Delayed perforation of the right ventricle as a complication of permanent cardiac pacing - is following the guidelines always the right choice? Non-standard treatment - a case report and literature review

被引:0
作者
Rydlewska, Anna
Malecka, Barbara
Zabek, Andrzej
Klimeczek, Piotr [1 ]
Lelakowski, Jacek
Pasowicz, Mieczyslaw [1 ]
Czajkowski, Marek [2 ]
Kutarski, Andrzej [3 ]
机构
[1] John Paul 2 Hosp, Ctr Diag Prevent & Telemed, Day Diagnost Unit 1, Krakow, Poland
[2] Med Univ, Dept Cardiol, Lublin, Poland
[3] Med Univ, Dept Cardiosurg, Lublin, Poland
关键词
permanent heart stimulation; complication; delayed perforation of the right ventricle; LEAD PERFORATION; ICD LEAD; PACEMAKER; HEART;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A case of a delayed perforation of the right ventricle by the pacemaker lead in a 67-year-old woman is presented. Perforation, mimicking stenocardial symptoms, was incidentally diagnosed on a computed tomography chest scan. Percutaneous lead extraction was successfully performed, with simultaneous implantation of a new pacemaker lead.
引用
收藏
页码:357 / 361
页数:5
相关论文
共 19 条
[1]   Late perforation of the heart, pericardium, and diaphragm by an active-fixation ventricular lead [J].
Akyol, A ;
Aydin, A ;
Erdinler, I ;
Oguz, E .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04) :350-351
[2]   Hiccup as a result of late lead perforation: report of two cases and review of the literature [J].
Celik, Turgay ;
Kose, Sedat ;
Bugan, Baris ;
Iyisoy, Atila ;
Akgun, Veysel ;
Cingoz, Faruk .
EUROPACE, 2009, 11 (07) :963-965
[3]  
Fisher JD, 2008, PACE, V31, P7, DOI 10.1111/j.1540-8159.2007.00918.x
[4]   Right ventricular lead perforation presenting as left chest wall muscle stimulation [J].
Greenberg, S ;
Lawton, J ;
Chen, J .
CIRCULATION, 2005, 111 (25) :E451-E452
[5]   Late Presentation of a Lead Perforation as a Complication of Permanent Pacemaker Insertion [J].
Haq, Salman A. ;
Heitner, John F. ;
Lee, Leonard ;
Kassotis, John T. .
ANGIOLOGY, 2008, 59 (05) :619-621
[6]   Computed tomography to assess possible cardiac lead perforation [J].
Henrikson, CA ;
Leng, CT ;
Yuh, DD ;
Brinker, JA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (05) :509-511
[7]  
Hirschl DA, 2007, PACE, V30, P28
[8]   Delayed lead perforation: A disturbing trend [J].
Khan, MN ;
Joseph, G ;
Khaykin, Y ;
Ziada, KM ;
Wilkoff, BL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (03) :251-253
[9]  
Krivan L, 2008, PACE, V31, P3, DOI 10.1111/j.1540-8159.2007.00917.x
[10]   Management of Subacute and Delayed Right Ventricular Perforation With a Pacing or an Implantable Cardioverter-Defibrillator Lead [J].
Laborderie, Julien ;
Barandon, Laurent ;
Ploux, Sylvain ;
Deplagne, Antoine ;
Mokrani, Bilel ;
Reuter, Sylvain ;
Le Gal, Francois ;
Jais, Pierre ;
Haissaguerre, Michel ;
Clementy, Jacques ;
Bordachar, Pierre .
AMERICAN JOURNAL OF CARDIOLOGY, 2008, 102 (10) :1352-1355