Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log files

被引:49
作者
Qian, Jianguo [1 ]
Lee, Louis [1 ]
Liu, Wu [1 ]
Chu, Karen [1 ]
Mok, Edward [1 ]
Luxton, Gary [1 ]
Le, Quynh-Thu [1 ]
Xing, Lei [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Sch Med, Stanford, CA 94305 USA
基金
美国国家科学基金会;
关键词
QUALITY-ASSURANCE; RADIATION-THERAPY; SCATTER CORRECTION; PROSTATE; IMRT; RADIOTHERAPY;
D O I
10.1088/0031-9155/55/13/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Volumetric modulated arc therapy (VMAT) has recently emerged as a new clinical modality for conformal radiation therapy. The aim of this work is to establish a methodology and procedure for retrospectively reconstructing the actual dose delivered in VMAT based on the pre-treatment cone-beam computed tomography (CBCT) and dynamic log files. CBCT was performed before the dose delivery and the system's log files were retrieved after the delivery. Actual delivery at a control point including MLC leaf positions, gantry angles and cumulative monitor units (MUs) was recorded in the log files and the information was extracted using in-house developed software. The extracted information was then embedded into the original treatment DICOM-radiation therapy (RT) file to replace the original control point parameters. This reconstituted DICOM-RT file was imported into the Eclipse treatment planning system (TPS) and dose was computed on the corresponding CBCT. A series of phantom experiments was performed to show the feasibility of dose reconstruction, validate the procedure and demonstrate the efficacy of this methodology. The resultant dose distributions and dose-volume histograms (DVHs) were compared with those of the original treatment plan. The studies indicated that CBCT-based VMAT dose reconstruction is readily achievable and provides a valuable tool for monitoring the dose actually delivered to the tumor target as well as the sensitive structures. In the absence of setup errors, the reconstructed dose shows no significant difference from the original pCT-based plan. It is also elucidated that the proposed method is capable of revealing the dosimetric changes in the presence of setup errors. The method reported here affords an objective means for dosimetric evaluation of VMAT delivery and is useful for adaptive VMAT in future.
引用
收藏
页码:3597 / 3610
页数:14
相关论文
共 32 条
  • [1] Quantification of volumetric and geometric changes occurring during fractionated radiotherapy for head-and-neck cancer using an integrated CT/linear accelerator system
    Barker, JL
    Garden, AS
    Ang, KK
    O'Daniel, JC
    Wang, H
    Court, LE
    Morrison, WH
    Rosenthal, DI
    Chao, KSC
    Tucker, SL
    Mohan, R
    Dong, L
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (04): : 960 - 970
  • [2] COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT)
    Bedford, James L.
    Warrington, Alan P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02): : 537 - 545
  • [3] Automated contour mapping with a regional deformable model
    Chao, Ming
    Li, Tianfang
    Schreibmann, Eduard
    Koong, Albert
    Xing, Lei
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 599 - 608
  • [4] Dose-guided radiation therapy with megavoltage cone-beam CT
    Chen, J.
    Morin, O.
    Aubin, M.
    Bucci, M. K.
    Chuang, C. F.
    Pouliot, J.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2006, 79 : S87 - S98
  • [5] Aperture modulated arc therapy
    Crooks, SM
    Wu, XD
    Takita, C
    Watzich, M
    Xing, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (10) : 1333 - 1344
  • [6] Ezzell G A, 2001, J Appl Clin Med Phys, V2, P138, DOI 10.1120/1.1386508
  • [7] Evidence of limited motion of the prostate by carefully emptying the rectum as assessed by daily MVCT image guidance with helical tomotherapy
    Fiorino, Claudio
    Di Mum, Nadia
    Broggi, Sara
    Cozzarini, Cesare
    Maggiulli, Eleonora
    Alongi, Filippo
    Valdagni, Riccardo
    Fazio, Ferruccio
    Calandrino, Riccardo
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (02): : 611 - 617
  • [8] Quality assurance of volumetric modulated arc therapy using Elekta Synergy
    Haga, Akihiro
    Nakagawa, Keiichi
    Shiraishi, Kenshiro
    Itoh, Saori
    Terahara, Atsuro
    Yamashita, Hideomi
    Ohtomo, Kuni
    Saegusa, Shigeki
    Imae, Toshikazu
    Yoda, Kiyoshi
    Pellegrini, Roberto
    [J]. ACTA ONCOLOGICA, 2009, 48 (08) : 1193 - 1197
  • [9] The use of megavoltage CT (MVCT) images for dose recomputations
    Langen, KM
    Meeks, SL
    Poole, DO
    Wagner, TH
    Willoughby, TR
    Kupelian, PA
    Ruchala, KJ
    Haimerl, J
    Olivera, GH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (18) : 4259 - 4276
  • [10] The use of EPID-measured leaf sequence files for IMRT dose reconstruction in adaptive radiation therapy
    Lee, Louis
    Mao, Weihua
    Xing, Lei
    [J]. MEDICAL PHYSICS, 2008, 35 (11) : 5019 - 5029