A palm-sized high-sensitivity near-infrared fluorescence imager for laparotomy surgery

被引:4
作者
Dorval, Paul [1 ,2 ]
Mangeret, Norman [2 ]
Guillermet, Stephanie [2 ]
Atallah, Ihab [3 ]
Righini, Christian [3 ]
Barabino, Gabriele [4 ]
Coll, Jean-Luc [3 ]
Rizo, Philippe [2 ,5 ]
Poulet, Patrick [1 ]
机构
[1] Univ Strasbourg, CNRS, Federat Med Translat Strasbourg, Lab Sci Ingn Informat & Imagerie,UMR 7357, F-67412 Illkirch Graffenstaden, France
[2] Fluoptics, F-38040 Grenoble, France
[3] Univ Grenoble 1, INSERM, U823, Inst Albert Bonniot, F-38000 Grenoble, France
[4] Ctr Hosp Univ St Etienne, F-42000 St Etienne, France
[5] CEA, LETI, Minatec, F-38054 Grenoble, France
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2016年 / 32卷 / 01期
关键词
Fluorescence image-guided surgery; Near-infrared fluorescence; Molecular imaging; Indocyanine green; INDOCYANINE GREEN; CANCER-SURGERY; GUIDED SURGERY; IN-VIVO;
D O I
10.1016/j.ejmp.2015.11.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In laparotomy surgery guided by near-infrared fluorescence imaging, the access to the field of operation is limited by the illumination and/or the imaging field. The side of cavities or organs such as the liver or the heart cannot be examined with the systems available on the market, which are too large and too heavy. In this article, we describe and evaluate a palm sized probe, whose properties, weight, size and sensitivity are adapted for guiding laparotomy surgery. Different experiments have been performed to determine its main characteristics, both on the illumination and imaging sides. The device has been tested for fluorescent molecular probe imaging in preclinical procedures, to prove its ability to be used in cancer nodule detection during surgery. This system is now CE certified for clinical procedures and Indocyanine Green imaging has been performed during clinical investigations: lymphedema and surgical resection of liver metastases of colorectal cancers. (C) 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:218 / 225
页数:8
相关论文
共 22 条
[21]   Design and characterization of an optimized simultaneous color and near-infrared fluorescence rigid endoscopic imaging system [J].
Venugopal, Vivek ;
Park, Minho ;
Ashitate, Yoshitomo ;
Neacsu, Florin ;
Kettenring, Frank ;
Frangioni, John V. ;
Gangadharan, Sidhu P. ;
Gioux, Sylvain .
JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (12)
[22]   Compact instrument for fluorescence image-guided surgery [J].
Wang, Xinghua ;
Bhaumik, Srabani ;
Li, Qing ;
Staudinger, V. Paul ;
Yazdanfar, Siavash .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (02)