Objective We compared an infrared temporal artery thermometer with our clinical standard axillary thermometer for temperature measurements in neonatal patients. Study Design We measured temporal artery (T-ta), axillary (T-ax, clinical standard), and rectal (T-r, gold standard) temperatures of 49 infants. The difference between T-r and T-ta was compared with that between Tr and Tax, and the data were analyzed based on bed type and postmenstrual age. Results The mean T-ta, T-ax, and T-r were 37.16 (SD 0.36) degrees C, 36.61 (SD 0.30) degrees C, and 36.82 (SD 0.30) degrees C, respectively. The measurements by these methods were all significantly different. The mean T-r-T-ax was 0.21 (SD 0.26) degrees C, and the mean T-r-T-ta was -0.34 (SD 0.37) degrees C, indicating that Tax was closer to Tr than was Tta (p < 0.0001). T-ta agreed more closely with Tr for infants in cribs than for those in incubators. Adjusting for bed type and body weight, with each week of postmenstrual age, the discrepancy between T-r-T-ta and T-r-T-ax decreased by 0.005 degrees C (p = 0.034). Conclusion Compared with the gold standard, T-r, Tta is not more accurate than T-ax. The temporal artery thermometer was less accurate for infants in incubators than for infants in cribs. The accuracy of temporal artery temperature increased with postmenstrual age.
引用
收藏
页码:1026 / 1031
页数:6
相关论文
共 50 条
[31]   A Neonate of Coronary Artery Fistula from Left Circumflex to Coronary Sinus [J].
Jingang Li ;
Hisanori Sakazaki ;
Kazuya Tsukuda ;
Keiichi Fujiwara .
Pediatric Cardiology, 2011, 32 :1212-1215
[32]   Heart failure and coronary ischemia in a neonate with right coronary artery fistula [J].
Shimada, Sorachi ;
Taketazu, Mio ;
Sato, Masayuki ;
Nii, Mitsumaro ;
Shirai, Masaru .
PEDIATRICS INTERNATIONAL, 2019, 61 (04) :417-418
[33]   Abnormal Origin of the Pulmonary Artery from the Ascending Aorta in the Neonate br [J].
Jonas, Marina Costa ;
Martins, Cristiane Nunes ;
Gontijo Filho, Bayard ;
Ramalho, Clarissa de Paula Freitas Rocha ;
Vrandecic, Erika Correa ;
Amaral, Fernando .
BRAZILIAN JOURNAL OF CARDIOVASCULAR SURGERY, 2022, 37 (06) :937-941
[34]   A Neonate of Coronary Artery Fistula from Left Circumflex to Coronary Sinus [J].
Li, Jingang ;
Sakazaki, Hisanori ;
Tsukuda, Kazuya ;
Fujiwara, Keiichi .
PEDIATRIC CARDIOLOGY, 2011, 32 (08) :1212-1215
[35]   tuf-PCR-temporal temperature gradient gel electrophoresis for molecular detection and identification of staphylococci: Application to breast milk and neonate gut microbiota [J].
Filleron, Anne ;
Simon, Margaux ;
Hantova, Stefaniya ;
Jacquot, Aurelien ;
Cambonie, Gilles ;
Marchandin, Helene ;
Jumas-Bilak, Estelle .
JOURNAL OF MICROBIOLOGICAL METHODS, 2014, 98 :67-75
[36]   Middle Cerebral Artery Stroke in a Neonate With a Congenital Hypercoagulable Condition Following Repair of an H-type Tracheoesophageal Fistula [J].
Garcia, Megan L. ;
Caldwell, Katharine ;
Kellund, Anna ;
Reyes, Cynthia .
JOURNAL OF PEDIATRIC HEMATOLOGY ONCOLOGY, 2016, 38 (07) :529-532
[37]   Anomalous Origin of the Right Coronary Artery from the Pulmonary Artery in a Neonate with Turner Syndrome and Aortic Arch Hypoplasia [J].
Stefek, Bryan P. ;
Imundo, Jason R. ;
Clark, Joseph B. .
TEXAS HEART INSTITUTE JOURNAL, 2019, 46 (03) :225-228
[38]   Temperature measurements in the neonate: the post-mercury thermometer era [J].
Semama, DS ;
Gouyon, JB .
ARCHIVES DE PEDIATRIE, 1999, 6 (06) :686-687
[39]   An Unusual Case of Brachial Artery Thrombosis in a Neonate Following Humerus fracture at Birth [J].
Raghavendra, Prashanth Ranya ;
Nair, Sruthi ;
Mujeeb, Athar Abdul ;
Haribalakrishna, Anitha .
JOURNAL OF CLINICAL NEONATOLOGY, 2024, 13 (02) :63-66
[40]   Evaluation of Temporal Artery Thermometry in Children 3-36 Months Old [J].
Holzhauer, Janet K. ;
Reith, Vicky ;
Sawin, Kathleen J. ;
Yen, Kenneth .
JOURNAL FOR SPECIALISTS IN PEDIATRIC NURSING, 2009, 14 (04) :239-244