Blood flow velocity and ultra-filtration velocity measured by CT imaging system inside a densely bundled hollow fiber dialyzer

被引:12
作者
Lu, Junfeng [2 ]
Lu, Wen-Qiang [1 ]
机构
[1] Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
[2] Chinese Acad Sci, Tech Inst Phys & Chem, Beijing 100190, Peoples R China
关键词
Blood flow velocity measurement; Ultra-filtration velocity measurement; Densely bundled hollow fiber dialyzer; CT imaging technology; Dialyzer design; HEMODIALYZERS;
D O I
10.1016/j.ijheatmasstransfer.2010.01.005
中图分类号
O414.1 [热力学];
学科分类号
摘要
Hollow fiber dialyzer is a device used for years in the therapy of hemodialysis treating people with severe kidney problems. in the design of such a device, the measurement of the blood flow velocity and the ultra-filtration velocity inside the dialyzer are very important. And non-Invasive measurement solutions are highly desired during experiment This paper presents a novel non-invasive measurement method to handle this work. The method adopts a CT (Computed Tomography) system to acquire blood motion data inside a densely bundled hollow fiber dialyzer, and then, by adopting a series of image processing methods, the blood flow velocity as well as the ultra-filtration velocity is measured. (C) 2010 Elsevier Ltd All rights reserved.
引用
收藏
页码:1844 / 1850
页数:7
相关论文
共 8 条
[1]  
Gonzalez R. C., 2004, Digital image processing using MATLAB, VSecond
[2]   Measurement of hollow fiber membrane transport properties in hemodialyzers [J].
Liao, ZJ ;
Klein, E ;
Poh, CK ;
Huang, ZP ;
Lu, JF ;
Hardy, PA ;
Gao, DY .
JOURNAL OF MEMBRANE SCIENCE, 2005, 256 (1-2) :176-183
[3]  
LU JF, 2005, THESIS U KENTUCKY LE
[4]  
LU JF, 2009, 6 INT S MULT FLOW HE
[5]   Effect of flow baffles on the dialysate flow distribution of hollow-fiber hemodialyzers: A nonintrusive experimental study using MRI [J].
Poh, CK ;
Hardy, PA ;
Liao, ZJ ;
Huang, ZP ;
Clark, WR ;
Gao, DY .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2003, 125 (04) :481-489
[6]   Effect of spacer yarns on the dialysate flow distribution of hemodialyzers: A magnetic resonance imaging study [J].
Poh, CK ;
Hardy, PA ;
Liao, ZJ ;
Huang, ZP ;
Clark, WR ;
Gao, DY .
ASAIO JOURNAL, 2003, 49 (04) :440-448
[7]  
POH CK, 2006, ARTIF ENCY BIOMATER, DOI DOI 10.1081/E-EBBE-120007344
[8]   Multi-template GAT correlation for character recognition with a limited quantity of data [J].
Wakahara, T .
EIGHTH INTERNATIONAL CONFERENCE ON DOCUMENT ANALYSIS AND RECOGNITION, VOLS 1 AND 2, PROCEEDINGS, 2005, :824-828