Comparison of two bioelectrical impedance devices and dual-energy X-ray absorptiometry to evaluate body composition in heart failure

被引:25
作者
Alves, F. D. [1 ,2 ]
Souza, G. C. [2 ,3 ]
Biolo, A. [1 ,2 ,3 ]
Clausell, N. [1 ,2 ,3 ]
机构
[1] Univ Fed Rio Grande do Sul, Fac Med, Postgrad Program Cardiovasc Sci, Porto Alegre, RS, Brazil
[2] Univ Fed Rio Grande do Sul, Fac Med, Dept Internal Med, Porto Alegre, RS, Brazil
[3] Hosp Clin Porto Alegre, Dept Cardiol, Porto Alegre, RS, Brazil
关键词
bioelectrical impedance; body composition; dual-energy X-ray absorptiometry; heart failure; MASS INDEX; MORTALITY; CHILDREN; DEXA;
D O I
10.1111/jhn.12218
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
Background: The utilisation of bioelectrical impedance analysis (BIA) in heart failure can be affected by many factors and its applicability remains controversial. The present study aimed to verify the adequacy of single-frequency BIA (SF-BIA) and multifrequency BIA (MF-BIA) compared to dual-energy x-ray absorptiometry (DEXA) for evaluating body composition in outpatients with heart failure. Methods: In this cross-sectional study, 55 patients with stable heart failure and left ventricle ejection fraction <= 45% were evaluated for fat mass percentage, fat mass and fat-free mass by DEXA and compared with the results obtained by SF-BIA (single frequency of 50 kHz) and MF-BIA (frequencies of 20 and 100 kHz). Results: MF-BIA and DEXA gave similar mean values for fat mass percentage, fat mass and fat-free mass, whereas values from SF-BIA were significantly different from DEXA. Both SF-BIA and MF-BIA measures of body composition correlated strongly with DEXA (r > 0.8; P < 0.001), except for fat mass assessed by SF-BIA, which showed a moderate correlation (r = 0.760; P < 0.001). MF-BIA also showed a better agreement with DEXA by Bland-Altman analysis in all measurements. However, both types of equipment showed wide limits of agreement and a significant relationship between variance and bias (Pitmans's test P > 0.05), except MF-BIA for fat-free mass. Conclusions: Compared with DEXA, MF-BIA showed better accuracy than SF-BIA, although both types of equipment showed wide limits of agreement. The BIA technique should be used with caution, and regression equations might be useful for correcting the observed variations, mainly in extreme values of body composition.
引用
收藏
页码:632 / 638
页数:7
相关论文
共 27 条
[1]  
Andreoli A, 2009, RADIOL MED, V114, P286, DOI 10.1007/s11547-009-0369-7
[2]   Wasting as independent risk factor for mortality in chronic heart failure [J].
Anker, SD ;
Ponikowski, P ;
Varney, S ;
Chua, TP ;
Clark, AL ;
WebbPeploe, KM ;
Harrington, D ;
Kox, WJ ;
PooleWilson, PA ;
Coats, AJS .
LANCET, 1997, 349 (9058) :1050-1053
[3]  
[Anonymous], Journal of Computers, DOI [DOI 10.1186/1475-2891-7-26, DOI 10.11897/SP.J.1016.2018.01448]
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Assessment of Body Composition in Peritoneal Dialysis Patients Using Bioelectrical Impedance and Dual-Energy X-Ray Absorptiometry [J].
Fuerstenberg, Antje ;
Davenport, Andrew .
AMERICAN JOURNAL OF NEPHROLOGY, 2011, 33 (02) :150-156
[6]   Comparison of Multifrequency Bioelectrical Impedance Analysis and Dual-Energy X-ray Absorptiometry Assessments in Outpatient Hemodialysis Patients [J].
Fuerstenberg, Antje ;
Davenport, Andrew .
AMERICAN JOURNAL OF KIDNEY DISEASES, 2011, 57 (01) :123-129
[7]   Comparison of body composition with bioelectric impedance (BIA) and dual energy X-ray absorptiometry (DEXA) among Singapore Chinese [J].
Gupta, Nidhi ;
Balasekaran, Govindasamy ;
Govindaswamy, Visvasuresh Victor ;
Hwa, Chia Yong ;
Shun, Lim Meng .
JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2011, 14 (01) :33-35
[8]  
Heyward VH, 2004, APPL BODY COMPOSITIO
[9]   Body mass index and prognosis in patients with chronic heart failure - Insights from the Candesartan in Heart failure: Assessment of Reduction in Mortality and morbidity (CHARM) program [J].
Kenchaiah, Satish ;
Pocock, Stuart J. ;
Wang, Duolao ;
Finn, Peter V. ;
Zornoff, Leonardo A. M. ;
Skali, Hicham ;
Pfeffer, Marc A. ;
Yusuf, Salim ;
Swedberg, Karl ;
Michelson, Eric L. ;
Granger, Christopher B. ;
McMurray, John J. V. ;
Solomon, Scott D. .
CIRCULATION, 2007, 116 (06) :627-636
[10]   Bioelectrical impedance analysis -: part II:: utilization in clinical practice [J].
Kyle, UG ;
Bosaeus, I ;
De Lorenzo, AD ;
Deurenberg, P ;
Elia, M ;
Gómez, JM ;
Heitmann, BL ;
Kent-Smith, L ;
Melchior, JC ;
Pirlich, M ;
Scharfetter, H ;
Schols, AMWJ ;
Pichard, C .
CLINICAL NUTRITION, 2004, 23 (06) :1430-1453