Real-time measurement and correction of both B0 changes and subject motion in diffusion tensor imaging using a double volumetric navigated (DvNav) sequence

被引:32
作者
Alhamud, A. [1 ]
Taylor, Paul A. [1 ,2 ]
van der Kouwe, Andre J. W. [3 ,4 ]
Meintjes, Ernesta M. [1 ]
机构
[1] Univ Cape Town, Dept Human Biol, MRC UCT Med Imaging Res Unit, ZA-7700 Rondebosch, South Africa
[2] AIMS, Muizenberg, South Africa
[3] Massachusetts Gen Hosp, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA USA
[4] Harvard Univ, Sch Med, Dept Radiol, Brookline, MA USA
关键词
Diffusion tensor imaging (DTI); Zero-order shim (frequency); The first-order shim (linear gradients); Navigated diffusion sequence (vNav); Double volumetric navigators (DvNav); B0; correction; ECHO-PLANAR IMAGES; INDUCED DISTORTION; FRACTIONAL ANISOTROPY; ARTIFACT REDUCTION; NORMAL BRAIN; CHILDREN; SPECTROSCOPY; MATURATION; CHILDHOOD; FREQUENCY;
D O I
10.1016/j.neuroimage.2015.11.022
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Diffusion tensor imaging (DTI) requires a set of diffusion weighted measurements in order to acquire enough information to characterize local structure. The MRI scanner automatically performs a shimming process by acquiring a field map before the start of a DTI scan. Changes in B0, which can occur throughout the DTI acquisition due to several factors (including heating of the iron shim coils or subject motion), cause significant signal distortions that result in warped diffusion tensor (DT) parameter estimates. In this work we introduce a novel technique to simultaneously measure, report and correct in real time subject motion and changes in B0 field homogeneity, both in and through the imaging plane. This is achieved using double volumetric navigators (DvNav), i.e. a pair of 3D EPI acquisitions, interleaved with the DTI pulse sequence. Changes in the B0 field are evaluated in terms of zero-order (frequency) and first-order (linear gradients) shim. The ability of the DvNav to accurately estimate the shim parameters was first validated in a water phantom. Two healthy subjects were scanned both in the presence and absence of motion using standard, motion corrected (single navigator, vNav), and DvNav DTI sequences. The difference in performance between the proposed 3D EPI field maps and the standard 3D gradient echo field maps of the MRI scanner was also evaluated in a phantom and two healthy subjects. The DvNav sequence was shown to accurately measure and correct changes in B0 following manual adjustments of the scanner's central frequency and the linear shim gradients. Compared to other methods, the DvNav produced DTI results that showed greater spatial overlap with anatomical references, particularly in scans with subject motion. This is largely due to the ability of the DvNav system to correct shim changes and subject motion between each volume acquisition, thus reducing shear distortion. (C) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:60 / 71
页数:12
相关论文
共 61 条
  • [1] Real-Time Optical Motion Correction for Diffusion Tensor Imaging
    Aksoy, Murat
    Forman, Christoph
    Straka, Matus
    Skare, Stefan
    Holdsworth, Samantha
    Hornegger, Joachim
    Bammer, Roland
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2011, 66 (02) : 366 - 378
  • [2] Motion Artifact Reduction in Pediatric Diffusion Tensor Imaging Using Fast Prospective Correction
    Alhamud, A.
    Taylor, Paul A.
    Laughton, Barbara
    van der Kouwe, Andre J. W.
    Meintjes, Ernesta M.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2015, 41 (05) : 1353 - 1364
  • [3] Volumetric navigators for real-time motion correction in diffusion tensor imaging
    Alhamud, A.
    Tisdall, M. Dylan
    Hess, Aaron T.
    Hasan, Khader M.
    Meintjes, Ernesta M.
    van der Kouwe, Andre J. W.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (04) : 1097 - 1108
  • [4] Alhamud A., 2011, P 19 ANN M ISMRM MAY
  • [5] How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging
    Andersson, JLR
    Skare, S
    Ashburner, J
    [J]. NEUROIMAGE, 2003, 20 (02) : 870 - 888
  • [6] Assaf Y, 1999, NMR BIOMED, V12, P335, DOI 10.1002/(SICI)1099-1492(199910)12:6<335::AID-NBM581>3.0.CO
  • [7] 2-A
  • [8] Diffusion tensor imaging using single-shot SENSE-EPI
    Bammer, R
    Auer, M
    Keeling, SL
    Augustin, M
    Stables, LA
    Prokesch, RW
    Stollberger, R
    Moseley, ME
    Fazekas, F
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (01) : 128 - 136
  • [9] MR DIFFUSION TENSOR SPECTROSCOPY AND IMAGING
    BASSER, PJ
    MATTIELLO, J
    LEBIHAN, D
    [J]. BIOPHYSICAL JOURNAL, 1994, 66 (01) : 259 - 267
  • [10] Non-invasive mapping of connections between human thalamus and cortex using diffusion imaging
    Behrens, TEJ
    Johansen-Berg, H
    Woolrich, MW
    Smith, SM
    Wheeler-Kingshott, CAM
    Boulby, PA
    Barker, GJ
    Sillery, EL
    Sheehan, K
    Ciccarelli, O
    Thompson, AJ
    Brady, JM
    Matthews, PM
    [J]. NATURE NEUROSCIENCE, 2003, 6 (07) : 750 - 757