Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR

被引:12
作者
Hippelainen, Eero [1 ,2 ]
Makela, Teemu [2 ,3 ]
Kaasalainen, Touko [3 ]
Kaleva, Erna [4 ]
机构
[1] Univ Helsinki, Helsinki Univ Hosp, HUS Med Imaging Ctr, Clin Physiol & Nucl Med, POB 340, FI-00029 Helsinki, Finland
[2] Univ Helsinki, Dept Phys, Helsinki, Finland
[3] Univ Helsinki, HUS Med Imaging Ctr, Radiol, Helsinki Univ Hosp, Helsinki, Finland
[4] Tampere Univ Hosp, Radiat Therapy Unit Lahti, Lahti, Finland
来源
EJNMMI PHYSICS | 2017年 / 4卷
关键词
Myocardial perfusion imaging; SPECT; Gated CT; IQ-SPECT; LEHR; VENTRICULAR VOLUME; PROGNOSTIC VALUE; TIME; PROTOCOL; SYSTEM;
D O I
10.1186/s40658-017-0187-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EFTrue). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. Results: The CT verification showed that the phantom EF settings were repeatable and accurate with the EFTrue being within 1% point from the manufacture's nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom's volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. Conclusions: The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent.
引用
收藏
页数:14
相关论文
共 23 条
[1]   Evaluation of patients with coronary artery disease IQ-SPECT protocol in myocardial perfusion imaging: Preliminary results [J].
Caobelli, F. ;
Pizzocaro, C. ;
Paghera, B. ;
Guerra, U. P. .
NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2013, 52 (05) :178-185
[2]   Feasibility of one-eighth time gated myocardial perfusion SPECT functional imaging using IQ-SPECT [J].
Caobelli, Federico ;
Thackeray, James T. ;
Soffientini, Alberto ;
Bengel, Frank M. ;
Pizzocaro, Claudio ;
Guerra, Ugo Paolo .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (12) :1920-1928
[3]   The importance of a correct positioning of the heart using IQ-SPECT system with multifocal collimators in myocardial perfusion imaging: A phantom study [J].
Caobelli, Federico ;
Kaiser, Stefano Ren ;
Thackeray, James Thomas ;
Bengel, Frank Michael ;
Chieregato, Matteo ;
Soffientini, Alberto ;
Pizzocaro, Claudio ;
Savelli, Giordano ;
Guerra, Ugo Paolo ;
Galelli, Marco ;
Zoccarato, Orazio .
JOURNAL OF NUCLEAR CARDIOLOGY, 2015, 22 (01) :57-65
[4]   IQ SPECT Allows a Significant Reduction in Administered Dose and Acquisition Time for Myocardial Perfusion Imaging: Evidence from a Phantom Study [J].
Caobelli, Federico ;
Kaiser, Stefano Ren ;
Thackeray, James T. ;
Bengel, Frank M. ;
Chieregato, Matteo ;
Soffientini, Alberto ;
Pizzocaro, Claudio ;
Savelli, Giordano ;
Galelli, Marco ;
Guerra, Ugo Paolo .
JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (12) :2064-2070
[5]   Clinical relevance of left ventricular volume assessment by gated myocardial SPET in patients with coronary artery disease [J].
De Winter, O ;
De Sutter, J ;
Dierckx, RA .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) :957-966
[6]   Advances in Cardiac Processing Software [J].
DePuey, Ernest Gordon, III .
SEMINARS IN NUCLEAR MEDICINE, 2014, 44 (04) :252-273
[7]   Corridor4DM: The Michigan method for quantitative nuclear cardiology [J].
Ficaro, Edward P. ;
Lee, Benjamin C. ;
Kritzman, James N. ;
Corbett, James R. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (04) :455-465
[8]  
GERMANO G, 1995, J NUCL MED, V36, P2138
[9]  
Havel M, 2014, HELL J NUCL MED, V17, P200
[10]   Validation of a short-scan-time imaging protocol for thallium-201 myocardial SPECT with a multifocal collimator [J].
Horiguchi, Yoriko ;
Ueda, Tomohiro ;
Shiomori, Tomofumi ;
Kanna, Masahiko ;
Matsushita, Hirooki ;
Kawaminami, Tomoko ;
Sudo, Yuta ;
Kikuchi, Shinnosuke ;
Sasaki, Ryo ;
Hoshimiya, Jun ;
Morita, Yukiko .
ANNALS OF NUCLEAR MEDICINE, 2014, 28 (08) :707-715