Predictors of lead break during transvenous lead extraction

被引:17
作者
Morita, Junji [1 ]
Yamaji, Kyohei [1 ]
Nagashima, Michio [1 ]
Kondo, Yusuke [2 ]
Sadohara, Yohei [1 ]
Hirokami, Jun [1 ]
Kuji, Rei [1 ]
Korai, Kengo [1 ]
Fukunaga, Masato [1 ]
Hiroshima, Kenichi [1 ]
Ando, Kenji [1 ]
Goya, Masahiko [3 ]
机构
[1] Kokura Mem Hosp, Dept Cardiol, Kitakyushu, Fukuoka, Japan
[2] Chiba Univ, Grad Sch Med, Dept Cardiovasc Med, Chiba, Japan
[3] Tokyo Med & Dent Univ, Dept Cardiovasc Med, Tokyo, Japan
关键词
coradial lead; lead break; pacemaker lead; passive lead; transvenous lead extraction;
D O I
10.1002/joa3.12524
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The incidence, predictors, and clinical impact of lead break during transvenous lead extraction (TLE) were previously unknown. Methods We included consecutive patients who underwent TLE between September 2013 and July 2019 at our institute. Lead break during removal was defined as lead stretching and becoming misshapen, as assessed by fluoroscopy. Results A total of 246 patients underwent TLE for 501 leads. At a patient level, complete success was achieved in 226 patients (91.9%). At a lead level, 481 leads (96.0%) were completely removed and 101 leads (20.1%) were broken during the procedure. Of 392 identified pacemaker leads, 71 (18.3%) were broken during the TLE procedure. A multivariable analysis confirmed high lead age (odds ratio [OR] 1.12, 95% confidence interval (CI) 1.07-1.17; P < .001), passive leads (OR 2.29 95% CI 1.09-4.80; P = .028), coradial leads (OR 3.45 95% CI 1.72-6.92; P < .001), and insulators made of nonpolyurethane (OR 2.38 95% CI 1.03-5.26; P = .04) as predictors of lead break. Broken leads needed longer procedure times and were associated with a higher rate of cardiac tamponade. Conclusions Lead age, coradial bipolar leads, passive leads, and leads without polyurethane insulation were predictors of lead break and could increase the difficulty of lead extraction.
引用
收藏
页码:645 / 652
页数:8
相关论文
共 21 条
[1]   2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS [J].
Bongiorni, Maria G. ;
Burri, Haran ;
Deharo, Jean C. ;
Starck, Christoph ;
Kennergren, Charles ;
Saghy, Laszlo ;
Rao, Archana ;
Tascini, Carlo ;
Lever, Nigel ;
Kutarski, Andrzej ;
Lozano, Ignacio Fernandez ;
Strathmore, Neil ;
Costa, Roberto ;
Epstein, Laurence ;
Love, Charles ;
Blomstrom-Lundqvist, Carina ;
Fauchier, Laurent ;
Defaye, Pascal ;
Arnar, David O. ;
Klug, Didier ;
Boveda, Serge ;
Nielsen, Jens Cosedis ;
Boriani, Giuseppe ;
Zhang, Shu ;
Martin, Andrew Paul ;
Prutkin, Jordan M. ;
de Zuloaga, Claudio .
EUROPACE, 2018, 20 (07) :1217-+
[2]   Current practice in transvenous lead extraction: a European Heart Rhythm Association EP Network Survey [J].
Bongiorni, Maria Grazia ;
Blomstrom-Lundqvist, Carina ;
Kennergren, Charles ;
Dagres, Nikolaos ;
Pison, Laurent ;
Svendsen, Jesper Hastrup ;
Auricchio, Angelo .
EUROPACE, 2012, 14 (06) :783-786
[3]   Extraction of Old Pacemaker or Cardioverter-Defibrillator Leads by Laser Sheath Versus Femoral Approach [J].
Bordachar, Pierre ;
Defaye, Pascal ;
Peyrouse, Eric ;
Boveda, Serge ;
Mokrani, Bilel ;
Marquie, Christelle ;
Barandon, Laurent ;
Fossaert, Emilie Marcant ;
Garrigue, Stephane ;
Reuter, Sylvain ;
Laborderie, Julien ;
Marijon, Eloi ;
Deharo, Jean-Claude ;
Jacon, Peggy ;
Kacet, Salem ;
Ploux, Sylvain ;
Deplagne, Antoine ;
Haissaguerre, Michel ;
Clementy, Jacques ;
Ritter, Philippe ;
Klug, Didier .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (04) :319-323
[4]   Lead Extraction in Pediatric and Congenital Heart Disease Patients [J].
Cecchin, Frank ;
Atallah, Joseph ;
Walsh, Edward P. ;
Triedman, John K. ;
Alexander, Mark E. ;
Berul, Charles I. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2010, 3 (05) :437-444
[5]   Trans-venous lead removal without the use of extraction sheaths, results of &gt;250 removal procedures [J].
de Bie, Mihaly K. ;
Fouad, Doaa A. ;
Borleffs, C. Jan Willem ;
van Rees, Johannes B. ;
Thijssen, Joep ;
Trines, Serge A. ;
Bootsma, Marianne ;
Schalij, Martin J. ;
van Erven, Lieselot .
EUROPACE, 2012, 14 (01) :112-116
[6]   Gross and microscopic pathological changes associated with nonthoracotomy implantable defibrillator leads [J].
Epstein, AE ;
Kay, GN ;
Plumb, VJ ;
Dailey, SM ;
Anderson, PG .
CIRCULATION, 1998, 98 (15) :1517-1524
[7]   Thin co-radial bipolar leads: technology and clinical performanceDünne koradiale bipolare Elektroden: Technologie und klinische Ergebnisse [J].
T. Fahraeus ;
C. W. Israel ;
M. Wöllenstein .
Herzschrittmachertherapie + Elektrophysiologie, 2001, 12 (3) :148-157
[8]   Lead extractions in patients with cardiac implantable electronic device infections: Single center experience [J].
Goya, Masahiko ;
Nagashima, Michio ;
Hiroshima, Ken-ichi ;
Hayashi, Kentaro ;
Makihara, Yu ;
Fukunaga, Masato ;
An, Yoshimori ;
Ohe, Masatsugu ;
Yamazato, So-ichiro ;
Sonoda, Ko-ichiro ;
Yamashita, Kennosuke ;
Katayama, Kouji ;
Ito, Tomoaki ;
Niu, Harushi ;
Ando, Kenji ;
Yokoi, Hiroyoshi ;
Iwabuchi, Masashi .
JOURNAL OF ARRHYTHMIA, 2016, 32 (04) :308-312
[9]   Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009 Increasing Complexity of Patients and Procedures [J].
Greenspon, Arnold J. ;
Patel, Jasmine D. ;
Lau, Edmund ;
Ochoa, Jorge A. ;
Frisch, Daniel R. ;
Ho, Reginald T. ;
Pavri, Behzad B. ;
Kurtz, Steven M. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (16) :1540-1545
[10]   Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction [J].
Hauser, Robert G. ;
Katsiyiannis, William T. ;
Gornick, Charles C. ;
Almquist, Adrian K. ;
Kallinen, Linda M. .
EUROPACE, 2010, 12 (03) :395-401