A new standard reference material for column evaluation in reversed-phase liquid chromatography

被引:67
作者
Sander, LC [1 ]
Wise, SA [1 ]
机构
[1] NIST, Gaithersburg, MD 20899 USA
关键词
RP-LC; standards; reference materials; test mixture;
D O I
10.1002/jssc.200390034
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new Standard Reference Material (SRM) for use in characterizing general aspects of liquid chromatographic column performance is described. SIRM 870 Column Performance Test Mixture for Liquid Chromatography contains five components: uracil, toluene, ethylbenzene, quinizarin, and amitriptyline, and is intended to provide indications of column void volume, hydrophobicity, methylene selectivity, activity toward organic bases, and activity toward metal chelators. The test is performed under neutral, buffered conditions. The measurement of retention and peak asymmetry factors for the components facilitates column classification. It is anticipated that this material will be useful (1) in column selection during method development, (2) as a control material to monitor column performance over time, and (3) in quality control in column manufacturing.
引用
收藏
页码:283 / 294
页数:12
相关论文
共 63 条
[1]   Characterization of a range of alkyl-bonded silica HPLC stationary phases: Correlation of quantitative surface analysis data with the retention behavior of neutral, acidic, and basic solutes [J].
Barrett, DA ;
Brown, VA ;
Davies, MC ;
Shaw, PN .
ANALYTICAL CHEMISTRY, 1996, 68 (13) :2170-2178
[2]   ROLE OF THE CHAIN-LENGTH OF CHEMICALLY BONDED PHASES AND THE RETENTION MECHANISM IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
BERENDSEN, GE ;
DEGALAN, L .
JOURNAL OF CHROMATOGRAPHY, 1980, 196 (01) :21-37
[3]   CHARACTERIZATION OF A RANGE OF ALKYL-BONDED SILICA HPLC STATIONARY PHASES - XPS AND TOF-SIMS STUDIES [J].
BROWN, VA ;
BARRETT, DA ;
SHAW, PN ;
DAVIES, MC ;
RITCHIE, HJ ;
ROSS, P ;
PAUL, AJ ;
WATTS, JF .
SURFACE AND INTERFACE ANALYSIS, 1994, 21 (05) :263-273
[4]   FLUORESCENCE STUDIES OF THE STATIONARY-PHASE CHEMICAL ENVIRONMENT IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY [J].
CARR, JW ;
HARRIS, JM .
ANALYTICAL CHEMISTRY, 1986, 58 (03) :626-631
[5]   Comparative study of test methods for reversed-phase columns for high-performance liquid chromatography [J].
Claessens, HA ;
van Straten, MA ;
Cramers, CA ;
Jezierska, M ;
Buszewski, B .
JOURNAL OF CHROMATOGRAPHY A, 1998, 826 (02) :135-156
[6]   Trends and progress in the characterization of stationary phases for reversed-phase liquid chromatography [J].
Claessens, HA .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (10) :563-583
[7]   SOME MODERN CONCEPTS IN REVERSED-PHASE LIQUID-CHROMATOGRAPHY ON CHEMICALLY BONDED ALKYL STATIONARY PHASES [J].
COOKE, NHC ;
OLSEN, K .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1980, 18 (10) :512-524
[8]   A SIMPLE TEST TO CHARACTERIZE C8/C18-BONDED PHASES, PARTICULARLY FOR TOXICOLOGICAL SCREENING [J].
DALDRUP, T ;
KARDEL, B .
CHROMATOGRAPHIA, 1984, 18 (02) :81-83
[9]   CHEMOMETRIC CLASSIFICATION OF REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COLUMNS [J].
DELANEY, MF ;
PAPAS, AN ;
WALTERS, MJ .
JOURNAL OF CHROMATOGRAPHY, 1987, 410 (01) :31-41
[10]   Characterization of C18-bonded liquid chromatographic stationary phases by Raman spectroscopy:: the effect of temperature [J].
Doyle, CA ;
Vickers, TJ ;
Mann, CK ;
Dorsey, JG .
JOURNAL OF CHROMATOGRAPHY A, 2000, 877 (1-2) :41-59