Effects of injection rate and dose on image quality in time-resolved magnetic resonance angiography (MRA) by using 1.0M contrast agents

被引:22
作者
Kramer, Harald
Michaely, Henrik J.
Requardt, Martin
Rohrer, Martin
Reeder, Scott
Reiser, Maximilian F.
Schoenberg, Stefan O.
机构
[1] Univ Munich, Univ Hosp Munich, D-81377 Munich, Germany
[2] Univ Munich, Univ Hosp Munich, Inst Clin Radiol, Munich, Germany
[3] Siemens Med Solut, Erlangen, Germany
[4] Schering AG, D-1000 Berlin, Germany
[5] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
关键词
MRA; time-resolved contrast-enhanced MRA; injection rate;
D O I
10.1007/s00330-006-0493-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In time-resolved MRA (TR MRA), injection parameters and contrast agent (CA) dose are important factors influencing image quality. In this study, three different injection schemes with different CA volumes were evaluated in 12 healthy volunteers. Injection rates between 0.2 and 0.8 ml/s were evaluated with CA volumes of 10 and 20 ml. To measure circulatory parameters, cine cardiac MRI was performed before each exam. Spatial resolution could be reduced to 2x1.4x2 mm(3), temporal resolution was 2.25 s/frame. To exclude signal saturation at high CA concentrations, a phantom with fixed CA concentrations was placed in the field of view. SNR was measured, and the area under the curve of the arterial signal of the different injection schemes was calculated. Results showed the largest diagnostic window at a relatively slow injection rate of 0.4 ml/s and a CA volume of 10 ml. Circulatory parameters have an important impact on CA arrival, so delay times have to be set depending on these parameters.
引用
收藏
页码:1394 / 1402
页数:9
相关论文
共 29 条
[1]   Normal human left and right ventricular dimensions for MRI as assessed by turbo gradient echo and steady-state free precession imaging sequences [J].
Alfakih, K ;
Plein, S ;
Thiele, H ;
Jones, T ;
Ridgway, JP ;
Sivananthan, MU .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (03) :323-329
[2]   The effect of injection rate on time-resolved contrast-enhanced peripheral MRA [J].
Carroll, TJ ;
Korosec, FR ;
Swan, JS ;
Hany, TF ;
Grist, TM ;
Mistretta, CA .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (04) :401-410
[3]   Signal-to-noise measurements in magnitude images from NMR phased arrays [J].
Constantinides, CD ;
Atalar, E ;
McVeigh, ER .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (05) :852-857
[4]   Contrast-enhanced peripheral magnetic resonance angiography using time-resolved vastly undersampled isotropic projection reconstruction [J].
Du, J ;
Carroll, TJ ;
Brodsky, E ;
Lu, AM ;
Grist, TM ;
Mistretta, CA ;
Block, WF .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (05) :894-900
[5]   SNR improvement for multiinjection time-resolved high-resolution CE-MRA of the peripheral vasculature [J].
Du, J ;
Carroll, TJ ;
Block, WF ;
Fain, SB ;
Korosec, FR ;
Grist, TM ;
Mistretta, CA .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (05) :909-917
[6]  
Fink C, 2005, INVEST RADIOL, V40, P40
[7]   Thorax: Low-dose contrast-enhanced three-dimensional MR angiography with subsecond temporal resolution - Initial results [J].
Finn, JP ;
Baskaran, V ;
Carr, JC ;
McCarthy, RM ;
Pereles, FS ;
Kroeker, R ;
Laub, GA .
RADIOLOGY, 2002, 224 (03) :896-904
[8]   Peripheral run-off CE-MRA with a 1.0 molar gadolinium chelate (Gadovist) with Intraarterial DSA comparison [J].
Gregor, M ;
Tombach, B ;
Hentsch, A ;
Reimer, P .
ACADEMIC RADIOLOGY, 2002, 9 :S398-S400
[9]   Accelerated dynamic Fourier velocity encoding by exploiting velocity-spatio-temporal correlations [J].
Hansen, MS ;
Baltes, C ;
Tsao, J ;
Kozerke, S ;
Pruessmann, KP ;
Boesiger, P ;
Pedersen, EM .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2004, 17 (02) :86-94
[10]   Aorta and runoff vessels: Single-injection MR angiography with automated table movement compared with multiinjection time-resolved MR angiography - Initial results [J].
Hany, TF ;
Carroll, TJ ;
Omary, RA ;
Eparza-Coss, E ;
Korosec, FR ;
Mistretta, CA ;
Grist, TM .
RADIOLOGY, 2001, 221 (01) :266-272