Mechanical evaluation of fracture fixation augmented with tricalcium phosphate bone cement in a porous osteoporotic cancellous bone model

被引:48
作者
Collinge, Cory
Merk, Bradley
Lautenschlager, Eugene P.
机构
[1] Harris Methodist Ft Worth Hosp, Ft Worth, TX USA
[2] Northwestern Univ, Dept Orthoped Surg, Chicago, IL 60611 USA
关键词
bone cement; tricalcium phosphate cement; screw augmentation; fracture fixation; osteoporotic fixation; screw pullout; resorbable cement; locked plating;
D O I
10.1097/BOT.0b013e318033093e
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: The purpose of this study was to examine the effects of resorbable bone cement on screw and plate-screw fracture fixation in a porous osteoporotic bone model. Methods: Experiment 1: Screw pullout strength was assessed for 4 sets of 4.5-mm cortical screws inserted into a synthetic osteoporotic cancellous bone model, including screws inserted without cement augmentation (control), screws augmented with tricalciurn phosphate (TCP) bone cement (Norian SRS; Synthes USA, Paoli, PA), and screws augmented with polyrnethylmethacrylate. Experiment 2: The effects of cement augmentation on plate-screw fixation strength were examined by performing cantilever bending tests on 4 sets of 8 plate-screw constructions, including nonaugmented and TCP-augmented standard and locked screw-plate constructions in a similar bone model. Results: Experiment 1: Cement augmentation with both TCP and polymethylmethacrylate increased screw pullout strength from a porous osteoporotic cancellous bone model by about 4-fold (P < 0.05), and there was no significant difference between the 2 cements (P > 0.1). Experiment 2: Fixation strength was 1.5 times higher for locked plates compared with standard plates when neither was augmented with cement (P = 0.07). Cement augmentation with TCP improved fixation strength by 3.6 times for a standard plate-screw construction (P < 0.05) and 3.3 times for a locked plate-screw construction (P < 0.05). The most stable construction was the TCP-augmented locked plate, in which a 5-fold increase was observed compared with that of standard plates without TCP (P < 0.05). Conclusions: This study indicates augmenting screws with TCP cement during osteosynthesis improves fixation strength in an osteoporotic cancellous bone model. Clinical Relevance: In fracture situations in which osteoporotic bone makes screw and screw-plate fixation tenuous, screw augmentation with TCP cement should be considered as adjunct treatment.
引用
收藏
页码:124 / 128
页数:5
相关论文
共 43 条
[1]  
[Anonymous], BIOMECHANICS ORTHOPE
[2]   Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[3]  
Collinge CA, 2000, CLIN ORTHOP RELAT R, P277
[4]   SKELETAL REPAIR BY IN-SITU FORMATION OF THE MINERAL PHASE OF BONE [J].
CONSTANTZ, BR ;
ISON, IC ;
FULMER, MT ;
POSER, RD ;
SMITH, ST ;
VANWAGONER, M ;
ROSS, J ;
GOLDSTEIN, SA ;
JUPITER, JB ;
ROSENTHAL, DI .
SCIENCE, 1995, 267 (5205) :1796-1799
[5]  
Cook Stephen D, 2004, Spine J, V4, P402, DOI 10.1016/j.spinee.2003.11.010
[6]  
Cornell Charles N, 2003, J Am Acad Orthop Surg, V11, P109
[7]  
DEGROOT K, 1990, HDB BIOACTIVE CERAMI, V2, P3
[8]   Age variations in the properties of human tibial trabecular bone [J].
Ding, M ;
Dalstra, M ;
Danielsen, CC ;
Kabel, J ;
Hvid, I ;
Linde, F .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (06) :995-1002
[9]   Biomechanics of locked plates and screws [J].
Egol, KA ;
Kubiak, EN ;
Fulkerson, E ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) :488-493
[10]   Biomechanical evaluation of calcium phosphate cement-augmented fixation of unstable intertrochanteric fractures [J].
Elder, S ;
Frankenburg, E ;
Goulet, J ;
Yetkinler, D ;
Poser, R ;
Goldstein, S .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2000, 14 (06) :386-393