Highly cross-linked polyethylene in primary total knee arthroplasty is associated with a lower rate of revision for aseptic loosening: a meta-analysis of 962,467 cases

被引:21
作者
Gkiatas, Ioannis [1 ]
Karasavvidis, Theofilos [2 ]
Sharma, Abhinav K. [3 ]
Xiang, William [1 ]
Malahias, Michael-Alexander [1 ]
Chalmers, Brian P. [3 ]
Sculco, Peter K. [1 ]
机构
[1] Hosp Special Surg, Stavros Niarchos Fdn, Complex Joint Reconstruct Ctr, New York, NY 10022 USA
[2] Aristotle Univ Thessaloniki, Fac Hlth Sci, Sch Med, Thessaloniki, Greece
[3] Hosp Special Surg, Adult Reconstruct & Joint Replacement, New York, NY 10022 USA
关键词
Total knee arthroplasty; Liner; Polyethylene; Highly cross linked; MOLECULAR-WEIGHT POLYETHYLENE; CONVENTIONAL POLYETHYLENE; JOINT REPLACEMENT; MECHANICAL-PROPERTIES; RADIOLUCENT LINES; WEAR PERFORMANCE; YOUNG-PATIENTS; TIBIAL POST; FOLLOW-UP; TOTAL HIP;
D O I
10.1007/s00402-021-03887-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background The evolution in total knee arthroplasty (TKA) includes the highly cross-linked polyethylene (HXLPE) which has been reported as an effective manner to reduce the wear of the polyethylene and the osteolysis. The purpose of the present study is to synthesize the results of comparative studies between HXLPE and conventional polyethylenes and determine their effect in primary TKA. Methods The US National Library of Medicine (PubMed/MEDLINE) and the Cochrane Database of Systematic Reviews were queried for publications utilizing the following keywords: "cross-linked", "polyethylene", "HXLPE", "conventional", "total knee arthroplasty", "TKA", "total knee replacement" and "TKR" combined with Boolean operators AND and OR. Results Ten studies met the inclusion criteria and were included in the present meta-analysis with 962,467 patients. No significant difference was found regarding the revision rate for any reason between the patients who received HXLPE and those with conventional liner (OR 0.67; 95% CI 0.39-1.18; I-2: 97.7%). In addition, there was no difference regarding the radiolucent lines between the two types of liners (OR 0.54; 95% CI 0.20-1.49; I-2: 69.4%). However, with data coming from seven studies enrolling a total of 411,543 patients, it was demonstrated that patients who received HXLPE were less likely to be revised due to aseptic loosening compared to the patients with conventional liners (OR 0.35; 95% CI 0.31-0.39; I-2: 0.0%). Conclusion The present meta-analysis showed that regarding the overall revision rate and radiographic outcomes there was no significant difference between the two types of liners. On the other hand, the significantly less revision rate due to loosening supports the routine continued use of HXLPE in primary TKA.
引用
收藏
页码:1177 / 1184
页数:8
相关论文
共 49 条
[1]   Aseptic loosening of total joint replacements: mechanisms underlying osteolysis and potential therapies [J].
Abu-Amer, Yousef ;
Darwech, Isra ;
Clohisy, John C. .
ARTHRITIS RESEARCH & THERAPY, 2007, 9 (Suppl 1)
[2]   Dose effects of cross-linking polyethylene for total knee arthroplasty on wear performance and mechanical properties [J].
Asano, Taiyo ;
Akagi, Masao ;
Clarke, Ian C. ;
Masuda, Shingo ;
Ishii, Tsunehiro ;
Nakamura, Takashi .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 83B (02) :615-622
[3]   Is Cross-Linked Polyethylene an Improvement Over Conventional Ultra-High Molecular Weight Polyethylene in Total Knee Arthroplasty? [J].
Boyer, Bertrand ;
Bordini, Barbara ;
Caputo, Dalila ;
Neri, Thomas ;
Stea, Susanna ;
Toni, Aldo .
JOURNAL OF ARTHROPLASTY, 2018, 33 (03) :908-914
[4]   Ultra-High Molecular Weight Polyethylene: Influence of the Chemical, Physical and Mechanical Properties on the Wear Behavior. A Review [J].
Bracco, Pierangiola ;
Bellare, Anuj ;
Bistolfi, Alessandro ;
Affatato, Saverio .
MATERIALS, 2017, 10 (07)
[5]   Fatigue crack propagation resistance of highly crosslinked polyethylene [J].
Bradford, L ;
Baker, D ;
Ries, MD ;
Pruitt, LA .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) :68-72
[6]   The 2012 John Charnley Award: Clinical Multicenter Studies of the Wear Performance of Highly Crosslinked Remelted Polyethylene in THA [J].
Bragdon, Charles R. ;
Doerner, Michael ;
Martell, John ;
Jarrett, Bryan ;
Palm, Henrik ;
Malchau, Henrik .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (02) :393-402
[7]   Update on the etiology of revision TKA - Evident trends in a retrospective survey of 1449 cases [J].
Calliess, Tilman ;
Ettinger, Max ;
Huelsmann, Natalie ;
Ostermeier, Sven ;
Windhagen, Henning .
KNEE, 2015, 22 (03) :174-179
[8]   Gamma irradiation alters fatigue-crack behavior and fracture toughness in 1900H and GUR 1050 UHMWPE [J].
Cole, JC ;
Lemons, JE ;
Eberhardt, AW .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 63 (05) :559-566
[9]   Lower prosthesis-specific 10-year revision rate with cross-linked than with non-crosslinked polyethylene in primary total knee arthroplasty 386,104 procedures from the Australian Orthopaedic Association National Joint Replacement Registry [J].
de Steiger, Richard N. ;
Muratoglu, Orhun ;
Lorimer, Michelle ;
Cuthbert, Alana R. ;
Graves, Stephen E. .
ACTA ORTHOPAEDICA, 2015, 86 (06) :721-727
[10]   Five cases of tibial post fracture in posterior stabilized total knee arthroplasty using Prolong highly cross-linked polyethylene [J].
Diamond, Owen J. ;
Howard, Lisa ;
Masri, Bassam .
KNEE, 2018, 25 (04) :657-662