Electrical Impedance of Stainless Steel Needle Electrodes

被引:24
作者
Kalvoy, Havard [1 ,2 ]
Tronstad, Christian [1 ,2 ]
Nordbotten, Bernt [2 ]
Grimnes, Sverre [1 ,2 ]
Martinsen, Orjan G. [1 ,2 ]
机构
[1] Oslo Univ Hosp, Rikshosp, Dept Clin & Biomed Engn, N-0027 Oslo, Norway
[2] Univ Oslo, Dept Phys, Oslo, Norway
关键词
Electrical properties; Medical grade; In vitro; EMG-electrodes; Long-term stability; SEM; TISSUE; POLARIZATION; STIMULATION; DESIGN;
D O I
10.1007/s10439-010-9989-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present experimental findings regarding variability and stability of the electrical impedance properties of medical grade stainless steel needle electrodes in vitro. Monopolar impedance spectra (1 Hz to 1 MHz) were measured and scanning electron microscope images were obtained for five needle types with active electrode area from 0.28 to 0.7 mm(2). A saline tank (0.9% NaCl) was used as tissue model. Measurements were done before and after electrolytic treatment with anodic and cathodic DC currents of 1 mu A. With active electrode areas below 1 mm(2), high influence from electrode polarization impedance (EPI) was expected at low frequencies (LF). For higher frequencies (HF) the EPI decreases and the impedance of the surrounding tissue is more pronounced. The hypothesis tested was that the EPI at LF would depend upon contact area, alloy composition, surface structure, and treatment of the active electrode, and at HF upon the electrode area geometry, and the specific resistivity of saline. Our results show large differences in electrical properties between needle types. After electrolytic treatment the EPI decreased. After 5-48 h of saline exposure the EPI increased, both for treated and untreated needles. Cathodic treatment gave lower impedance and drift than anodic or no treatment.
引用
收藏
页码:2371 / 2382
页数:12
相关论文
共 28 条
[1]   Reduction of the contribution of electrode polarization effects in the radiowave dielectric measurements of highly conductive biological cell suspensions [J].
Bordi, F ;
Cametti, C ;
Gili, T .
BIOELECTROCHEMISTRY, 2001, 54 (01) :53-61
[2]   ELECTRICAL-STIMULATION WITH PT ELECTRODES .1. METHOD FOR DETERMINATION OF REAL ELECTRODE AREAS [J].
BRUMMER, SB ;
TURNER, MJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1977, 24 (05) :436-439
[3]  
Buchthal F., 1966, Brain Res, V3, P1, DOI 10.1016/0006-8993(66)90056-4
[4]   Sources and effects of electrode impedance during deep brain stimulation [J].
Butson, CR ;
Maks, CB ;
McIntyre, CC .
CLINICAL NEUROPHYSIOLOGY, 2006, 117 (02) :447-454
[5]  
Cooper R., 1946, J I ELECT ENG LOND, V93, P69
[6]   ELECTRICAL-PROPERTIES OF COMMERCIAL CONCENTRIC EMG ELECTRODES [J].
DORFMAN, LJ ;
MCGILL, KC ;
CUMMINS, KL .
MUSCLE & NERVE, 1985, 8 (01) :1-8
[7]  
Fricke R, 1932, PHILOS MAG, V14, P310
[8]   IMPEDANCE OF STAINLESS-STEEL ELECTRODES [J].
GEDDES, LA ;
DACOSTA, CP ;
WISE, G .
MEDICAL & BIOLOGICAL ENGINEERING, 1971, 9 (05) :511-&
[9]  
Gray J.R., 2004, ENV INSTRUMENTATION, P491, DOI [DOI 10.1002/0471473332.CH23, 10.1002/0471473332.CH23]
[10]   IMPEDANCE MEASUREMENT OF INDIVIDUAL SKIN SURFACE ELECTRODES [J].
GRIMNES, S .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1983, 21 (06) :750-755