The influence of different technique factors on image quality for chest radiographs:: Application of the recent CEC image quality criteria

被引:12
作者
Lanhede, B [1 ]
Tingberg, A
Månsson, LG
Kheddache, S
Widell, M
Björneld, L
Sund, P
机构
[1] Sahlgrens Univ Hosp, Dept Radiat Phys, SE-41345 Gothenburg, Sweden
[2] Malmo Univ Hosp, Dept Radiat Phys, SE-20502 Malmo, Sweden
[3] Sahlgrens Univ Hosp, Dept Radiol, SE-41345 Gothenburg, Sweden
[4] Malmo Univ Hosp, Dept Diagnost Radiol, SE-20502 Malmo, Sweden
[5] GSF, Natl Res Ctr Environm & Hlth, Neuherberg, Germany
关键词
D O I
10.1093/oxfordjournals.rpd.a033119
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The aim of this work was to evaluate and possibly improve the CEC image criteria for radiographic chest images. Chest images of healthy Volunteers were acquired using different technique factors. The image criteria were used as a tool to discriminate between the different images. The technique factors were chosen so that the image quality would differ slightly. Four different technique parameters, each with two possible settings used in clinical practice today, were used: tube voltage, 102 and 141 kV; screen-film speed, 160 and 320; maximum optical density in the parenchyma, 1.3 and 1.8; method for scatter reduction, air gap and moving grid. The results showed that the image criteria were able to distinguish different technique groups. Optical density 1.8 was better than 1.3 independent of the other parameters. No difference was seen for screen-film speed. No correlation was seen between the ranking of the systems and patient dose.
引用
收藏
页码:203 / 206
页数:4
相关论文
共 5 条
[1]  
ALMEN A, UNPUB BR J RADIOL
[2]  
*EUR COMM, 1996, 16260 EN EUR COMM
[3]  
KHEDDACHE S, IN PRESS EUROPEAN CO
[4]  
MACCIA C, 1990, 12952 EUR
[5]  
ZANKL M, UNPUB INFLUENCE DIFF