Intrasac pressure waveforms after endovascular aneurysm repair (EVAR) are a reliable marker of type I endoleaks, but not type II or combined types: An experimental study

被引:25
作者
Chaudhuri, A
Ansdell, LE
Grass, AJ
Adiseshiah, M
机构
[1] UCL Hosp, Vasc Endovasc Unit, London, England
[2] UCL, Dept Civil Engn, London, England
关键词
pressure waveforms; thrombus; negative interference; sac decompression;
D O I
10.1016/j.ejvs.2004.07.006
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose. To ascertain the nature of the pressure wave transmitted through aneurysm thrombus and the changes produced after endovascular repair and the development of type I and II endoleaks. Methods. A 25 mm Talent endovascular graft was deployed in a latex model of an abdominal aortic aneurysm, which was incorporated in a pulsatile flow unit. The graft was surrounded by thrombus analogue to simulate conditions in vivo. Pressure waveforms in the sac were captured over 5 s at 1000 Hz in these settings: W no endoleaks (baseline), after introduction of (ii) type I (iii) type II and (iv) combined type I and II endoleaks. The arterial blood pressure settings used were 140/100 and 130/90 mmHg, denoted the high and low settings, respectively. ANOVA in Minitab 13 was applied for statistical analysis. Results. Pulsatile waveforms were transmitted through the thrombus. Intrasac pressure after stent-grafting reduced to 110/107, 99/96 mmHg (p < 0.001) (high, low settings, respectively). Introduction of a type I endoleak caused this to rise to 120/112, 115/107 mmHg (p < 0.001, vs. baseline); after producing a type II endoleak these were 101/98, 91/88 mmHg (P < 0.001, vs. baseline). A combined type I and II endoleak produced intrasac pressures identical to that of a type I endoleak. Conclusions. Intrasac pressure waveforms following EVAR are easily defined following a type I endoleak. Waveforms obtained following type II endoleak simulation resemble the baseline waveform in an attenuated form. Intrasac pressures are, therefore, a reliable marker for type I, but not a type II endoleak. In the case of a combined endoleak, the type I endoleak waveform effectively masks that of the type II. Intrasac thrombus faithfully transmits intrasac pressures.
引用
收藏
页码:373 / 378
页数:6
相关论文
共 16 条
[1]  
Chaudhuri A, 2003, J ENDOVASC THER, V10, P1097, DOI 10.1583/1545-1550(2003)010<1097:NLAAAM>2.0.CO
[2]  
2
[3]  
Flora HS, 2002, J ENDOVASC THER, V9, P665, DOI 10.1583/1545-1550(2002)009<0665:AWSATT>2.0.CO
[4]  
2
[5]  
Maldonado Thomas S, 2003, Vasc Endovascular Surg, V37, P1, DOI 10.1177/153857440303700101
[6]   Reduction in aortic aneurysm size: Early results after endovascular graft placement [J].
Matsumura, JS ;
Pearce, WH ;
McCarthy, WJ ;
Yao, JST .
JOURNAL OF VASCULAR SURGERY, 1997, 25 (01) :113-123
[7]   MEASUREMENT OF RADIOACTIVE MICROSPHERES IN BIOLOGICAL SAMPLES [J].
MERNAGH, JR ;
SPIERS, EW ;
ADISESHIAH, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (04) :646-650
[8]   Intra-aneurysmal pressure after incomplete endovascular exclusion [J].
Parodi, JC ;
Berguer, R ;
Ferreira, LM ;
La Mura, R ;
Schermerhorn, ML .
JOURNAL OF VASCULAR SURGERY, 2001, 34 (05) :909-914
[9]  
Schurink GWH, 2000, J VASC SURG, V31, P501
[10]   Endoleakage after stent-graft treatment of abdominal aneurysm: Implications on pressure and imaging - an in vitro study [J].
Schurink, GWH ;
Aarts, NJM ;
Wilde, J ;
van Baalen, JM ;
Chuter, TAM ;
Kool, LJ ;
van Bockel, JH .
JOURNAL OF VASCULAR SURGERY, 1998, 28 (02) :234-241