Enhancing the estimation of blood pressure using pulse arrival time and two confounding factors

被引:43
作者
Baek, Hyun Jae [2 ]
Kim, Ko Keun [3 ]
Kim, Jung Soo [3 ]
Lee, Boreom [4 ]
Park, Kwang Suk [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Seoul, South Korea
[2] Seoul Natl Univ, Grad Program Bioengn, Seoul, South Korea
[3] Seoul Natl Univ, Interdisciplinary Program Biomed Engn, Seoul, South Korea
[4] Seoul Natl Univ, Med Res Ctr, Inst Med & Biol Engn, Seoul, South Korea
关键词
BP; PAT; heart rate; arterial stiffness; multiple-regression analysis; TRANSIT-TIME; CARDIOVASCULAR-DISEASE; RISK;
D O I
10.1088/0967-3334/31/2/002
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A new method of blood pressure (BP) estimation using multiple regression with pulse arrival time (PAT) and two confounding factors was evaluated in clinical and unconstrained monitoring situations. For the first analysis with clinical data, electrocardiogram (ECG), photoplethysmogram (PPG) and invasive BP signals were obtained by a conventional patient monitoring device during surgery. In the second analysis, ECG, PPG and non-invasive BP were measured using systems developed to obtain data under conditions in which the subject was not constrained. To enhance the performance of BP estimation methods, heart rate (HR) and arterial stiffness were considered as confounding factors in regression analysis. The PAT and HR were easily extracted from ECG and PPG signals. For arterial stiffness, the duration from the maximum derivative point to the maximum of the dicrotic notch in the PPG signal, a parameter called TDB, was employed. In two experiments that normally cause BP variation, the correlation between measured BP and the estimated BP was investigated. Multiple-regression analysis with the two confounding factors improved correlation coefficients for diastolic blood pressure and systolic blood pressure to acceptable confidence levels, compared to existing methods that consider PAT only. In addition, reproducibility for the proposed method was determined using constructed test sets. Our results demonstrate that non-invasive, non-intrusive BP estimation can be obtained using methods that can be applied in both clinical and daily healthcare situations.
引用
收藏
页码:145 / 157
页数:13
相关论文
共 23 条
  • [1] Nonintrusive Biological Signal Monitoring in a Car to Evaluate a Driver's Stress and Health State
    Baek, Hyun Jae
    Lee, Haet Bit
    Kim, Jung Soo
    Choi, Jong Min
    Kim, Ko Keun
    Park, Kwang Suk
    [J]. TELEMEDICINE JOURNAL AND E-HEALTH, 2009, 15 (02): : 182 - 189
  • [2] Heart rate-arterial blood pressure relationship in conscious rat before vs. after spinal cord transection
    Baldridge, BR
    Burgess, DE
    Zimmerman, EE
    Carroll, JJ
    Sprinkle, AG
    Speakman, RO
    Li, SG
    Brown, DR
    Taylor, RF
    Dworkin, S
    Randall, DC
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 2002, 283 (03) : R748 - R756
  • [3] COMPARISON OF BRACHIAL AND RADIAL ARTERIAL-PRESSURE MONITORING IN PATIENTS UNDERGOING CORONARY-ARTERY BYPASS-SURGERY
    BAZARAL, MG
    WELCH, M
    GOLDING, LAR
    BADHWAR, K
    [J]. ANESTHESIOLOGY, 1990, 73 (01) : 38 - 45
  • [4] ACUTE RADIAL NERVE INJURY FROM USE OF AN AUTOMATIC BLOOD-PRESSURE MONITOR
    BICKLER, PE
    SCHAPERA, A
    BAINTON, CR
    [J]. ANESTHESIOLOGY, 1990, 73 (01) : 186 - 188
  • [5] ASSESSMENT OF SPONTANEOUS BAROREFLEX SENSITIVITY IN RATS - A NEW METHOD USING THE CONCEPT OF STATISTICAL DEPENDENCE
    CERUTTI, C
    DUCHER, M
    LANTELME, P
    GUSTIN, MP
    PAULTRE, C
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-REGULATORY INTEGRATIVE AND COMPARATIVE PHYSIOLOGY, 1995, 268 (02) : R382 - R388
  • [6] Blood pressure evaluation based on arterial pulse wave velocity
    Franchi, D
    Bedini, R
    Manfredini, F
    Berti, S
    Palagi, G
    Ghione, S
    Ripoli, A
    [J]. COMPUTERS IN CARDIOLOGY 1996, 1996, : 397 - 400
  • [7] PULSE TRANSIT-TIME AS AN INDICATOR OF ARTERIAL BLOOD-PRESSURE
    GEDDES, LA
    VOELZ, MH
    BABBS, CF
    BOURLAND, JD
    TACKER, WA
    [J]. PSYCHOPHYSIOLOGY, 1981, 18 (01) : 71 - 74
  • [8] Ambulatory blood pressure monitoring and risk of cardiovascular disease: A population based study
    Hansen, TW
    Jeppesen, J
    Rasmussen, S
    Ibsen, H
    Torp-Pedersen, C
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2006, 19 (03) : 243 - 250
  • [9] Smoking and atherosclerotic cardiovascular disease in men with low levels of serum cholesterol - The Korea Medical Insurance Corporation Study
    Jee, SH
    Suh, I
    Kim, IS
    Appel, LJ
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 282 (22): : 2149 - 2155
  • [10] A new approach for non-intrusive monitoring of blood pressure on a toilet seat
    Kim, JS
    Chee, YJ
    Park, JW
    Choi, JW
    Park, KS
    [J]. PHYSIOLOGICAL MEASUREMENT, 2006, 27 (02) : 203 - 211