A novel microfluidic flow-injection analysis device with fluorescence detection for cation sensing.: Application to potassium

被引:27
作者
Destandau, Emilie [1 ]
Lefevre, Jean-Pierre [1 ]
Eddine, Assia Chouai Fakhr [1 ]
Desportes, Serge [1 ]
Jullien, Marie Caroline [1 ]
Hierle, Rolland [1 ]
Leray, Isabelle [1 ]
Valeur, Bernard [1 ]
Delaire, Jacques Alexis [1 ]
机构
[1] ENS Cachan, DAlembert Inst, F-94235 Cachan, France
关键词
mu TAS; fluorimetric detection; potassium sensor; flow-injection analysis; passive mixer;
D O I
10.1007/s00216-007-1132-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A microfabricated device has been developed for fluorimetric detection of potassium ions without previous separation. It is based on use of a fluorescent molecular sensor, calix-bodipy, specially designed to be sensitive to and selective for the target ion. The device is essentially made of a Y-shape microchannel moulded in PDMS fixed on a glass substrate. A passive mixer is used for mixing the reactant and the analyte. The optical detection arrangement uses two optical fibres, one for excitation by a light-emitting diode, the other for collection of the fluorescence. This system enabled the flow-injection analysis of the concentration of potassium ions in aqueous solutions with a detection limit of 0.5 mmol L-1 and without interference with sodium ions. A calibration plot was constructed using potassium standard solutions in the range 0-16 mmol L-1, and was used for the determination of the potassium content of a pharmaceutical pill.
引用
收藏
页码:2627 / 2632
页数:6
相关论文
共 15 条
[1]   Fluorescent ion-selective optode membranes incorporated onto a centrifugal microfluidics platform [J].
Badr, IHA ;
Johnson, RD ;
Madou, MJ ;
Bachas, LG .
ANALYTICAL CHEMISTRY, 2002, 74 (21) :5569-5575
[2]   Determination of inorganic ions using microfluidic devices [J].
Evenhuis, CJ ;
Guijt, RM ;
Macka, M ;
Haddad, PR .
ELECTROPHORESIS, 2004, 25 (21-22) :3602-3624
[3]   CALCULATION OF EQUILIBRIUM-CONSTANTS FROM MULTIWAVELENGTH SPECTROSCOPIC DATA .1. MATHEMATICAL CONSIDERATIONS [J].
GAMPP, H ;
MAEDER, M ;
MEYER, CJ ;
ZUBERBUHLER, AD .
TALANTA, 1985, 32 (02) :95-101
[4]  
Guijt RM, 2001, ELECTROPHORESIS, V22, P2537, DOI 10.1002/1522-2683(200107)22:12<2537::AID-ELPS2537>3.0.CO
[5]  
2-C
[6]  
Guijt RM, 2001, ELECTROPHORESIS, V22, P235, DOI 10.1002/1522-2683(200101)22:2<235::AID-ELPS235>3.0.CO
[7]  
2-O
[8]   On-chip integration of neutral ionophore-based ion pair extraction reaction [J].
Hisamoto, H ;
Horiuchi, T ;
Tokeshi, M ;
Hibara, A ;
Kitamori, T .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1382-1386
[9]   A highly selective fluorescent molecular sensor for potassium based on a calix[4]bisazacrown bearing boron-dipyrromethene fluorophores [J].
Malval, JP ;
Leray, I ;
Valeur, B .
NEW JOURNAL OF CHEMISTRY, 2005, 29 (08) :1089-1094
[10]   Single electrode conductivity detection for electrophoretic separation systems [J].
Prest, JE ;
Baldock, SJ ;
Bektas, N ;
Fielden, PR ;
Brown, BJT .
JOURNAL OF CHROMATOGRAPHY A, 1999, 836 (01) :59-65