Intra-Aortic Balloon Pump Malposition Reduces Visceral Artery Perfusion in an Acute Animal Model

被引:4
作者
Vondran, Maximilian [1 ]
Rastan, Ardawan J. [1 ]
Tillmann, Eugen [1 ]
Seeburger, Joerg [1 ]
Schroeter, Thomas [1 ]
Dhein, Stefan [1 ]
Bakhtiary, Farhad [1 ]
Mohr, Friedrich-Wilhelm [1 ]
机构
[1] Univ Leipzig, Dept Cardiac Surg, Heart Ctr Leipzig, D-04289 Leipzig, Germany
关键词
Animal model; Circulatory assistance; Heart failure; Intensive care; Mesenteric ischemia; Cardiac surgery; Intra-aortic balloon pump; Mesenteric artery; MESENTERIC ISCHEMIA; GASTROINTESTINAL COMPLICATIONS; VASCULAR COMPLICATIONS; CELIAC AXIS; COUNTERPULSATION; OUTCOMES; SURGERY; RISK; FLOW;
D O I
10.1111/aor.12563
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Visceral artery perfusion can be potentially affected by intra-aortic balloon pump (IABP) catheters. We utilized an animal model to quantify the acute impact of a low balloon position on mesenteric artery perfusion. In six pigs (78 +/- 7kg), a 30-cc IABP was placed in the descending aorta in a transfemoral procedure. The celiac artery (CA) and the cranial mesenteric artery (CMA) were surgically dissected. Transit time blood flow was measured for (i) baseline, (ii) 1:1 augmentation with the balloon proximal to the visceral arteries, and (iii) 1:1 augmentation with the balloon covering the visceral arteries. Blood flow in the CMA and CA was reduced by 17 and 24%, respectively, when the balloon compromised visceral arteries compared with a position above the visceral arteries (flow in mL/min: CMA: (i) 1281 +/- 512, (ii) 1389 +/- 287, (iii) 1064 +/- 276, P<0.05 for 3 vs. 1 and 3 vs. 2; CA: (i) 885 +/- 370, (ii) 819 +/- 297, (iii) 673 +/- 315; P<0.05 for 3 vs. 1). The covering of visceral arteries by an IABP balloon causes a significant reduction of visceral artery perfusion; thus, the positioning of this device during implantation is critical for obtaining a satisfactory outcome.
引用
收藏
页码:334 / 340
页数:7
相关论文
共 32 条
[1]  
Aouifi A, 1999, CAN J ANAESTH, V46, P114, DOI 10.1007/BF03012544
[2]   Vascular complications of the intraaortic balloon pump in patients undergoing open heart operations: 15-year experience [J].
Arafa, OE ;
Pedersen, TH ;
Svennevig, JL ;
Fosse, E ;
Geiran, OR .
ANNALS OF THORACIC SURGERY, 1999, 67 (03) :645-651
[3]   End-diastolic flow reversal limits the efficacy of pediatric intra-aortic balloon pump counterpulsation [J].
Bartoli, Carlo R. ;
Rogers, Benjamin D. ;
Ionan, Constantine E. ;
Pantalos, George M. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 147 (05) :1660-1667
[4]   The intraaortic balloon pump in cardiac surgery [J].
Baskett, RJF ;
Ghali, WA ;
Maitland, A ;
Hirsch, GM .
ANNALS OF THORACIC SURGERY, 2002, 74 (04) :1276-1287
[5]   Acute increase in reversal blood flow during counterpulsation is associated with vasoconstriction and changes in the aortic mechanics [J].
Bia, Daniel ;
Zocalo, Yanina ;
Armentano, Ricardo ;
de Forteza, Eduardo ;
Cabrera-Fischer, Edmundo .
2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, :3986-3989
[6]   MESENTERIC ISCHEMIA - A MULTIDISCIPLINARY APPROACH [J].
BRADBURY, AW ;
BRITTENDEN, J ;
MCBRIDE, K ;
RUCKLEY, CV .
BRITISH JOURNAL OF SURGERY, 1995, 82 (11) :1446-1459
[7]   Potential Risk for Intra-Aortic Balloon-Induced Obstruction to the Celiac Axis or the Renal Artery in the Asian Population [J].
Byon, H-J ;
Kim, H. ;
Kim, H-C ;
Kim, J-T ;
Kim, H-S ;
Lee, S-C ;
Kim, C. S. .
THORACIC AND CARDIOVASCULAR SURGEON, 2011, 59 (02) :99-102
[8]   Should We Consider the Ethnic Difference in Selecting Size of Intraaortic Balloon by Commercial Guideline? [J].
Cho, Young-Seok ;
Lim, Cheong ;
Han, Mi-Jung ;
Chun, Eun Ju ;
Choi, Sang Il ;
Park, Kay-Hyun ;
Chae, In-Ho ;
Choi, Dong-Ju .
ASAIO JOURNAL, 2009, 55 (05) :519-522
[9]   Trends in intraaortic balloon counterpulsation complications and outcomes in cardiac surgery [J].
Christenson, JT ;
Cohen, M ;
Ferguson, JJ ;
Freedman, RJ ;
Miller, MF ;
Ohman, M ;
Reddy, RC ;
Stone, GW ;
Urban, PM .
ANNALS OF THORACIC SURGERY, 2002, 74 (04) :1086-1090
[10]   Intra-aortic balloon counterpulsation in US and non-US centres:: Results of the Benchmark® Registry [J].
Cohen, M ;
Urban, P ;
Christenson, JT ;
Joseph, DL ;
Freedman, RJ ;
Miller, MF ;
Ohman, EM ;
Reddy, RC ;
Stone, GW ;
Ferguson, JJ .
EUROPEAN HEART JOURNAL, 2003, 24 (19) :1763-1770