Real-time measurement and correction of both B0 changes and subject motion in diffusion tensor imaging using a double volumetric navigated (DvNav) sequence

被引:32
作者
Alhamud, A. [1 ]
Taylor, Paul A. [1 ,2 ]
van der Kouwe, Andre J. W. [3 ,4 ]
Meintjes, Ernesta M. [1 ]
机构
[1] Univ Cape Town, Dept Human Biol, MRC UCT Med Imaging Res Unit, ZA-7700 Rondebosch, South Africa
[2] AIMS, Muizenberg, South Africa
[3] Massachusetts Gen Hosp, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA USA
[4] Harvard Univ, Sch Med, Dept Radiol, Brookline, MA USA
关键词
Diffusion tensor imaging (DTI); Zero-order shim (frequency); The first-order shim (linear gradients); Navigated diffusion sequence (vNav); Double volumetric navigators (DvNav); B0; correction; ECHO-PLANAR IMAGES; INDUCED DISTORTION; FRACTIONAL ANISOTROPY; ARTIFACT REDUCTION; NORMAL BRAIN; CHILDREN; SPECTROSCOPY; MATURATION; CHILDHOOD; FREQUENCY;
D O I
10.1016/j.neuroimage.2015.11.022
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Diffusion tensor imaging (DTI) requires a set of diffusion weighted measurements in order to acquire enough information to characterize local structure. The MRI scanner automatically performs a shimming process by acquiring a field map before the start of a DTI scan. Changes in B0, which can occur throughout the DTI acquisition due to several factors (including heating of the iron shim coils or subject motion), cause significant signal distortions that result in warped diffusion tensor (DT) parameter estimates. In this work we introduce a novel technique to simultaneously measure, report and correct in real time subject motion and changes in B0 field homogeneity, both in and through the imaging plane. This is achieved using double volumetric navigators (DvNav), i.e. a pair of 3D EPI acquisitions, interleaved with the DTI pulse sequence. Changes in the B0 field are evaluated in terms of zero-order (frequency) and first-order (linear gradients) shim. The ability of the DvNav to accurately estimate the shim parameters was first validated in a water phantom. Two healthy subjects were scanned both in the presence and absence of motion using standard, motion corrected (single navigator, vNav), and DvNav DTI sequences. The difference in performance between the proposed 3D EPI field maps and the standard 3D gradient echo field maps of the MRI scanner was also evaluated in a phantom and two healthy subjects. The DvNav sequence was shown to accurately measure and correct changes in B0 following manual adjustments of the scanner's central frequency and the linear shim gradients. Compared to other methods, the DvNav produced DTI results that showed greater spatial overlap with anatomical references, particularly in scans with subject motion. This is largely due to the ability of the DvNav system to correct shim changes and subject motion between each volume acquisition, thus reducing shear distortion. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:60 / 71
页数:12
相关论文
共 61 条
  • [11] Real-time RF pulse adjustment for B0 drift correction
    Benner, Thomas
    van der Kouwe, Andre J. W.
    Kirsch, John E.
    Sorensen, A. Gregory
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2006, 56 (01) : 204 - 209
  • [12] Diffusion Imaging With Prospective Motion Correction and Reacquisition
    Benner, Thomas
    van der Kouwe, Andre J. W.
    Sorensen, A. Gregory
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (01) : 154 - 167
  • [13] Advances in functional neuroanatomy: A review of combined DTI and fMRI studies in healthy younger and older adults
    Bennett, Ilana J.
    Rypma, Bart
    [J]. NEUROSCIENCE AND BIOBEHAVIORAL REVIEWS, 2013, 37 (07) : 1201 - 1210
  • [14] Bhat H., 2012, P 19 ANN M ISMRM MAY
  • [15] Bhat H., 2014, ISMRM WORKSH MOT COR
  • [16] Ghost artifact reduction for echo planar imaging using image phase correction
    Buonocore, MH
    Gao, LS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (01) : 89 - 100
  • [17] AFNI: Software for analysis and visualization of functional magnetic resonance neuroimages
    Cox, RW
    [J]. COMPUTERS AND BIOMEDICAL RESEARCH, 1996, 29 (03): : 162 - 173
  • [18] Monitoring and correcting spatio-temporal variations of the MR scanner's static magnetic field
    El-Sharkawy, Abdel Monem
    Schar, Michael
    Bottomley, Paul A.
    Atalar, Ergin
    [J]. MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2006, 19 (05) : 223 - 236
  • [19] Diffusion-weighted MR imaging in the brain in children:: Findings in the normal brain and in the brain with white matter diseases
    Engelbrecht, V
    Scherer, A
    Rassek, M
    Witsack, HJ
    Mödder, U
    [J]. RADIOLOGY, 2002, 222 (02) : 410 - 418
  • [20] Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging
    Foerster, BU
    Tomasi, D
    Caparelli, EC
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (05) : 1261 - 1267