Correlation between Central Corneal Thickness and Intraocular Pressure in Various Age Groups

被引:3
作者
Marjanovic, Ivan [1 ]
Kontic, Djordje [1 ]
Hentova-Sencanic, Paraskeva [1 ]
Markovic, Vujica [1 ]
Bozic, Marija [1 ]
机构
[1] Clin Ctr Serbia, Inst Eye Dis, Belgrade 11000, Serbia
关键词
central corneal thickness; dynamic contour tonometry; Goldmann applanation tonometry; age groups; DYNAMIC CONTOUR TONOMETRY; GOLDMANN APPLANATION TONOMETER; GLAUCOMA; ACCURACY;
D O I
10.2298/SARH1006279M
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction The Pascal Dynamic Contour Tonometer (DCT) is an ophthalmic investigational device which directly calculates the dynamic pulsatile fluctuations in intraocular pressure (IOP) using a piezoelectric pressure sensor embedded in the devices tip. Objective The aim of this study was to compare IOP results obtained with Goldmann applanation tonometer (GAT), and their correlation with central corneal thickness (CCT) in patients of various age groups. Methods We studied 37 patients (17 male and 20 female), divided into three age groups: younger than 40 years; 40-60 years old, and older than 60 years. In the first and second group there were patients rarely and in the third mostly diagnosed and medicamentously compensated glaucoma. In topical anaesthesia, first we measured CCT, with ultrasound pachimeter, then IOP with DCT and GAT. Results Statistically significant difference between measurements of IOP with the DCT and GAT appeared in all groups: I mean cliff. -1.71 +/- 1.27 mm; p<0.0001; II mean cliff. -1.19 +/- 1.06 mm; p<0.0001; Ill mean cliff. -1.69 +/- 1.67 mm; p<0.0001. CCT was in indirect correlation with the values of lop measured both with the DCT and GAT in the first and third, while it was in the direct correlation with these values in the second group. Conclusion CCT had no influence on IOP measurements both with the DCT and GAT in none of the groups. The DCT cannot replace GAT, but it is a reliable device for the measurement of lop particularly in corneal deformations (keratoconus, after corneal refractive surgery, corneal scars, etc.).
引用
收藏
页码:279 / 286
页数:8
相关论文
共 25 条
[1]  
Bhan A, 2002, INVEST OPHTH VIS SCI, V43, P1389
[2]   Dynamic contour tonometry in comparison to intracameral IOP measurements [J].
Boehm, Andreas G. ;
Weber, Anja ;
Pillunat, Lutz E. ;
Koch, Rainer ;
Spoerl, Eberhard .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (06) :2472-2477
[3]  
BOHM AG, 2008, EFFECT AGE OCULAR PU
[4]   Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS) [J].
Brandt, JD ;
Beiser, JA ;
Kass, MA ;
Gordon, MO .
OPHTHALMOLOGY, 2001, 108 (10) :1779-1788
[5]   RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES [J].
DIELEMANS, I ;
VINGERLING, JR ;
HOFMAN, A ;
GROBBEE, DE ;
DEJONG, PTVM .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1994, 232 (03) :141-144
[6]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[7]   Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness [J].
Doyle, A ;
Lachkar, Y .
JOURNAL OF GLAUCOMA, 2005, 14 (04) :288-292
[8]   Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings [J].
Feltgen, N ;
Leifert, D ;
Funk, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (01) :85-87
[9]  
GOLDMANN H, 1957, Ophthalmologica, V134, P221
[10]   Central corneal thickness as a risk factor for advanced glaucoma damage [J].
Herndon, LW ;
Weizer, JS ;
Stinnett, SS .
ARCHIVES OF OPHTHALMOLOGY, 2004, 122 (01) :17-21