppler echocardiography; CARDIOVASCULAR MAGNETIC-RESONANCE; OFFSET ERRORS; BLOOD-FLOW; QUANTIFICATION; ECHOCARDIOGRAPHY; PULMONARY; VARIABILITY; ACCURACY; SOCIETY;
D O I
10.1186/s12880-016-0128-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Magnetic Resonance Imaging (MRI) 2D phase-contrast flow measurement has been regarded as the gold standard in blood flow measurements and can be performed with free breathing or breath held techniques. We hypothesized that the accuracy of flow measurements obtained with segmented phase-contrast during breath holding, and in particular higher number of k-space segments, would be non-inferior compared to navigator phase-contrast. Volumes obtained from anatomic segmentation of cine MRI and Doppler echocardiography were used for additional reference. Methods: Forty patients, five women and 35 men, mean age 65 years (range 53-80), were randomly selected and consented to the study. All underwent EKG-gated cardiac MRI including breath hold cine, navigator based free-breathing phase-contrast MRI and breath hold phase-contrast MRI using k-space segmentation factors 3 and 5, as well as transthoracic echocardiography within 2 days. Results: In navigator based free-breathing phase-contrast flow, mean stroke volume and cardiac output were 79.7 +/- 17.1 ml and 5071 +/- 1192 ml/min, respectively. The duration of the acquisition was 50 +/- 6 s. With k-space segmentation factor 3, the corresponding values were 77.7 ml +/- 17.5 ml and 4979 +/- 1211 ml/min (p = 0.15 vs navigator). The duration of the breath hold was 17 +/- 2 s. K-space segmentation factor 5 gave mean stroke volume 77.9 +/- 16.4 ml, cardiac output 5142 +/- 1197 ml/min (p = 0.33 vs navigator), and breath hold time 11 +/- 1 s. Anatomical segmentation of cine gave mean stroke volume and cardiac output 91.2 +/- 20.8 ml and 5963 +/- 1452 ml/min, respectively. Echocardiography was reliable in 20 of the 40 patients. The mean diameter of the left ventricular outflow tract was 20.7 +/- 1.5 mm, stroke volume 78.3 ml +/- 15.2 ml and cardiac output 5164 +/- 1249 ml/min. Conclusions: In forty consecutive patients with coronary heart disease, breath holding and segmented k-space sampling techniques for phase-contrast flow produced stroke volumes and cardiac outputs similar to those obtained with free-breathing navigator based phase-contrast MRI, using less time. The values obtained agreed fairly well with Doppler echocardiography while there was a larger difference when compared with anatomical volume determinations using SSFP (steady state free precession) cine MRI.
引用
收藏
页数:8
相关论文
共 27 条
[1]   Concomitant gradient terms in phase contrast MR: Analysis and correction [J].
Bernstein, MA ;
Zhou, XHJ ;
Polzin, JA ;
King, KF ;
Ganin, A ;
Pelc, NJ ;
Glover, GH .
MAGNETIC RESONANCE IN MEDICINE, 1998, 39 (02) :300-308
[2]   How we perform cardiovascular magnetic resonance flow assessment using phase-contrast velocity mapping [J].
Chai, P ;
Mohiaddin, R .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2005, 7 (04) :705-716
[3]   Baseline correction of phase contrast images improves quantification of blood flow in the great vessels [J].
Chernobelsky, Alexander ;
Shubayev, Oleg ;
Comeau, Cindy R. ;
Wolff, Steven D. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2007, 9 (04) :681-685
[4]  
Douglas GA, 1991, PRACTICAL STAT MED R, P365
[5]   Applications of phase-contrast flow and velocity imaging in cardiovascular MRI [J].
Gatehouse, PD ;
Keegan, J ;
Crowe, LA ;
Masood, S ;
Mohiaddin, RH ;
Kreitner, KF ;
Firmin, DN .
EUROPEAN RADIOLOGY, 2005, 15 (10) :2172-2184
[6]   A multi-center inter-manufacturer study of the temporal stability of phase-contrast velocity mapping background offset errors [J].
Gatehouse, Peter D. ;
Rolf, Marijn P. ;
Bloch, Karin Markenroth ;
Graves, Martin J. ;
Kilner, Philip J. ;
Firmin, David N. ;
Hofman, Mark B. M. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2012, 14
[7]   Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements [J].
Gatehouse, Peter D. ;
Rolf, Marijn P. ;
Graves, Martin J. ;
Hofman, Mark B. M. ;
Totman, John ;
Werner, Beat ;
Quest, Rebecca A. ;
Liu, Yingmin ;
von Spiczak, Jochen ;
Dieringer, Matthias ;
Firmin, David N. ;
van Rossum, Albert ;
Lombardi, Massimo ;
Schwitter, Juerg ;
Schulz-Menger, Jeanette ;
Kilner, Philip J. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2010, 12
[8]   Head-to-Head Comparison of Left Ventricular Function Assessment with 64-Row Computed Tomography, Biplane Left Cineventriculography, and Both 2- and 3-Dimensional Transthoracic Echocardiography Comparison With Magnetic Resonance Imaging as the Reference Standard [J].
Greupner, Johannes ;
Zimmermann, Elke ;
Grohmann, Andrea ;
Duebel, Hans-Peter ;
Althoff, Till ;
Borges, Adrian C. ;
Rutsch, Wolfgang ;
Schlattmann, Peter ;
Hamm, Bernd ;
Dewey, Marc .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (21) :1897-1907
[9]   Accuracy of Right and Left Ventricular Functional Assessment by Short-Axis vs Axial Cine Steady-State Free-Precession Magnetic Resonance Imaging: Intrapatient Correlation With Main Pulmonary Artery and Ascending Aorta Phase-Contrast Flow Measurements [J].
James, Susan H. ;
Wald, Rachel ;
Wintersperger, Bernd J. ;
Jimenez-Juan, Laura ;
Deva, Djeven ;
Crean, Andrew M. ;
Elsie Nguyen ;
Paul, Narinder S. ;
Ley, Sebastian .
CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2013, 64 (03) :213-219
[10]   The effects of breath-holding on pulmonary regurgitation measured by cardiovascular magnetic resonance velocity mapping [J].
Johansson, Bengt ;
Babu-Narayan, Sonya V. ;
Kilner, Philip J. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2009, 11