A numerical model to reproduce squeaking of ceramic-on-ceramic total hip arthroplasty. Influence of design and material

被引:10
作者
Piriou, P. [1 ]
Ouenzerfi, G. [2 ]
Migaud, H. [3 ]
Renault, E. [4 ]
Massi, F. [2 ]
Serrault, M. [5 ]
机构
[1] Clin Parc, 6 Ave Morvan, F-71400 Autun, France
[2] Inst Natl Sci Appl Lyon INSA, Lab Mecan Contacts & Struct, Batiment Jean dAlembert,18-20,Rue Sci, F-69621 Villeurbanne, France
[3] Univ Lille, Dept Univ Chirurg Orthoped & Traumatol, Pl Verdun, F-59000 Lille, France
[4] Tornier, 161 Rue Lavoisier, F-38334 Montbonnot St Martin, France
[5] Clin St Domin, 99 Rue Messei, F-61100 Flers, France
关键词
Total hip arthroplasty; Squeaking; Finite element analysis; Ceramic bearing; FOLLOW-UP; BRAKE SQUEAL; IN-VITRO; BEARINGS; ARTICULATIONS; FRICTION; IMPLANTS; WEAR;
D O I
10.1016/j.otsr.2016.03.005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Modern ceramic (CoC) bearings for hip arthroplasty (THA) have been used in younger patients who expect improved survivorship. However, audible squeaking produced by the implant is an annoying complication. Previous numerical simulations were not able to accurately reproduce in vitro and in vivo observations. Therefore, we developed a finite element model to: (1) reproduce in vitro squeaking and validate the model by comparing it with in vivo recordings, (2) determine why there are differences between in vivo and in vitro squeaking frequencies, (3) identify the stem's role in this squeaking, (4) predict which designs and materials are more likely to produce squeaking. Hypothesis: A CoC THA numerical model can be developed that reproduces the squeaking frequencies observed in vivo. Material and methods: Numerical methods (finite element analysis [ANSYS]) and experimental methods(using a non-lubricated simulated hip with a cementless 32 mm CoC THA) were developed to reproduce squeaking. Numerical analysis was performed to identify the frequencies that cause vibrations perceivedas an acoustic emission. The finite element analysis (FEA) model was enhanced by adjusting periprosthetic bone and soft tissue elements in order to reproduce the squeaking frequencies recorded in vivo. A numerical method (complex eigenvalue analysis) was used to find the acoustic frequencies of the squeaking noise. The frequencies obtained from the model and the hip simulator were compared to those recorded in vivo. Results: The numerical results were validated by experiments with the laboratory hip simulator. The frequencies obtained (mean 2790Hz with FEA, 2755 Hz with simulator, decreasing to 1759 Hz when bone and soft tissue were included in the FEA) were consistent with those of squeaking hips recorded in vivo (1521 Hz). The cup and ceramic insert were the source of the vibration, but had little influence on the diffusion of the noise required to make the squeaking audible to the human ear. The FEA showed that diffusion of squeaking was due to an unstable vibration of the stem during frictional contact. The FEA predicted a higher rate of squeaking (at a lower coefficient of friction) when TZMF (TM) alloy is used instead of Ti6Al4V and when an anatomic press-fit stem is used instead of straight self-locking designs. Discussion: The current FEA model is reliable; it can be used to assess various stem designs and alloys to predict the different rates of squeaking that certain stems will likely produce. (C) 2016 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:S229 / S234
页数:6
相关论文
共 35 条
[1]   Acoustics of friction [J].
Akay, A .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2002, 111 (04) :1525-1548
[2]   Is alumina-on-alumina ceramic bearings total hip replacement the right choice in patients younger than 50 years of age? A 7-to 15-year follow-up study [J].
Boyer, P. ;
Huten, D. ;
Loriaut, P. ;
Lestrat, V. ;
Jeanrot, C. ;
Massin, P. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2010, 96 (06) :616-622
[3]   Retrieval analysis of squeaking ceramic implants: Are there related specific features? [J].
Chevillotte, C. ;
Trousdale, R. T. ;
An, K. -N. ;
Padgette, D. ;
Wright, T. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2012, 98 (03) :281-287
[4]   Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty [J].
Chevillotte, Christophe ;
Pibarot, Vincent ;
Carret, Jean-Paul ;
Bejui-Hugues, Jacques ;
Guyen, Olivier .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (11) :1599-1604
[5]   Occurrence of noise in alumina-on-alumina total hip arthroplasty. A survey on 284 consecutive hips [J].
Cogan, A. ;
Nizard, R. ;
Sedel, L. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2011, 97 (02) :206-210
[6]   A proposed mechanism for squeaking of ceramic-on-ceramic hips [J].
Currier, John H. ;
Anderson, Devon E. ;
Van Citters, Douglas W. .
WEAR, 2010, 269 (11-12) :782-789
[7]   Alumina ceramic bearings for total hip arthroplasty - Five-year results of a prospective randomized study [J].
D'Antonio, J ;
Capello, W ;
Manley, M ;
Naughton, M ;
Sutton, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (436) :164-171
[8]   Analysis of squeaking on ceramic hip endoprosthesis using the complex eigenvalue method [J].
Fan, N. ;
Chen, G. X. ;
Qian, L. M. .
WEAR, 2011, 271 (9-10) :2305-2312
[9]  
Fan N, 2012, J BIOMECH, V45, pS269
[10]  
Feder BJ, 2008, NY TIMES