Average of Patient Deltas: Patient-Based Quality Control Utilizing the Mean Within-Patient Analyte Variation

被引:13
作者
Cembrowski, George S. [1 ,2 ]
Xu, Qian [3 ]
Cervinski, Mark A. [4 ,5 ]
机构
[1] Univ Alberta, Lab Med & Pathol, Edmonton, AB, Canada
[2] Cembrowski Cembrowski Qual Control Consulting, Edmonton, AB, Canada
[3] Univ Manitoba, Sch Med, Winnipeg, MB, Canada
[4] Geisel Sch Med Dartmouth, Dept Pathol & Lab Med, Hanover, NH USA
[5] Dartmouth Hitchcock Med Ctr, Dept Pathol & Lab Med, One Med Ctr Dr, Lebanon, NH 03756 USA
关键词
HEALTHY-YOUNG MALES; MOVING AVERAGE; BLOOD-GAS; TIME; OPTIMIZATION; BISPEBJERG; IMPLEMENTATION; COMMUTABILITY; PARAMETERS; TOOL;
D O I
10.1093/clinchem/hvab057
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Because traditional QC is discontinuous, laboratories use additional strategies to detect systematic error. One strategy, the delta check, is best suited to detect large systematic error. The moving average (MA) monitors the mean patient analyte value but cannot equitably detect systematic error in skewed distributions. Our study combines delta check and MA to develop an average of deltas (AoD) strategy that monitors the mean delta of consecutive, intrapatient results. METHODS: Arrays of the differences (delta) between paired patient results collected within 20-28 h of each other were generated from historical data. AoD protocols were developed using a simulated annealing algorithm in MatLab (Mathworks) to select the number of patient delta values to average and truncation limits to eliminate large deltas. We simulated systematic error by adding bias to arrays for plasma albumin, alanine aminotransferase, alkaline phosphatase, amylase, aspartate aminotransferase, bicarbonate, bilirubin (total and direct), calcium, chloride, creatinine, lipase, sodium, phosphorus, potassium, total protein, and magnesium. The average number of deltas to detection (ANDED) was then calculated in response to induced systematic error. RESULTS: ANDED varied by combination of assay and AoD protocol. Errors in albumin, lipase, and total protein were detected with a mean of 6 delta pairs. The highest ANDED was calcium, with a positive 0.6-mg/dL shift detected with an ANDED of 75. However, a negative 0.6-mg/dL calcium shift was detected with an ANDED of 25. CONCLUSIONS: AoD detects systematic error with relatively few paired patient samples and is a patient-based QC technique that will enhance error detection.
引用
收藏
页码:1019 / 1029
页数:11
相关论文
共 30 条
[1]   Diurnal rhythm in clinical chemistry: An underrated source of variation [J].
Abou El Hassan, Mohamed ;
Delvin, Edgard ;
Elnenaei, Manal O. ;
Hoffman, Barry .
CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2018, 55 (08) :516-534
[2]   Understanding Patient-Based Real-Time Quality Control Using Simulation Modeling [J].
Bietenbeck, Andreas ;
Cervinski, Mark A. ;
Katayev, Alex ;
Loh, Tze Ping ;
van Rossum, Huub H. ;
Badrick, Tony .
CLINICAL CHEMISTRY, 2020, 66 (08) :1072-1083
[3]   A mechanism-based way to evaluate commutability of control materials for enzymatic measurements. The example of gamma-glutamyltransferase [J].
Carobene, Anna ;
Guerra, Elena ;
Ceriotti, Ferruccio .
CLINICA CHIMICA ACTA, 2013, 424 :153-158
[4]  
Cembrowski G, 2020, J LAB PRECIS MED, V5
[5]   Five-Year Two-Center Retrospective Comparison of Central Laboratory Glucose to GEM 4000 and ABL 800 Blood Glucose: Demonstrating the (In)adequacy of Blood Gas Glucose [J].
Cembrowski, George ;
Jung, Joanna ;
Mei, Junyi ;
Xu, Eric ;
Curic, Tihomir ;
Gibney, Rt Noel ;
Jacka, Michael ;
Sadrzadeh, Hossein .
JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2020, 14 (03) :535-545
[6]   Impaired clinical utility of sequential patient GEM blood gas measurements associated with calibration schedule [J].
Cembrowski, George S. ;
Xu, Qian ;
Cembrowski, Adam R. ;
Mei, Junyi ;
Sadrzadeh, Hossein .
CLINICAL BIOCHEMISTRY, 2017, 50 (16-17) :936-941
[7]   The use of serial patient blood gas, electrolyte and glucose results to derive biologic variation: a new tool to assess the acceptability of intensive care unit testing [J].
Cembrowski, George S. ;
Tran, David V. ;
Higgins, Trefor N. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (10) :1447-1454
[8]  
CEMBROWSKI GS, 1975, CLIN CHEM, V21, P1396
[9]  
CLSI, 2003, US EV REAG LOT VAR US EV REAG LOT VAR, V1st
[10]   Changing the paradigm of laboratory quality control through implementation of real-time test results monitoring: For patients by patients [J].
Fleming, James K. ;
Katayev, Alexander .
CLINICAL BIOCHEMISTRY, 2015, 48 (7-8) :508-513