Nano Liquid Chromatography Directly Coupled to Electron Ionization Mass Spectrometry for Free Fatty Acid Elucidation in Mussel

被引:59
作者
Rigano, Francesca [1 ,2 ]
Albergamo, Ambrogina [1 ]
Sciarrone, Danilo [1 ]
Beccaria, Marco [2 ]
Purcaro, Giorgia [2 ]
Mondello, Luigi [1 ,2 ]
机构
[1] Univ Messina, Dipartimento Sci Chim Biol Farmaceut & Ambientali, Viale Annunziata, I-98168 Messina, Italy
[2] Univ Messina, Chromaleont Srl, Viale Annunziata, I-98168 Messina, Italy
关键词
AEROSOL GENERATION INTERFACE; GAS-CHROMATOGRAPHY; DERIVATIZATION; PERFORMANCE; EXTRACTION; SEPARATION; DETECTOR; SAMPLES; ESTERS; PHASE;
D O I
10.1021/acs.analchem.6b00328
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Recently the miniaturization of liquid chromatography (LC) systems and progresses in mass spectrometry instrumentation have enabled direct introduction of the effluent coming from a nanoLC column into the high-vacuum region of an electron ionization source. In the present research, a nanoLC system was directly coupled to an electron ionization mass spectrometer (EI-MS) without any interface or modification of the ion source. The advantage with respect to atmospheric pressure ionization techniques, normally coupled with LC, is major identification power because of a more extensive and reproducible fragmentation pattern, without any matrix effect or mobile-phase interference. In particular, a nanoLC/EI-MS method was developed for elucidation of the free fatty acid profile in mussel samples, avoiding a previous derivatization step, required when gas chromatographic analysis is involved. A total of 20 fatty acids were reliably identified through the comparison with commercial libraries. A quantitative determination was also carried out by using the response factors approach along with the internal standard method, allowing for quantification of 14 fatty acids. Among them, palmitic acid resulted the most abundant, followed by omega 6 arachidonic acid. The quantitative data were compared with those obtained by a well-established technique, such as gas chromatography with flame ionization detection (GC-FID). Both nanoLC/EI-MS and GC-FID methods were validated and similar results were obtained in terms of limit of detection and quantification, resulting in the picomole range, and sensitivity, as well was not significantly different, as demonstrated by comparing the slope values of the calibration curves (p < 0.05, from a t-test).
引用
收藏
页码:4021 / 4028
页数:8
相关论文
共 37 条
[1]   Isolation and structure determination of algicidal compounds from Ulva fasciata [J].
Alamsjah, MA ;
Hirao, S ;
Ishibashi, F ;
Fujita, Y .
BIOSCIENCE BIOTECHNOLOGY AND BIOCHEMISTRY, 2005, 69 (11) :2186-2192
[2]  
BLIGH EG, 1959, CAN J BIOCHEM PHYS, V37, P911
[3]   SEPARATION OF LONG-CHAIN FATTY-ACIDS AS PHENACYL ESTERS BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY [J].
BORCH, RF .
ANALYTICAL CHEMISTRY, 1975, 47 (14) :2437-2439
[4]   Advanced liquid chromatography-mass spectrometry interface based on electron ionization [J].
Cappiello, A. ;
Famiglini, G. ;
Pierini, E. ;
Palma, P. ;
Trufelli, H. .
ANALYTICAL CHEMISTRY, 2007, 79 (14) :5364-5372
[5]   Liquid chromatography-electron ionization mass spectrometry: Fields of application and evaluation of the performance of a Direct-EI interface [J].
Cappiello, A ;
Famiglini, G ;
Palma, P ;
Siviero, A .
MASS SPECTROMETRY REVIEWS, 2005, 24 (06) :978-989
[7]   MICRO FLOW-RATE PARTICLE BEAM INTERFACE FOR CAPILLARY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
CAPPIELLO, A ;
BRUNER, F .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1281-1287
[8]   A simple approach for coupling liquid chromatography and electron ionization mass spectrometry [J].
Cappiello, A ;
Famiglini, G ;
Mangani, F ;
Palma, P .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2002, 13 (03) :265-273
[9]   New trends in the application of electron ionization to liquid chromatography-mass spectrometry interfacing [J].
Cappiello, A ;
Famiglini, G ;
Mangani, F ;
Palma, P .
MASS SPECTROMETRY REVIEWS, 2001, 20 (02) :88-104
[10]   Overcoming Matrix Effects in Liquid Chromatography-Mass Spectrometry [J].
Cappiello, Achille ;
Famiglini, Giorgio ;
Palma, Pierangela ;
Pierini, Elisabetta ;
Termopoli, Veronica ;
Trufelli, Helga .