Automated 3D ultrasound elastography of the breast: a phantom validation study

被引:27
作者
Hendriks, Gijs A. G. M. [1 ]
Hollander, Branislav [1 ]
Menssen, Jan [1 ]
Milkowski, Andy [2 ]
Hansen, Hendrik H. G. [1 ]
de Korte, Chris L. [1 ]
机构
[1] Radboud Univ Nijmegen Med Ctr, Dept Radiol & Nucl Med, Med UltraSound Imaging Ctr MUSIC, Nijmegen, Netherlands
[2] Siemens Ultrasound, Issaquah, WA USA
关键词
3D elastography; strain; ultrasound; ABVS; breast cancer; plane wave; MAMMOGRAPHY; MORTALITY; DENSITY; VECTOR; WOMEN; RISK;
D O I
10.1088/0031-9155/61/7/2665
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In breast cancer screening, the automated breast volume scanner (ABVS) was introduced as an alternative for mammography since the latter technique is less suitable for women with dense breasts. Although clinical studies show promising results, clinicians report two disadvantages: long acquisition times (> 90 s) introducing breathing artefacts, and high recall rates due to detection of many small lesions of uncertain malignant potential. Technical improvements for faster image acquisition and better discrimination between benign and malignant lesions are thus required. Therefore, the aim of this study was to investigate if 3D ultrasound elastography using plane-wave imaging is feasible. Strain images of a breast elastography phantom were acquired by an ABVS-mimicking device that allowed axial and elevational movement of the attached transducer. Pre-and post-deformation volumes were acquired with different constant speeds (between 1.25 and 40.0 mm s(-1)) and by three protocols: Go-Go (pre-and post-volumes with identical start and end positions), Go-Return (similar to Go-Go with opposite scanning directions) and Control (pre-and post-volumes acquired per position, this protocol can be seen as reference). Afterwards, 2D and 3D cross-correlation and strain algorithms were applied to the acquired volumes and the results were compared. The Go-Go protocol was shown to be superior with better strain image quality (CNRe and SNRe) than Go-Return and to be similar as Control. This can be attributed to applying opposite mechanical forces to the phantom during the Go-Return protocol, leading to out-of-plane motion. This motion was partly compensated by using 3D cross-correlation. However, the quality was still inferior to Go-Go. Since these results were obtained in a phantom study with controlled deformations, the effect of possible uncontrolled in vivo tissue motion artefacts has to be addressed in future studies. In conclusion, it seems feasible to implement 3D ultrasound quasi-static elastography on an ABVS-like system and to reduce scan times within one breath-hold (similar to 10 s) by plane-wave acquisitions.
引用
收藏
页码:2665 / 2679
页数:15
相关论文
共 50 条
  • [31] Weakly supervised Branch Network with Template Mask for Classifying Masses in 3D Automated Breast Ultrasound
    Kim, Daekyung
    Nam, Chang-Mo
    Park, Haesol
    Jang, Mijung
    Lee, Kyong Joon
    2022 IEEE WINTER CONFERENCE ON APPLICATIONS OF COMPUTER VISION (WACV 2022), 2022, : 3212 - 3219
  • [32] Development of a physical 3D anthropomorphic breast phantom
    Carton, Ann-Katherine
    Bakic, Predrag
    Ullberg, Christer
    Derand, Helen
    Maidment, Andrew D. A.
    MEDICAL PHYSICS, 2011, 38 (02) : 891 - 896
  • [33] Initial results of the FUSION-X-US prototype combining 3D automated breast ultrasound and digital breast tomosynthesis
    Schaefgen, Benedikt
    Heil, Joerg
    Barr, Richard G.
    Radicke, Marcus
    Harcos, Aba
    Gomez, Christina
    Stieber, Anne
    Hennigs, Andre
    von Au, Alexandra
    Spratte, Julia
    Rauch, Geraldine
    Rom, Joachim
    Schuetz, Florian
    Sohn, Christof
    Golatta, Michael
    EUROPEAN RADIOLOGY, 2018, 28 (06) : 2499 - 2506
  • [34] Multi-modal artificial intelligence for the combination of automated 3D breast ultrasound and mammograms in a population of women with predominantly dense breasts
    Tan, Tao
    Rodriguez-Ruiz, Alejandro
    Zhang, Tianyu
    Xu, Lin
    Beets-Tan, Regina G. H.
    Shen, Yingzhao
    Karssemeijer, Nico
    Xu, Jun
    Mann, Ritse M.
    Bao, Lingyun
    INSIGHTS INTO IMAGING, 2023, 14 (01)
  • [35] Development of automated detection and classification methods of masses on 3D breast ultrasound images
    Hara, T
    Fukuoka, D
    Fujita, H
    Endo, T
    Moon, WK
    CARS 2002: COMPUTER ASSISTED RADIOLOGY AND SURGERY, PROCEEDINGS, 2002, : 794 - 799
  • [36] Automatic Nipple Detection on 3D Images of an Automated Breast Ultrasound System (ABUS)
    Moghaddam, Mandana Javanshir
    Tan, Tao
    Karssemeijer, Nico
    Platel, Bram
    MEDICAL IMAGING 2014: IMAGE PROCESSING, 2014, 9034
  • [37] Automated 3D ultrasound image segmentation to aid breast cancer image interpretation
    Gu, Peng
    Lee, Won-Mean
    Roubidoux, Marilyn A.
    Yuan, Jie
    Wang, Xueding
    Carson, Paul L.
    ULTRASONICS, 2016, 65 : 51 - 58
  • [38] Contrast ultrasound versus ultrasound elastography for diagnosis of breast lumps A cross-sectional study
    Zhao, Wei
    Yan, Kaituo
    Liu, Yingying
    Zhang, Zaibin
    MEDICINE, 2019, 98 (26)
  • [39] COMPARATIVE STUDY OF AUTOMATED BREAST 3-D ULTRASOUND AND HANDHELD B-MODE ULTRASOUND FOR DIFFERENTIATION OF BENIGN AND MALIGNANT BREAST MASSES
    Chen, Lin
    Chen, Yue
    Diao, Xue-Hong
    Fang, Liang
    Pang, Yun
    Cheng, Ai-Qun
    Li, Wei-Ping
    Wang, Yi
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2013, 39 (10) : 1735 - 1742
  • [40] Shear wave elastography plaque characterization with mechanical testing validation: a phantom study
    Widman, E.
    Maksuti, E.
    Larsson, D.
    Urban, M. W.
    Bjallmark, A.
    Larsson, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (08) : 3151 - 3174