Low-dose interpolated average CT for attenuation correction in cardiac PET/CT

被引:6
作者
Wu, Tung-Hsin [4 ]
Zhang, Geoffrey [3 ]
Wang, Shyh-Jen [2 ,4 ]
Chen, Chih-Hao [2 ]
Yang, Bang-Hung [2 ,4 ]
Wu, Nien-Yun [2 ]
Huang, Tzung-Chi [1 ]
机构
[1] China Med Univ, Dept Biomed Imaging & Radiol Sci, Taichung, Taiwan
[2] Taipei Vet Gen Hosp, Dept Nucl Med, Taipei, Taiwan
[3] H Lee Moffitt Canc Ctr & Res Inst, Dept Radiat Oncol, Tampa, FL USA
[4] Natl Yang Ming Univ, Dept Biomed Imaging & Radiol Sci, Taipei, Taiwan
关键词
Attenuation correction; Cine CT; Deformable image registration; PET; OPTICAL-FLOW; IMAGES;
D O I
10.1016/j.nima.2009.11.001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Because of the advantages in the use of high photon flux and thus the short scan times of CT imaging, the traditional Ge-68 scans for positron emission tomography (PET) image attenuation correction have been replaced by CT scans in the modern PET/CT technology. The combination of fast CT scan and slow PET scan often causes image misalignment between the PET and CT images due to respiration motion. Use of the average CT derived from cine CT images is reported to reduce such misalignment. However, the radiation dose to patients is higher with cine CT scans. This study introduces a method that uses breath-hold CT images and their interpolations to generate the average CT for PET image attenuation correction. Breath-hold CT sets are taken at end-inspiration and end-expiration. Deformable image registration is applied to generate a voxel-to-voxel motion matrix between the two CT sets. The motion is equally divided into 5 steps from inspiration to expiration and 5 steps from expiration to inspiration, generating a total of 8 phases of interpolated CT sets. An average CT image is generated from all the 10 phase CT images, including original inhale/exhale CT and 8 interpolated CT sets. Quantitative comparison shows that the reduction of image misalignment artifacts using the average CT from the interpolation technique for PET attenuation correction is at a similar level as that using cine average CT, while the dose to the patient from the CT scans is reduced significantly. The interpolated average CT method hence provides a low dose alternative to cine CT scans for PET attenuation correction. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:361 / 364
页数:4
相关论文
共 50 条
  • [31] Validation of the CT iterative reconstruction technique for low-dose CT attenuation correction for improving the quality of PET images in an obesity-simulating body phantom and clinical study
    Matsutomo, Norikazu
    Nagaki, Akio
    Sasaki, Masayuki
    NUCLEAR MEDICINE COMMUNICATIONS, 2015, 36 (08) : 839 - 847
  • [32] Reducing radiation dose in rest-stress cardiac PET/CT by single poststress cine CT for attenuation correction: Quantitative validation
    Gould, K. Lance
    Pan, Tinsu
    Loghin, Catalin
    Johnson, Nils P.
    Sdringola, Stefano
    JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (05) : 738 - 745
  • [33] Low-dose CT and contrast-medium CT in hybrid PET/CT systems for oncologic patients
    Chiaravalloti, Agostino
    Rubello, Domenico
    Chondrogiannis, Sotirios
    Giammarile, Francesco
    Colletti, Patrick M.
    Schillaci, Orazio
    NUCLEAR MEDICINE COMMUNICATIONS, 2015, 36 (09) : 867 - 870
  • [34] CT-Based Attenuation Correction on the FLEX Triumph Preclinical PET/CT Scanner
    Prasad, Rameshwar
    Ay, Mohammad R.
    Ratib, Osman
    Zaidi, Habib
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (01) : 66 - 75
  • [35] Anatomical-based partial volume correction for low-dose dedicated cardiac SPECT/CT
    Liu, Hui
    Chan, Chung
    Grobshtein, Yariv
    Ma, Tianyu
    Liu, Yaqiang
    Wang, Shi
    Stacy, Mitchel R.
    Sinusas, Albert J.
    Liu, Chi
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (17) : 6751 - 6773
  • [36] A Curve Fitting Approach Using ANN for Converting CT Number to Linear Attenuation Coefficient for CT-based PET Attenuation Correction
    Lai, Chia-Lin
    Lee, Jhih-Shian
    Chen, Jyh-Cheng
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (01) : 164 - 170
  • [37] Automated PET/CT Brain Registration for Accurate Attenuation Correction
    Khurshid, Khawar
    Berger, Kevin L.
    McGough, Robert J.
    2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 5805 - 5808
  • [38] Attenuation Correction for PET/MRI Using MRI-based Pseudo CT
    Wang, Tonghe
    Lei, Yang
    Dong, Xue
    Higgins, Kristin
    Liu, Tian
    Curran, Walter J.
    Mao, Hui
    Nye, Jonathon A.
    Yang, Xiaofeng
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [39] Attenuation correction of PET images with respiration-averaged CT images in PET/CT
    Pan, TS
    Mawlawi, C
    Nehmeh, SA
    Erdi, YE
    Luo, DS
    Liu, HH
    Castillo, R
    Mohan, R
    Liao, ZX
    Macapinlac, HA
    JOURNAL OF NUCLEAR MEDICINE, 2005, 46 (09) : 1481 - 1487
  • [40] Revealing Anatomical Structures in PET to Generate CT for Attenuation Correction
    Pan, Yongsheng
    Liu, Feihong
    Jiang, Caiwen
    Huang, Jiawei
    Xia, Yong
    Shen, Dinggang
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION, MICCAI 2023, PT X, 2023, 14229 : 24 - 33